Details.

08:30 am RR on.

09:00 am

Test the power on/off sequence.
  • to see that the power off park the cube translator. align the power on/off with SX.

Hexapod positions:

-0.884, 0.431,-0.5,-205,10.5

09:10am

Telescope zenith

09:12am

Briefing

Bin2 calib, check bin1

Power on WFS.

Time out bayside stageX and Y.

Bayside stage X and Y not current positions: -0.067 current 0.01A

Y: position -0.032, current 0.01A

Power off.

114778|INF| 119|2022-12-05 16:17:17.843388| MAIN > Parking the XYZ bayside stages
WfsArbitrator_114778|INF| 120|2022-12-05 16:18:17.944043| MAIN > Stage movement timeout after 60 seconds. Retry 1 of 3

WfsArbitratator _114778|INF| 121|2022-12-05 16:19:21.080648| MAIN > Stage movement timeout after 60 seconds. Retry 2 of 3

Bayside stage Drivers off doesn't reflect the status should be off.

Manual power off.

w_stop/start_all

09:27am

WFS Power on.

BaysidestageX and Y not doing the home.
  • a single command didnt move the stage.
  • Call Alessandro for support.
9:58am

Move RFBG rotaqtor to 120

Power off WFs manual processes.

w_stop/start_all.

Bayside stages don't responds to the manula commands.


PARALLEL ACTIVITY. CHECK THE SPARE SOULWFS MACHINE

spare SX WFS machine in parallel testing.


10:14 am

power on manual. Not responding the stages; all three.

Alessandro checking the electronics diagram and request a photo of the stages.

Trying to move manula the Z stage it moved.

The X and Y stages didn;t move.

11:00am

Chris went ot take a photo.

X stage was in midrange.

Move the rotation from 0 to 300.

The stages moved after the 24v ispower cycle.

move the X and Y stages outside the home positions, Power off

restart soft.

11:30am

Operate.

Bayside Stage Y didn't move. timeout of the process

X and Z moved/homed.

After a power cycle of 24v the stage moved.

Error in the power on : Operate -FATAL cube stage

Fabio; checks the power on sequence

11:40am

check operate.

Same error. Debug.

Fabio checks later on.

Poweron cube stage manual to finish the power on and debg the power on sequence later.

Revert the Power on sequence.

11:55am

operate.

Ocam

ok

cool down ocam

12:01PM

Doug calls mountain crew to get status of M1 works.

Still doing some work in M1.Ready soon. Light chamber still on

Load board setup: 2022_1201_RR

Update hexapod posiions to:

-1.5,0.9,-0.5,-254.5,-57.7,0

RFBG rotator=210

Adsec flat.
  • 623 modes applied.
12:17PM

lights off

CCD47 change binn for reset the camera and bcu47

ccd47 on

Acquire source

hexapod:

-0.9.1.3,-0.72,-238,-3.8,0

open loop

TN:194223

TN:20221205_195126

close loop

b1,600,3l/d,63

mag 5.9

gain:0.95,0.05,0.05

TN:20221205_200416

save/load flat:

SOUL-LUCI_600m_20221205

ose loop/open

to see the flat.

20220512 20:09:06.0870 Alert: IlluminationErr: OCAM
20220512 20:09:06.2574 Alert: IlluminationErr: OCAM
20220512 20:09:06.4286 Alert: IlluminationErr: OCAM

loaded flat:

/20220419_225059_bump_low_force.sav

close loop tt:0.65,0.05,0.05

TN:201423

01:20PM

Guido test a idl load flat limiting the max force applied.

AdOpt >print,fsm_load_shape('/local/aomeas/adsec_calib/flat/Applicable/default.sav',force_max=100)

1:10PM
  • Acquire INMAt
  • KL_29
  • 20220607_184919
  • iter=4
  • cl
  • M2C
  • KL_29
  • Rec:20221201_201809
  • gain:0.3,0.05,0.00
  • output:20221205_202823
Analyse InMat.

Looks very low disturbances. Bad IntMAt

New Inmat changing the TT gains to 0.85

Same parameters as before.

output:20221205_203552

Analyse

  • Improve signals, but still requires to have a better signal, increased the number of iterations.
  • output:Rec:Rec_20221205_203927
  • Bad intMat not used
New Intmat.
  • same params as before
  • iter=8
  • output:20221205_204246
Analyse

modes: