16:00 Spoke with Brian on the summit. Morning fills have been completed. I will reauthorize as LBTI on both sides and insert stowpins. I am clear to turn off the lights as needed. Checkouts will follow the procedures outlined:

https://docs.google.com/spreadsheets/d/1NKQxa0I7EDg1spfKgGOTaOy69zN9MAKZkcMgddbpaWA/edit?usp=sharing

16:15 Found the DX WFS machine to have high disk usage. I am zipping the 2022 logs. Need to find out if the LBTI WFS machines have the same monitoring for space as the SOUL LUCI WFS machines.

16:20 Found the DX Flow meters off (normally on, #6) but the DX PI Tip-Tilt SPC Controller (#7) left on. The SPC controller should not be on if the UBC is not cold so this was obviously in error. I also found the SX WFS software was left up. This should be stopped after use. Proceeding with SX checkout while DX computer cleanup is underway

16:30 The SX software took 3 tries to restart after I took it down. Once up, ok. OCAM live upon binning and when changing configurations through the WFS Arb remains so. The checklist still has the changing of binning through the hardware GUI and this should be amended. SX Filterwheel 1 did take a minute to stabilize to first position but no issues with repeated motions after homing. SX adsec set with 639 modes.

16:50 Deploying both SX and DX Argos Source

Space is much better on both SX and DX machines after gzipping up old logs:

DX:  Filesystem            Size  Used Avail Use% Mounted on /dev/mapper/vg_lbtidxwfs6-lv_root                       462G  299G  140G  69% / tmpfs                 7.8G   68K  7.8G   1% /dev/shm /dev/sda1             477M   42M  410M  10% /boot 192.168.39.55:/local/aomeas                       7.3T  5.8T  1.2T  84% /local/aomeas SX: Filesystem            Size  Used Avail Use% Mounted on /dev/mapper/vg_lbtisxwfs6-lv_root                       462G  339G  100G  78% / tmpfs                 7.8G   68K  7.8G   1% /dev/shm /dev/sda1             477M   42M  410M  10% /boot 192.168.39.54:/local/aomeas                       7.2T  5.7T  1.2T  83% /local/aomeas

17:09 Starting up DX WFS software. Failed to bring up housekeeper the first try, took 3x stop_all and start all to be successful. DX flat set with 547 modes.

17:18 Actuator 164 is not working correctly

17:21 Bringing up OCAM2K for DX, the top left and bottom right quadrants are very noisy. Rebinning to 2x2 thought the wfsarb did not resolve. Cycling the BCU2K through the PDU and rebinning through the wfsarb resolved.

17:48 SX source aligned on CCD 47 and sending presetAO for bin1. rr_mode disabled with presetAO. And the ACE parameters did not update in the mirror GUI (modal basis, reconstructor, filtering). PresetAO hung at running in AOS commands GUI. Waited 15 min for a timeout but it never did. Tried to stopAO but no impact and no error in the AOS log. Rested shell and stopped and started AOG left. Flat set with 639 modes. Tertiary in default position
SX Hexapod Positions:
SX Argos Alignemnt.png

17:28 ARGOS source aligned on DX. When loop closed the CCD47 went now live causing loop to go crazy and open.
  • DX Hexapod Positions, tertiary in default authorized position.
    DX Argos Alignment.png BCU47 I now unpingable: ping 192.168.153.55

It is showing offline on the hardware GUI. Writing note to IT8760. WIthout the BUC47 there is little that can be done besides ACE table checkouts. I will park the ARGOS swing arms and hand back over to the mountain while the other admin work is executed.

18:50 Shutting down DX. Set safe state (parking filterwheels, warming OCAM, set frame rate), rest the shell. Removing hexapod offsets, re enabling look up tables, and disabling rr mode.

18:54 Shutting down SX. Set safe state (parking filterwheels, warming OCAM, set frame rate), rest the shell. Removing hexapod offsets, re enabling look up tables, and disabling rr mode.

19:00 Turning off ARGOS sources and parking. Chamber lights on. DX parked, shutter fully closed and off, SX parked, shutter fully closed and off.

I am configuring the ACE tables and setups for 4runner and SHARK

-- JenniferPower - 01 Mar 2023
  • *
I Attachment Action Size Date Who Comment
DX Argos Alignment.pngpng DX Argos Alignment.png manage 9 K 01 Mar 2023 - 18:40 JenniferPower DX Hexapod Positions
SX Argos Alignemnt.pngpng SX Argos Alignemnt.png manage 10 K 01 Mar 2023 - 18:40 JenniferPower SX Hexapod Positions
Topic revision: r1 - 01 Mar 2023, JenniferPower
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback