Staff:

Lbto: Juan carlos

Observer: Fernando, Simone (Shark-vs team). Steward remote control room

Summary

Details.

power on 10:00

ocam ok

test close loop

1300

gain2,0.20,0.3

10:30

Check with the adsecDX, Guido and Xianyu check

11:10

close loop for halpha

ncpa on for halpha_direct.

Focus Halpha checkoput. -25.47

insert Halpha + Cube.

baysede:

x=43.66,y=-35.72

z->-20.43

save focus file

11:40

loop open skip frames.

to investgate.

reclose

optimize NCPA for halpha_cube

values saved as Halpha_direct. generated fits and txt

12:29

saved 1ms open loop (no hvs internal tt), loop 1200Hz, good for daytime jitter estimation compare with the AO telemetry

Tn:20240112_193017

13:00

change to lmicrNO

check ncpa and focus.

saved ncpa file

moving Z

coil disabled 20:24
  • AOS-> FLAO alert [code:0] error.
chmber temperature -4.72

adsc_stop/start

flat ->546 modes.

look ok

w_stop/start_all

coil disabled from fastdiagnostic 20:20:01
  • 20:23:59 timeout-will emergencyReact
  • 20:24:01 timeout-will emergencyReact Now.
  • 20:24:01. EmergencyReact=1

change to HalphaNo.

Check ncpa

loop close.

move baysde y =0.2mm loop skipping.

stopAO didn;t reflat.

adsec rest/set

slope zero event.

restart gopt and loop restablished.

focus HalphaNo to -25.73

to change the file.

02:04

starting ncps scanning Z5#

saved file

02:23 StopAO

restract argos.

heaxspod: -4.7,-4.5,0.70 245 -50

chamber temp -4.6

z7# appled -1800

-- JuanCarlosGuerra - 12 Jan 2024
Topic revision: r1 - 12 Jan 2024, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.