Summary.

Pupil wobble for Bin3 and Bin4. The bin3 and bin4 are in the operational range. The Bin3 touches the limit in X negative but the pupil errors are <0.1, so it is inside of the operation ranges. A shift of one pixel n x positive done and the performance were not good, looking a the PSF corrected. it should require to build another pupil shifted and tests it.

The error of the enable RR mode comes after the presetAO is completed. This new behaviour probably is due to the software update. IT: related to the enable/disable of the RR mode: #8830

Details.

11:30AM
  • Power on
  • OK
  • Ocam
  • OK

Argos cal unit deployed:
  • time:
  • 18:33:49.15 command.
  • 18:34:01 reach working positions.
Sun Feb 12 18:46:27.105 2023 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1534] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • Fan_47 and FAN_LN47 off.
Time to reach temperature ->20 minutes from -1 to 7C

adsec flat.
  • 18:55:12.375426| MAIN > Maximum mode index used to apply flat -> # 639

Bin3 rotation

12:15PM
  • backward.
  • ok
  • forward
  • ok
  • Possible one pixel shit positive in X.
  • Errors <0.1 subaperture,
  • pupil still in the operational range.
bn3_20230212.png

bin4 rotation.

12:34PM
  • shutter 50
  • mag 15
  • TN:20230212_193848
12:39PM
  • backward.
  • ok
  • forward.
  • ok
bin4_20230212.png

Shift of Bin3.

Astropy v1.2.0 and above [astropy]
>>> makePupils.movePupils('20220901_070135',1,0,240,4)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20230212_202302/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20230212_202302/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 148
out_of_pupil_slopes 2552
lines_to_add 11
residual_pixels_to_add 38
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20230212_202302
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20230212_202302"
'20230212_202302

The oginal pupil of the rec is: 20200210_000000

test pupil
  • TN:_202723
  • backward.
  • ok
  • forward
  • ok

RR mode enable and disables after the presetAO completed.

Sun Feb 12 20:49:25.007 2023 aos.info - Updated variable AOARB.L.RR_ENABLED: 1
Sun Feb 12 20:50:30.866 2023 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1542] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sun Feb 12 20:50:30.882 2023 aos.info - WFS source set to: FLAOWFS
Sun Feb 12 20:50:30.936 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits

bin3 presetAO with default pupil and image quality

modalSR 55%

TN+luci(J): _205718,_210002

luci1.20230213.0002, _0003.fits

luci dark: luci1.20230213.0004.fits

loaded the new Bin3 pupil

Sun Feb 12 21:02:50.716 2023 aos.Stop{FLAO}.complete - Stop{FLAO}[1547]
Sun Feb 12 21:02:58.773 2023 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1548] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sun Feb 12 21:02:58.791 2023 aos.info - WFS source set to: FLAOWFS
Sun Feb 12 21:02:58.844 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits

Sun Feb 12 21:13:15.436 2023 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1553] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sun Feb 12 21:13:15.455 2023 aos.info - WFS source set to: FLAOWFS
Sun Feb 12 21:13:15.507 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits
Sun Feb 12 21:13:16.210 2023 aos.info - FLAO: WFS: loading setup for LUCIFER

un Feb 12 21:13:47.023 2023 aos.info - FLAO: WFS: preset done
Sun Feb 12 21:13:47.041 2023 aos.info - Updated variable AOARB.L.LAB_MODE: 0
Sun Feb 12 21:13:47.057 2023 aos.info - Updated variable AOARB.L.RR_ENABLED: 0
Sun Feb 12 21:13:47.074 2023 aos.info - AO loop opened, Sound=leftAOOPENED.wav
  • optimize gain:
  • 3.4,0.85,0.10
  • not good corrections.
  • TN+Luci: _211822,_212008

roll back the AO table.

Close loop check Bin3

12 21:27:02.490 2023 aos.info - FLAO: WFS: preset done
Sun Feb 12 21:27:02.507 2023 aos.info - Updated variable AOARB.L.LAB_MODE: 0
Sun Feb 12 21:27:02.524 2023 aos.info - Updated variable AOARB.L.RR_ENABLED: 0
Sun Feb 12 21:27:02.541 2023 aos.info - AO loop opened, Sound=leftAOOPENED.wav
Sun Feb 12 21:27:02.563 2023 aos.info - Updated variable AOARB.L.SL_READY: 1

optimize: 3.15,0.82,0.1

gopt: 0.65,0.17,0.04

TN+luci : _213154,_213301

modalSr:50

Good corrections.

2:33PM Stop

park agos: 21:35:32

recevedd comd: 21:35:44

IT: related to the enable/disable of the RR mode: #8830

-- JuanCarlosGuerra - 12 Feb 2023
Topic revision: r2 - 13 Feb 2023, JuanCarlosGuerra