20221206_SX_Day SX S-LUCI Check after spare WFS tests

18:00 Everything ready and aligned

18:20 Loop closed bin 1
  • no seeing
  • Gains 2.0, 0.05, 0.05
  • About 3x jitter on auxLoop plot due to LN pumps

18:30:17 Switch BCU start having troubles (according to Xianyu)

18:39:05 AdSec RIP and Failure

18:47 AdSec running and set again.
  • Did not restart WFS SW (test if needed according to Fabio)

18:53 Loop Closed
  • no seeing
  • Gain 1.6, 0.05, 0.05
  • About 3x jitter on auxLoop plot due to LN pumps
  • TN 185857, 190001, 190037

19:26 Closed loop several time
  • WFS SW did not know we were in RR mode
    • Stop started WFS Arbitrator: Nope
    • disable/enable_rr_mode: Yep
  • 1.0" (2.0") seeing
  • Optimize Gain 3.22, 0.67, 0.62
  • Optical Gain 0.47
  • TN 193019, 193304, 193318
  • See 127 Hz vibration in jitter plot
  • Optimize gain again, 2.93, 2.85 and do not see 127 Hz
  • Optimize again. couple clicks to 3.18. See a bit of 127
  • Optimize again, 3,22, 0.71, 0.62
  • Change autogain.conf so TT use 0.8 instead of 0.9
  • Optimize gain: 2.46, 0.67, 0.62
  • Optical Gain
  • TN 200344, 200410, 200422
  • Optimize gain: TT

20:38 Changed pointer to ao table
  • cp table_LUCIFER_ACE-AO.txt.20210121 table_LUCIFER_ACE-AO.txt.20221206
  • ln -sib table_LUCIFER_ACE-AO.txt.20221206 table_LUCIFER_ACE-AO.txt
# Columns:
# 0    1      2   3     4      5      6                7                          8                 9    10       11       12       13        14    15     16      17         18        19        20     21    22
Rmag   flux   bin freq  emGain base   rec              filtering                  pupils            tt   fw1_acq  fw2_acq  fw1_obs  fw2_obs   TVbin TVfreq TVavg   TVflux CL->freq     emG  GOpt->mode   amp  freq

-1.5 1.054E9  1  1700  1       KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    3         2       0        2         4     4.3     20     1250000     200        1        30    10e-9  80.0
0.0   2.65E8  1  1700  1       KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    3         2       0        2         4     4.3     20     1250000     200        1        30    10e-9  80.0
3.5  10.54E6  1  1700  10      KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    3         2       0        2         4     4.3     20       15000     200        1        30    10e-9  80.0
4.5   4.20E6  1  1700  20      KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        2         4     4.3     20       15000     200        3        30    10e-9  80.0
5.5   1.67E6  1  1700  30      KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        2         4     4.3     20        5000     200        6        30    10e-9  80.0
6.5   6.65E5  1  1700  100     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        2         4     4.3     20        2000     200       12        30    10e-9  80.0
7.5   2.65E5  1  1700  100     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        2         4     4.3     20         800     200       25        30    10e-9  80.0
8.5   1.05E5  1  1700  100     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        2         4     4.3     20         800     200       50        30    10e-9  80.0
9.5   4.20E4  1  1700  300     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         2       0        1         4     4.3     20         800     200      100        30    10e-9  80.0
10.5  1.67E4  1  1700  600     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         1       0        1         4     4.3     20         800     200      300        30    10e-9  80.0
11.4  7.30E3  1  1350  600     KL_v20 20190930_161000  iir1700HzV003              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
11.4  7.30E3  1  1350  600     KL_v20 20190930_161000  iir1350HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
11.7  5.53E3  1  1000  600     KL_v20 20190930_161000  iir1350HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
11.7  5.53E3  1  1000  600     KL_v20 20190930_161000  iir1000HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
12.5    2648  1   750  600     KL_v20 20190930_161000  iir1000HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
12.5    2648  1   750  600     KL_v20 20191213_000000  iir0750HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
12.7    2202  1   625  600     KL_v20 20191213_000000  iir0750HzV001              20181103_001410   3    4         1       0        1        16     12.6    20         800     200      600        30    10e-9  80.0
12.7    2202  2  1250  600     KL_v20 20191011_125000  iir1350HzV001              20220901_074516   3    4         1       0        1        16     12.6    20         300     200      600        30    15e-9  80.0
13.0    1671  2  1000  600     KL_v20 20191011_125000  iir1000HzV003              20220901_074516   3    4         1       0        1        16     12.6    20         300     200      600        30    15e-9  80.0
14.0     665  2   800  600     KL_v20 20191011_125000  iir1000HzV003              20220901_074516   3    4         1       0        1        16     12.6    20         300     200      600        30    15e-9  80.0
14.5     420  2   750  600     KL_v20 20191011_125000  iir1000HzV003              20220901_074516   3    4         1       0        1        16     12.6    20         300     200      600        30    15e-9  80.0
14.5     420  3  1000  600     KL_v20 20200922_150000  iir1000HzV004              20220901_070135   3    4         1       0        1        16     12.6    20         300     200      600        30    20e-9  80.0
15.5     167  3   750  600     KL_v20 20200922_150000  iir1000HzV004              20220901_070135   3    4         1       0        1        16     12.6    50         300     200      600        30    20e-9  80.0
15.5     167  3   750  600     KL_v20 20200922_150000  iir0750HzV003              20220901_070135   3    4         1       0        1        16     12.6    50         300     200      600        30    20e-9  80.0
15.8     135  3   500  600     KL_v20 20200922_150000  iir0750HzV003              20220901_070135   3    4         1       0        1        16     12.6    50         300     200      600        30    20e-9  80.0
15.8     135  3   500  600     KL_v20 20200922_153500  forgettingFactorBin4v2SOUL 20220901_070135   3    4         1       0        1        16     12.6    50         100     200      600        30    20e-9  80.0
16.5      67  3   250  600     KL_v20 20200922_153500  forgettingFactorBin4v2SOUL 20220901_070135   3    4         1       0        1        16     12.6    50         100     200      600        30    20e-9  80.0
16.5      67  4   500  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50         100     200      600        30    30e-9  45.0
17.1    38.3  4   285  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50         100     200      600        30    30e-9  45.0
17.5    26.5  4   190  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50        1000     100      600        30    30e-9  45.0
18.5    10.5  4   100  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50         100     100      600        30    30e-9  45.0
19.5     4.2  4   100  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50          10     100      600        30    50e-9  45.0
20.5     1.7  4   100  600     KL_v20 20200923_180000  forgettingFactorBin4v2SOUL 20220909_160320   3    4         1       0        1        16     12.6    50          10     100      600        30    50e-9  45.0

