Summary.

Night plan:

https://wiki.lbto.org/Commissioning/Com20220417

Last night we were actually able to accomplish way more than expected - collected data for field aberrations for SX and DX - a complete set for DX - we had a small break (1-1.5hr) in the thick clouds w/good seeing.

Start tonight is LUCI-Bino, we'll try to swtich to LBC a little past night center for somet extra-focal pupils and PHS - and depending on efficiency (or seeing) switch to MODS at the end of the night.

Weather predictions are mixed - poor seeing from ALTA, clear skies from everyone, but high winds 1st half from NOAA.

If we can't start with AO at the begiining or first half, then we'll start with AO as soon as we can open and then sort the plan from there.

-Barry

Details.

05:30 PM Power on

during power up, some process stopped and the guis disapeared

[AOeng@soul-sxwfs source]$ w_check

msgdrtdb running
mirrorctrl running and connected to Msgd
pinger running and connected to Msgd
powerboard running and connected to Msgd
slopecompctrl running and connected to Msgd
techviewctrl running and connected to Msgd
ocam2 running and connected to Msgd
ccd47 running and connected to Msgd
ttctrl running and connected to Msgd
adc1 running and connected to Msgd
adc2 running and connected to Msgd
rerotator running and connected to Msgd
filterwheel1 running and connected to Msgd
filterwheel2 NOT running
mercury NOT running
cubestage NOT running
baysidex NOT running
baysidey NOT running
baysidez NOT running
cameralensctrl NOT running
picctrl NOT running
optloopdiag NOT running
pupilcheckocam2 NOT running
masterdiagnostic NOT running
housekeeperWFS NOT running
dbfiller NOT running
wfsarb NOT running
od_gopt NOT running
gopt NOT running
auxloops NOT running
[AOeng@soul-sxwfs source]$

Stop/Start the processes.

stopProcessByName(): process msgdrtdb killed
Exception
Exception
WARNING: Supervisor process ocam2 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ocam2 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ocam2 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ocam2 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ccd47 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ccd47 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ccd47 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ccd47 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process truesense is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process truesense is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process truesense is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process wfseng is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process idl is still running by user (zabbix
), please kill it first.
WARNING: Supervisor process idl is still running by user (zabbix
), please kill it first.
WARNING: Supervisor process defunct is still running by user (AOeng
), please kill it first.

a second stop_all

stopProcessByName(): process msgdrtdb down
Exception
Exception
WARNING: Supervisor process ocam2 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process ccd47 is still running by user (AOeng
), please kill it first.
WARNING: Supervisor process idl is still running by user (zabbix
), please kill it first.
WARNING: Supervisor process idl is still running by user (zabbix
), please kill it first.
clean_ts8 script is disabled

top

top_powerup20220416.png

power on

5:51 PM ok

Ocam warm

6:37 PM cool down

Lookign at the viewer shows to the trinagle pattern.

Power cycle BCu2K + change 1x1crop.

7:00 PM left shutter not open

8:50 PM back to operations.

guider_L_20220417.png

Seeing variable 1.4"-1.5"

seeing 2"

04-17 04:03:51.777095| MAIN > thGetVar ttctrl.L.ENABLEOFFSET.REQ error(-3606): No connection with MsgD
WfsArbitrator_104656|DEB| 1|2022-04-17 04:07:47.534981| MAIN > Connected as pyarg

Rtestart process.

Power up without init of the bayside stages..?

make to have a speed init.

04-17 04:03:51.777095| MAIN > thGetVar ttctrl.L.ENABLEOFFSET.REQ error(-3606): No connection with MsgD
WfsArbitrator_104656|DEB| 1|2022-04-17 04:07:47.534981| MAIN > Connected as pyarg

techviewctrl.L.202204170000.log:FrameGrabberCtrl_191464|ERR| 141|2022-04-17 04:03:01.165174 MAIN > Error in TestComm(): [AOException] Timeout error in sendMultiBcuCommand() (code -5001) Timeout error File bcucommand.cpp line 164
techviewctrl.L.202204170000.log:FrameGrabberCtrl_191464|ERR| 142|2022-04-17 04:03:02.165293 MAIN > Lost communication with frame grabber!
techviewctrl.L.202204170000.log:FrameGrabberCtrl_191464|ERR| 145|2022-04-17 04:03:05.179979 MAIN > Error in TestComm(): [AOException] Timeout error in sendMultiBcuCommand() (code -5001) Timeout error File bcucommand.cpp line 164
techviewctrl.L.202204170000.log:FrameGrabberCtrl_191464|ERR| 148|2022-04-17 04:03:09.181215 MAIN > Error in TestComm(): [AOException] Timeout error in sendMultiBcuCommand() (code -5001) Timeout error File bcucommand.cpp line 164
techviewctrl.L.202204170000.log:FrameGrabberCtrl_191464|ERR| 151|2022-04-17 04:03:13.182395 MAIN > Error in TestComm(): [AOException] Timeout error in sendMultiBcuCommand() (code -5001) Timeout error File bcucommand.cpp line 164

change to seeing linited.

mean time Troubleshooting.

Reset BCU47 again.

Everything seems OK.

CCD47 takes images/darks.

seeing 1.5"

9:45 PM back to AO

=> Command line arguements processed
Set 0 Tracknum 20220417_050002 LuciName unknown
Set 1 Tracknum 20220417_050015 LuciName unknown
Set 2 Tracknum 20220417_050026 LuciName unknown
-> wrote file /home/AOeng/doug/get_telemetry/OSCO/20220417_050002.log
telescope to Everyone (10:03 PM)
2nd position telemetry w/ NCPA:

[telescope@obs1 ~]$ ./get_ao_telemetry

=> Command line arguements processed
Set 0 Tracknum 20220417_050248 LuciName unknown
Set 1 Tracknum 20220417_050300 LuciName unknown
Set 2 Tracknum 20220417_050311 LuciName unknown
-> wrote file /home/AOeng/doug/get_telemetry/OSCO/20220417_050248.log

seeing :1.4"


TN:_045711,045732.

ramping gain down to reduce the skip frames.

seeing up to 1.8", syart to skip frames 30%

reduce HO2 one step 01.13 to 0.09

reduced the skip but image quality suffered.

PSF not good..SR ~30. Modal SR:55%

10:12 PM Bin2 ESM

seeing 1.2"

seeing up to 1.7"

rot angle 313 CL good. EL 46

rms :0.6

10:20 rot angle 373.2 EL 46

OK CL

05:22:13.173206| MAIN > source_acq.acquire(): star found at X=627.78,Y=277.46
WfsArbitrator_154605|INF| 633|2022-04-17 05:22:13.173264| MAIN > source_acq.acquire(): current error X=1.092 mm, Y=1.785 mm
WfsArbitrator_154605|INF| 634|2022-04-17 05:22:21.750511| MAIN > Star found at x=31.2, y=31.1, flux=8449.2
WfsArbitrator_154605|INF| 635|2022-04-17 05:22:21.751834| MAIN > source_acq.acquire(): star found at X=499.21,Y=496.97
WfsArbitrator_154605|INF| 636|2022-04-17 05:22:21.751897| MAIN > source_acq.acquire(): current error X=0.088 mm, Y=0.071 mm
WfsArbitrator_154605|INF| 637|2022-04-17 05:22:21.751933| MAIN > source_acq.acquire(): target reached.

new preset, same magnitude, seeing cond more less the same.

Loaded the previous gains.

rms:0.6"

10:25 stop preset for wind in the dome.

10:29 Bin4

seeing: 1.7"

estm mag: 15.72

gain: 0.39,0.58,0.10

esm_bin4_20220417.png

rms:0.53

seeing 1:9"-2"

offset 3mm stage ~5 arcsec.

good image correction ~4.8,

7 05:42:44.412860| MAIN > arbFunc offsetXYinternal (, 'FLAOWFS', 'LUCIFER', -179.75, -185.85499999999999, True)
WfsArbitrator_154605|INF| 889|2022-04-17 05:42:44.413123| MAIN > offsetXY(): starting offset of -178.840, -186.731 mm
WfsArbitrator_154605|INF| 890|2022-04-17 05:42:44.413901| MAIN > offsetXYZ(): Focus correction (absolute): 0.000 mm
WfsArbitrator_154605|ERR| 891|2022-04-17 05:42:44.414263| MAIN > RETRY: Offset outside FoV limits. Target xyz is -193.78 -237.40 53.25

22-04-17 05:44:18.990914| MAIN > arbFunc offsetXYinternal (, 'FLAOWFS', 'LUCIFER', 0.00096476699999999995, -8.9732900000000004, True)
WfsArbitrator_154605|INF| 893|2022-04-17 05:44:18.991169| MAIN > offsetXY(): starting offset of 0.045, -8.973 mm

10:46

seeing 2"

gain:0.54,0.58,0.10

10:52"seeing 2.2-1.6""

variable seeing.

10:54 binocular.

left side ready, right side still collimating, the WFS is ready to proceed.

go with the soulsX.

Star faint in the CCD47.

center pupil doesn't get anything.

change from bin4 to bin1. ocam reconfigured but it is not getting anything...life but not seeing anything.

11:10- azimuth drive error...

very faint in the CCD47.

change magnitude to 14.1

11:14 send preset. Error Az driver error.

11:20 preset

adjust pointing.

variable seeing buble of large seeing >2"

loaded the previous gain

seems stable

optimize gain.

rms:0.49

gain:1.59,0.64,010

skipping

reducing gain

1.34,0.59,0.10

_063410

some skipping frames

_63517

skipping

063544

pause skipping frames.

7 06:36:04.777619| MAIN > Skip frame detected!
20221704 06:36:23.7978 Alert: HighOrderOffload: HighOrderOffload vecto

loopPause_skipactivated_20220417_1137.png

Check the opffsety in the AOS was -12 and 12

the WFS moved -11 and 11

offset moved. AOs failed.

skipping frames in pause.

send back offsetys to 0,0 no star in the CCD47 or PyWFS.

Cancel preset.

11:50 seeing >2" blow up.

center pupil failed.

close loop straigh.. seeing very bad...

loop close ..

loaded gain:0.39,0.58

loop not stable.

11:58 cancel preset.

really bad seeing.

change to seeing limited back up programs.

12:45 seeing variable still seeing limited.

guider_20220417_1247.png

1:33am variable seeing 1.4"-2.4"

gcs_20220417_134.png

seeing plot...

seeing_20220417.png

09:09:07.829360(UT) Az drive fault ...troubleshooting..

3:06 variable seeing 1.3-2"

gcs_20220417_306.png

variable seeing not good for AO.

4:00 am Power off

-- JuanCarlosGuerra - 17 Apr 2022
Topic revision: r3 - 17 Apr 2022, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.