20221219_SX_Day Test Presets for LBTI and Shark

Start with LBTI Binocular Presets

07:05 David authorize LBTI on both side and power on HBS and Drives

07:08 Power on LBTI SX and DX
  • DX bcu47 will not power on

07:42 LBTI Binoc Preset
  • DX Preset Fail
  • IIFGUI
Mon Dec 19 07:42:05.355 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: Right side swing arm (teritiary) configuration wrong
Mon Dec 19 07:42:05.383 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: Stop{FLAO}[514]
Mon Dec 19 07:42:05.400 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: PresetAO{FLAO}[515] - [AOException] presetAO: FATAL - Timeout waiting for techviewctrl.R.STATUS (-5001) TIMEOUT_ERROR
Mon Dec 19 07:42:05.418 2022 right Preset command failed for LBTI mode ADAPTIVEACE_TRACK with guidestars, AO reference, Sound=presetFailed.wav
Mon Dec 19 07:42:06.017 2022 left Preset command wait for telescope complete for LBTI
  • The DX message was expected as the DX bcu47 is not powered on
  • SX preset successful
Mon Dec 19 07:42:49.471 2022 left Preset alert for LBTI mode ADAPTIVEACE_TRACK: Left side swing arm (teritiary) configuration wrong
Mon Dec 19 07:42:49.485 2022 left Preset alert for LBTI mode ADAPTIVEACE_TRACK: Stop{FLAO}[526]
Mon Dec 19 07:42:49.502 2022 left Preset alert for LBTI mode ADAPTIVEACE_TRACK: PresetAO{FLAO}[527] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:900 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
Mon Dec 19 07:42:49.520 2022 left Preset command complete with warnings for LBTI mode ADAPTIVEACE_TRACK with guidestars, AO reference, Sound=presetWarning.wav
,
  • AOSGUI left
Mon Dec 19 07:54:37.559 2022 aos.Stop{FLAO}.started - Stop{FLAO}[528] (msg:Starting Preset)
Mon Dec 19 07:54:37.592 2022 aos.Stop{FLAO}.complete - Stop{FLAO}[528]
Mon Dec 19 07:54:41.399 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[529] (aomode:ACE-AO,focst:bentGregorianCenter,instr:LBTI,wfs:LBTIWFS,sox:0[0],soy:0,rox:-0.00414804[-0.00414804],roy:0.00381537,mag:8.25,cindex:1.88,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Dec 19 07:54:41.423 2022 aos.info - WFS source set to: LBTIWFS
Mon Dec 19 07:54:42.047 2022 aos.info - FLAO: WFS: loading setup for LBTI
Mon Dec 19 07:54:42.098 2022 aos.info - FLAO: WFS: configuring CCDs
Mon Dec 19 07:54:58.406 2022 aos.info - FLAO: WFS: preset done
Mon Dec 19 07:54:58.423 2022 aos.info - Updated variable AOARB.L.LAB_MODE: 0
Mon Dec 19 07:54:58.442 2022 aos.info - AO loop opened, Sound=leftAOOPENED.wav
Mon Dec 19 07:54:58.465 2022 aos.info - Updated variable AOARB.L.SL_READY: 1
Mon Dec 19 07:54:58.484 2022 aos.info - Updated variable AOARB.L.AO_READY: 0
Mon Dec 19 07:54:59.366 2022 aos.info - Updated variable AOARB.L.AO_READY: 1
Mon Dec 19 07:54:59.366 2022 aos.PresetAO{FLAO}.complete - PresetAO{FLAO}[529] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:900 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0

