Staff.

LBTO: Andrew C, Josh W, Alex B, Juan Carlos G Dave T.

Observer: Enrico et al. Arcetri(remote)

Forecast.

forecast0_20230221.png

Details.

  • Officially call the night at 3.30PM
  • Dave runs some Luci tests.
  • Second half for SoulDX.
  • Configuring at midnight.
11:30PM
  • Power on WFS.
  • Chamber temp 4.5
  • ocam OK
  • adsec flat
06:50:01.972909| MAIN > Maximum mode index used to apply flat -> # 623

11:53

test preset.

Wed Feb 22 06:53:23.673 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[433] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 06:53:23.753 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1700.00_sd3892.fits

optimize: 20230221_070025

TN+Luci:20230222_070306,_070613 (0053),_070728 (0054),_070843 (0055)

TN:_070437,_070501,_070517,_070532,_070554

ModalSR:85

luci dark:uci2.20230222.0056.fits

12:14am
  • Preset Bin4
  • mag 16.521
  • dist s1.5@500
  • check flux moved to bin3 freq 268Hz.
  • TN+Luci (J): 202302222_072130
12:25am

presetAO

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 07:25:35.421 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

TN+Luci:_073835,_074033,_074225,_074448,_075112

luci dark: 0062,0063

subframe:_080524,_080726,_080852,_081713,_081918,_082246

dark:0072(20230222_081216),0073

Point for today with Enrico:
  • Bin4 mag 16.51 performance with luci with two seeing 1.5 and 1

  • Bin3 and Bin4 where to change vs mag

  • Bin3 and Bin4 optimizing frequency.

1:31am

rise TT gain to check the 87Hz.

TN:_083139,_083211,_083240,_083256

tt_bin4_87_2023022.png

01:39am

presetAO
  • same setup but the frequency is changed to go slower (300Hz)
Wed Feb 22 08:39:37.969 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[451] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 08:39:37.987 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 08:39:38.061 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

optimize gain several times. Enrico changed the scanning to go further.

TN+luci:_085233,_085401,_085536,_085701,_085922

02:00

Wed Feb 22 09:06:25.636 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 09:06:25.704 2023 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[455] - [AOException] presetAO: FATAL - [Errno 28] No space left on device

remove the core dumps from live/aoroot/bin.

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 09:08:43.837 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

Andrew just commented that shark-NIR is doing motors tests.

TN+luci:_091902,_092419,_092637,_092223,_092836

2:37

PresetAO
  • changed the setup f frequency for 750Hz.
  • AO table-> table_LUCIFER_ACE-AO_20230222_500m_TESTING02.txt
  • TN+luci:_094602,_094742,_095108,_095256,_095432,_095610

3:03

PresetAO
  • Keep the set but changing frequency to 300Hz.
  • AO table -> table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO_20230222_500m_TESTING01.txt
Wed Feb 22 10:03:07.500 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[464] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 10:03:07.569 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.00_sd3892.fits
Optmimze:20230222_100643

TN+Luci:_100850,101031,_

3:12 big jump the psf moved away from the Py

loop stopped.

3:14
  • preset again.
  • optmize:_101730
  • TN+luci:101920,_102054,_102224,_102353,_102535,_102706
3:32

presetAO
  • set to default frequency 500Hz.
  • tableA0->table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO_20230221_500m_v01.txtt

ocam not live.

interrupt command from the FLAO

stop/start the AOS arbitrator.

3:41

PresetAO
  • the camera lens is not moving correctly as soon as the loop is closed the update of the diff x and y is not happening.
  • power cycle bcu2k and reconfigure.
  • same behaviour.
  • power cycle W soft
eb 22 10:57:23.776 2023 aos.AcquireRef{FLAO}.complete - AcquireRef{FLAO}[476] - dx:2.19953 dy:-0.52009 slNull: f1:0.0 f2:1.0 freq:491 mag:0 nBins:4 nModes:55 r0:nan snMode:0 strehl:0 ttMod:0
Wed Feb 22 10:57:46.903 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[477] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 10:57:46.922 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 10:57:46.968 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

290|ERR| 95|2023-02-22 10:58:20.375922| MAIN > Cannot turn on PROG47. Reason: Timeout waiting for powerboard.R.PROG_B
  • reconfigure CCD7 from hw gui
  • ok
  • ocam2k not live
  • reconfgure
  • ok
Wed Feb 22 11:01:29.308 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[478] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 11:01:29.327 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 11:01:29.371 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

The loop in teh center pupils is showing to be very slow. change the refresh of the .

Check clse loop wth gain zero and it shows the same behaviour.

reduced_loop_20230222.png

Adsec rest/set

send presetAO

same behaviour.

4:16
  • adsc_stop/start
  • and w_stop/start_all
AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 11:21:23.438 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 11:21:23.469 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits
  • optimize:_112751
  • TN+luci:112937,_113110,_113316,_113441,_113842,_114027
  • modalSR20
4:42

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.49,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 22 11:42:37.907 2023 aos.info - WFS source set to: FLAOWFS
Wed Feb 22 11:42:37.999 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.00_sd3892.fits
  • optimize:115042
  • TN+Luci:_115311,_115506,_115722,_115936,_120113,_120307. The TN shows the PSF off the subframe.
  • recenter the PSF using a new subframe region
  • TN+luci:_121510,_121652,_121816,_121948,_122122,_122300
  • luci dark: 0128,0129

5:34
  • preset to check the resonance at 87Hz.
  • no disturbance
  • rise the shutter to 85
  • mag 15.1
Wed Feb 22 12:32:37.205 2023 aos.Stop{FLAO}.complete - Stop{FLAO}[491]
Wed Feb 22 12:33:56.401 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[492] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.51,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

manula gain

_123848,_123909,_123924,_124001,_124023,_124049

No visible 87Hz.

5:47am

STOP

-- JuanCarlosGuerra - 22 Feb 2023