After discussion Fabio accepted his pull request for the adsec software with the if clause.
  • rested shell, powered off
  • adsc stop
  • On dxadsec, use_soul mixed_mode_control.20221028 && use_soul
  • This did not update the .bashrc for mixed_mode_control.20221028 but remained mixed_mode_control despite use_soul saying it was updated for mixed_mode_control.20221028
  • retry
  • no luck
  • Fabio created a backup of the .bashrc and hand edited the .bashrc for mixed_mode_control.20221028
  • success!
  • Adsc_start, power on, load program and set shell with 625 modes
Closing loop on ARGOS Source 11.2 mag
  • first attempt at NCPA failed. KL_v29 phase inverse not found.
  • 9:06:24 One singular PI shape while operating at 900Hz