07:57 DX bcu47 now reports powered on
  • DX presetAO successful
  • IIFGUI
Mon Dec 19 07:57:21.558 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: Right side swing arm (teritiary) configuration wrong
Mon Dec 19 07:57:21.576 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: Stop{FLAO}[518]
Mon Dec 19 07:57:21.598 2022 right Preset alert for LBTI mode ADAPTIVEACE_TRACK: PresetAO{FLAO}[519] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:1700 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
Mon Dec 19 07:57:21.619 2022 right Preset command complete with warnings for LBTI mode ADAPTIVEACE_TRACK with guidestars, AO reference, Sound=presetWarning.wav
  • AOSGUI right
Mon Dec 19 07:56:14.935 2022 aos.Stop{FLAO}.started - Stop{FLAO}[518] (msg:Starting Preset)
Mon Dec 19 07:56:14.953 2022 aos.Stop{FLAO}.complete - Stop{FLAO}[518]
Mon Dec 19 07:56:18.545 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[519] (aomode:ACE-AO,focst:bentGregorianCenter,instr:LBTI,wfs:LBTIWFS,sox:0[0],soy:0,rox:-0.00417119[-0.00417119],roy:0.00565338,mag:8.25,cindex:1.88,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Dec 19 07:56:18.579 2022 aos.info - WFS source set to: LBTIWFS
Mon Dec 19 07:56:19.163 2022 aos.info - FLAO: WFS: loading setup for LBTI
Mon Dec 19 07:56:19.176 2022 aos.info - FLAO: WFS: configuring CCDs
Mon Dec 19 07:57:20.692 2022 aos.info - FLAO: WFS: preset done
Mon Dec 19 07:57:20.713 2022 aos.info - Updated variable AOARB.R.LAB_MODE: 0
Mon Dec 19 07:57:20.750 2022 aos.info - AO loop opened, Sound=rightAOOPENED.wav
Mon Dec 19 07:57:20.792 2022 aos.info - Updated variable AOARB.R.SL_READY: 1
Mon Dec 19 07:57:20.831 2022 aos.info - Updated variable AOARB.R.AO_READY: 0
Mon Dec 19 07:57:21.448 2022 aos.info - Updated variable AOARB.R.AO_READY: 1
Mon Dec 19 07:57:21.449 2022 aos.PresetAO{FLAO}.complete - PresetAO{FLAO}[519] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:1700 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0

08:00 iif_binoc_runao
Mon Dec 19 08:00:48.776 2022 aos.AcquireRef{FLAO}.started - AcquireRef{FLAO}[530] Mode: Move stages, Automatic acquire
on Dec 19 08:00:48.766 2022 aos.AcquireRef{FLAO}.started - AcquireRef{FLAO}[520] Mode: Move stages, Automatic acquire
  • Cancel preset on both FLAO GUIs
  • OK
  • IIFGUI shows runAO failed on both sides

08:02 Modify IDL routines
  • station_define.pro: change LBTI name to SHARK
  • instrument_common.pro: Add LSHARK and RSHARK

08:11 iif_binoc_preset
  • Both SX and DX were successful
  • IIFGUI
Mon Dec 19 08:11:29.021 2022 right Preset alert for SHARK mode ADAPTIVEACE_TRACK: Right side swing arm (teritiary) configuration wrong
Mon Dec 19 08:11:29.033 2022 right Preset alert for SHARK mode ADAPTIVEACE_TRACK: Stop{FLAO}[526]
Mon Dec 19 08:11:29.049 2022 left Preset command wait for telescope complete for SHARK
Mon Dec 19 08:11:29.050 2022 right Preset alert for SHARK mode ADAPTIVEACE_TRACK: PresetAO{FLAO}[527] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:1700 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
Mon Dec 19 08:11:29.093 2022 right Preset command complete with warnings for SHARK mode ADAPTIVEACE_TRACK with guidestars, AO reference, Sound=presetWarning.wav
Mon Dec 19 08:11:29.198 2022 left Preset alert for SHARK mode ADAPTIVEACE_TRACK: Left side swing arm (teritiary) configuration wrong
Mon Dec 19 08:11:29.220 2022 left Preset alert for SHARK mode ADAPTIVEACE_TRACK: Stop{FLAO}[536]
Mon Dec 19 08:11:29.239 2022 left Preset alert for SHARK mode ADAPTIVEACE_TRACK: PresetAO{FLAO}[537] - dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:900 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
Mon Dec 19 08:11:29.257 2022 left Preset command complete with warnings for SHARK mode ADAPTIVEACE_TRACK with guidestars, AO reference, Sound=presetWarning.wav
  • In WFS log on WFS Control GUI
WfsArbitrator_8752 |DEB|       164|2022-12-19 08:11:08.348242|             MAIN > Loading setup file /home/sxwunit/live/aoroot/calib/wfs/current/setups/SHARK_ACE-AO
WfsArbitrator_18893|DEB|       179|2022-12-19 08:11:08.387380|             MAIN > Loading setup file /home/dxwunit/live/aoroot/calib/wfs/current/setups/SHARK_ACE-AO
  • Thus, the AO Tables for SHARK on both sides were found
  • The command and return from the IIF to my IDL routines was<
IDL> iif_register, /TCS
IDL> iif_register
 
**********************************************
*
*  =>  LEFT Registered:   LCBG: LSHARK
*  => RIGHT Registered:   RCBG: RSHARK
*
**********************************************
 
IDL> iif_preset, AO247, MODE=ACE, GCS=TRACK, GS=0, RS=0, /LEFT
IDL> iif_preset, AO247, MODE=ACE, GCS=TRACK, GS=0, RS=0, /right
IDL> iif_binoc_preset
         # Target star: 0   AO247
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 1.646
         # Guide star: 0
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 0.630
         # AO Reference star: 0
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 0.85
  
     IRCL&: preset, AO247, MODE=ADAPTIVEACE_TRACK, GS=0, RS=0, PA=0.00000     (iif_preset)
  
         # Target star: 0   AO247
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 1.646
         # Guide star: 0
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 0.630
         # AO Reference star: 0
           06:21:50.481 +31:56:44.77 RADEC 2000.0 J2000 9.5 -7.3 8.25 R 1.88 B_V 0.85
  
      IRCR: preset, AO247, MODE=ADAPTIVEACE_TRACK, GS=0, RS=0, PA=0.00000     (iif_preset)
  
  => Warning in irc_presetTelescope
        msg = Resmsg[0]  = PresetTelescope result status: Warning
        msg = Resmsg[1]  = Right side swing arm (teritiary) configuration wrong
        msg = Resmsg[2]  = Stop{FLAO}[526]
        msg = Resmsg[3]  = PresetAO{FLAO}[527] -  dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:1700 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
===> Error in left irc_preset (iif_read_preset_output
        err =Resmsg[0]  = PresetTelescope result status: Warning
        err =Resmsg[1]  = Left side swing arm (teritiary) configuration wrong
        err =Resmsg[2]  = Stop{FLAO}[536]
        err =Resmsg[3]  = PresetAO{FLAO}[537] -  dx:nan dy:nan slNull: f1:1.0 f2:2.0 freq:900 mag:0 nBins:1 nModes:500 r0:nan snMode:0
        err = strehl:0 ttMod:0

IDL> iif_binoc_runao  (cancel preset on both FLAO GUIs)
===> Error in right irc_runao
        err = Resmsg[0]  = RunAO result status: Error
        err = Resmsg[1]  = AcquireRef{FLAO}[528] -  dx:0 dy:0 slNull: f1:1.0 f2:2.0 freq:1700 mag:0 nBins:1 nModes:500 r0:nan snMode:0 strehl:0 ttMod:0
        err = Resmsg[2]  = StartAO{FLAO}[529] - Illegal command for state AOArbitrator.Ready
        err = Resmsg[3]  = StartAO{FLAO}[529] - Illegal command for state AOArbitrator.Ready
% Compiled module: IIF_READ_RUNAO_OUTPUT.
===> Error in left irc_runao (iif_read_runao_output)
        err = Resmsg[0]  = RunAO result status: Error
        err = Resmsg[1]  = AcquireRef{FLAO}[538] -  dx:0 dy:0 slNull: f1:1.0 f2:2.0 freq:900 mag:0 nBins:1 nModes:500 r0:nan snMode:0 s
        err = trehl:0 ttMod:0
        err = Resmsg[2]  = StartAO{FLAO}[539] - Illegal command for state AOArbitrator.Ready
        err = Resmsg[3]  = StartAO{FLAO}[539] - Illegal command for state AOArbitrator.Ready

08:20 Preset Tests Complete
   * LBTI PresetAO and RunAO work
   * SHARK PresetAO and RunAO work (the same as LBTI but with SHARK AO Table read)

08:22 Power Off LBTIs
   * Off with PDU 5, 8, 15, 16
   * w_stop_all on both
   * Note, SX OCAM reports ambient at 1.7C and DX as 6.0C

DONE

-- Main.DougMiller - 19 Dec 2022