20:41: PresetAO with 1.7 kHz table
  • AcqureRefAO
  • 1.7 kHz, 3 l/D, EMg 490
  • Gain 1.7, 0.41, 0.45
  • GOPT 0.47
  • TN 204658, 204729, 204743

20:48 Optimize gain
  • gain skipframe on second TT and loop open
  • re-close loop
  • opt gain 1.74, 0.44, 0.54

20:55 Modify autogain.conf
  • changed max_iterations from 10 to 3

20:59 Open Loop, PresetAO
  • AcquireRefAO
  • gain 1.51, 0.41, 0.47
  • TN 210351
  • Opt gain: 1.55, 0.41, 0.45
  • TN 210745

21:12 Mdifiy AO table
  • Set EMG to LBTI Values for 8.5 -> 3.5 (up so more flux)
  • Set bin1_fo_end_tt from 3.0 to 2.0
  • gain 1.56, 0.43, 0.43
  • gain 1.6, 0.44, 0.42
  • GOPT
  • TN 212651
  • gain 1.58, 0.44, 0.45
  • gain 1.56, 0.42, 0.45
  • gain 1.52, 0.41, 0.45

21:47 Set ARGOS light to 68
  • mag 11.4, send 11.41
  • AcquireRefAO
  • 1334 Hz, 3 l/D, 600
  • gain 1.57, 0.52, 0.45
  • Change bin1_f1_ent_tt from 1.6 to 2.0
  • Change bin1_f1_max_iteration from 10 to 2
  • gain 1.62, 0.51, 0.45
  • gain ...
  • TN 220304
  • Change bin1_f1_factor for TT from 0.8 to 0.7
  • gain 1.38, 0.45, 0.45
  • gain 1.27, 0.46, 0.46

22:09 Mag 12.65
  • 653 Hz, 3 l/D, 600
  • gain 2.38, 0.66, 0.66
    bin1_f3_sequence   string "tt ho1 ho2 tt ho1 ho2"
    bin1_f3_factors    string "0.5 0.9 0.9 0.8 0.9 0.9"
    bin1_f3_fmin       float  400.01
    bin1_f3_fmax       float  750
    bin1_f3_min        float  0.1
    bin1_f3_start      float  0.1
    bin1_f3_end_tt     float  3.0
    bin1_f3_end_ho1    float  2.0
    bin1_f3_end_ho2    float  1.5
    bin1_f3_step_tt    float  0.5
    bin1_f3_step_ho1   float  0.2
    bin1_f3_step_ho2   float  0.1
    bin1_f3_cycles     int    5
    bin1_f3_nframes    int    50
    bin1_f3_max_tt     float  12.0
    bin1_f3_max_ho1    float  12.0
    bin1_f3_max_ho2    float  12.0
    bin1_f3_ho_middle  int    120
    bin1_f3_quadratic_tt        int    0
    bin1_f3_refine_down_factor  float  0.5
    bin1_f3_refine_up_factor    float  1.5
    bin1_f3_max_iterations      int 3
    
  • gain 2.73, 0.65, 0.64
  • Change max iter from 3 to 2
  • gain 1.95, 0.78, 0.62
  • gain 2.74, 0.65, 0.1
  • gain 2.69, 0.78. 0.53
  • Tn 222025, 222042

-- DougMiller - 06 Dec 2022
Topic revision: r1 - 06 Dec 2022, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback