Bin3 test preset

OK.

Loaded disturbance.

Sat Mar 5 17:59:36.021 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sat Mar 5 17:59:36.043 2022 aos.info - WFS source set to: FLAOWFS
Sat Mar 5 17:59:36.100 2022 aos.info - FLAO: WARNING: added atmospheric

angle 341.

optimize gain.

4.39,3.41,0.10

gopt

0.46,0.36,0.03

skip frames1-4%

TN:20220305_180829,_180942,_181135

relationship, drop SR vs skip frames %

with ncpa skip to 15%

_181337,_181507

peak high the mode:#20 (?)

ncpa off.

skip frames up to 20%

_181701

skiiping

set gain zero:_181949

optimize gain

still skipping.5-15%

4.52,3.83,0.10

gopt: 0.47,0.40,0.04

_182350,_182532

skipping frame0-20%

stopAO

Sat Mar 5 18:27:35.936 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[7] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sat Mar 5 18:27:35.952 2022 aos.info - WFS source set to: FLAOWFS
Sat Mar 5 18:27:36.012 2022 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s2.0_L040.0_v15.0_ovfreq200.0000_sd3892.fits
Sat Mar 5 18:27:36.711 2022 aos.info - FLAO: WFS: loading setup for LUCIFE

rotangle rerot:180.1

skip frame up to 7%

better than before.

_183755,183909

ncpa on

_184033,_184139,

rerotator on

skipframe up to 10%

_184330,_184436

gain0.1

move LFBG to 360

optimize with ncpa on

4.39,3.73,0.10

optimize_ncpa_on.png

ncpa off

optimize

gopt

0.54,0.45,0.04

gopt_360.png

_185202,_185330

ncpaon

185459,

skip frame 100% (?)

stop loop.

reflat

Sat Mar 5 18:59:16.629 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[12] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sat Mar 5 18:59:16.651 2022 aos.info - WFS source set to: FLAOWFS
Sat Mar 5 18:59:16.712 2022 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s2.0_L040.0_v15.0_ovfreq200.0000_sd3892.fits
Sat Mar 5 18:59:17.422 2022 aos.info - FLAO: WFS: loading setup for LUCIFE

rot 300

optimize gain.

no skip frames.

_190601, gopt signal visible.,_190755

gopt-rot300.png

ncpaon

_190953,_191113

start to skipTT jitter.

_76351|WAR| 3591|2022-03-05 19:10:54.207420| MAIN > Skip frame detected! (1.48% of frames skipped.)
AdSecArbitrator_76351|WAR| 3593|2022-03-05 19:10:57.270317| MAIN > Skip frame detected! (3.32% of frames skipped.)
AdSecArbitrator_76351|WAR| 3594|2022-03-05 19:10:58.277639| MAIN > Skip frame detected! (1.48% of frames skipped.)
AdSecArbitrator_76351|WAR| 3595|2022-03-05 19:10:59.285141| MAIN > Skip frame detected! (1.11% of frames skipped.)
AdSecArbitrator_76351|WAR| 3596|2022-03-05 19:11:00.292501| MAIN > Skip frame detected! (0.37% of frames skipped.)
AdSecArbitrator_76351|WAR| 3597|2022-03-05 19:11:01.299762| MAIN > Skip frame detected! (3.32% of frames skipped.)
AdSecArbitrator_76351|WAR| 3598|2022-03-05 19:11:02.314635| MAIN > Skip frame detected! (1.11% of frames skipped.)
AdSecArbitrator_76351|WAR| 3599|2022-03-05 19:11:03.325618| MAIN > Skip frame detected! (2.21% of frames skipped.)

change shutter to 48, it seems to bright at rot 322 (?) better PSF.

mag.15.2 (?)

rot 270

ar 5 19:22:30.547 2022 aos.info - TTF offload: previous 2 requests ignored
Sat Mar 5 19:22:30.560 2022 aos.info - TTF Offload request (FLAO) Z: -0.657313 -1.15525 -0.000158682
Sat Mar 5 19:22:33.217 2022 aos.info - TTF offload command end
Sat Mar 5 19:22:33.551 2022 aos.info - TTF offload: previous 2 requests ignored
Sat Mar 5 19:22:33.564 2022 aos.info - TTF Offload request (FLAO) Z: -0.259592 -0.451364 -0.000135515
Sat Mar 5 19:22:36.202 2022 aos.info - TTF offload command end
Sat Mar 5 19:22:37.581 2022 aos.info - TTF offload: previous 2 requests ignored
Sat Mar 5 19:22:37.604 2022 aos.info - TTF Offload request (FLAO) Z: 0.0753803 -0.16584 -0.0002859

skip frames 1-3.3%

optimize

7.94,3.34,0.10

gopt_rot270.png

192640,

skip frames to 100% (?) maybe catched with the TN_192804

lost PSF.

Sat Mar 5 19:28:19.350 2022 aos.info - TTF Offload request (FLAO) Z: -0.12832 0.0750563 5.25219e-06
Sat Mar 5 19:28:22.567 2022 aos.info - TTF offload command end
Sat Mar 5 19:28:23.353 2022 aos.info - TTF offload: previous 3 requests ignored
Sat Mar 5 19:28:23.374 2022 aos.info - TTF Offload request (FLAO) Z: -0.0567563 -0.146671 0.000104784
Sat Mar 5 19:28:26.585 2022 aos.info - TTF offload command end
Sat Mar 5 19:28:27.356 2022 aos.info - TTF offload: previous 3 requests ignored
Sat Mar 5 19:28:27.376 2022 aos.info - TTF Offload request (FLAO) Z: -0.147133 -0.239708 -6.78478e-06
Sat Mar 5 19:28:30.079 2022 aos.info - TTF offload command end
Sat Mar 5 19:28:30.372 2022 aos.info - TTF offload: previous 2 requests ignored
Sat Mar 5 19:28:30.387 2022 aos.info - TTF Offload request (FLAO) Z: -0.147133 -0.239708 -6.78478e-06
Sat Mar 5 19:28:33.075 2022 aos.info - TTF offload command end
Sat Mar 5 19:28:33.375 2022 aos.info - TTF offload: previous 2 requests ignored

Sat Mar 5 19:32:13.570 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[19] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sat Mar 5 19:32:13.584 2022 aos.info - WFS source set to: FLAOWFS
Sat Mar 5 19:32:13.639 2022 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s2.0_L040.0_v15.0_ovfreq200.0000_sd3892.fits
Sat Mar 5 19:32:14.317 2022 aos.info - FLAO: WFS: loading setup for LUCIFER

rot 245.

optimize

skip frames 1-3.3%

gopt

0.48,0.33,0.03

optimize_rot245.png

193904,_194028,

ncpon

_194225,_194347

stop

Sat Mar 5 19:46:28.176 2022 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[23] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Sat Mar 5 19:46:28.188 2022 aos.info - WFS source set to: FLAOWFS
Sat Mar 5 19:46:28.253 2022 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s2.0_L040.0_v15.0_ovfreq200.0000_sd3892.fits
Sat Mar 5 19:46:28.962 2022 aos.info - FLAO: WFS: loading setup for LUCIFER

_195136,_195250

skip frame to 20% during the tn__195250

at Mar 5 19:52:04.135 2022 aos.info - TTF offload command end
Sat Mar 5 19:52:14.095 2022 aos.info - TTF offload: previous 2 requests ignored
Sat Mar 5 19:52:14.110 2022 aos.info - TTF Offload request (FLAO) Z: 0.0352441 0.0275048 -4.67427e-06
Sat Mar 5 19:52:16.648 2022 aos.info - TTF offload command end
Sat Mar 5 19:52:17.099 2022 aos.info - TTF Offload request (FLAO) Z: 0.0301673 0.0280479 -5.58931e-05
Sat Mar 5 19:52:19.660 2022 aos.info - TTF offload command end

it is niot offloading anymore (?)


Sat Mar 5 19:54:28.668 2022 aos.Stop{FLAO}.complete - Stop{FLAO}[26]

handover 1PM

-- JuanCarlosGuerra - 05 Mar 2022
I Attachment Action Size Date Who Comment
gopt-rot300.pngpng gopt-rot300.png manage 144 K 05 Mar 2022 - 19:07 JuanCarlosGuerra  
gopt_360.pngpng gopt_360.png manage 150 K 05 Mar 2022 - 18:53 JuanCarlosGuerra  
gopt_rot270.pngpng gopt_rot270.png manage 143 K 05 Mar 2022 - 19:28 JuanCarlosGuerra  
optimize_ncpa_on.pngpng optimize_ncpa_on.png manage 217 K 05 Mar 2022 - 18:50 JuanCarlosGuerra  
optimize_rot245.pngpng optimize_rot245.png manage 182 K 05 Mar 2022 - 19:38 JuanCarlosGuerra  
skip_100_192640.pngpng skip_100_192640.png manage 415 K 05 Mar 2022 - 19:35 JuanCarlosGuerra  
Topic revision: r1 - 05 Mar 2022, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback