DE Time

07:36
  • Waiting for handover.
08:03
  • Heading to the OSCO star.
08:24
  • OSCOing. The Seeing is not great, but......
  • GOPT, No NCPA
  • Still a few safe skips.
  • 20220208_082936
    20220208_082955
    20220208_083010
    20220208_083023
    20220208_083037
  • Edge trying to roll.
  • 20220208_083122
  • 20220208_083239
  • NCPA On
  • 20220208_083532
    20220208_083556
    20220208_083611
    20220208_083631
    20220208_083654
    20220208_083707
    20220208_083728

08:48
  • Seeing is trending up. Difficult to keep loop closed.
08:55
  • Long slew to a new bin1 target.
09:26
  • Had trouble getting it to close. There was a seeing bubble that made the "Check Flux" step read mag 14, so we went to bin2
  • I had to go manual on the gain: 9.95, 2.7, 2.0. Those were about the low points on the optimize gain graphs. The stupid script went much higher and the loop opened.
  • 20220208_092927
    20220208_092941
    20220208_092957
    20220208_093008
    20220208_093018
  • Showing mag of 12.5, now. Bin1 should be possible, but we'll go to another target.
09:40
  • Moved to a new target. It's showing mag 10.7.
  • Psychotic auto-gain procedure loves to drive us to 60%+ safe skips
  • Seeing bubble and the edge rolled, 09:43.
  • 20220208_094847
    20220208_094903
    20220208_094918
  • AdSecArbitrator_140457|WAR| 2687|2022-02-08 09:46:04.997275| MAIN > Skip frame detected!
    20220802 09:48:04.6080 Alert: HighOrderOffload: HighOrderOffload vector ready
    20220802 09:48:29.6606 Alert: HighOrderOffload: HighOrderOffload vector ready
    AdSecArbitrator_140457|INF| 2737|2022-02-08 09:48:48.401692| COMMANDHANDLER > FSM (status: AORunning) has received command 2024 (Save status)
    Last message repeated
    AdSecArbitrator_140457|INF| 2738|2022-02-08 09:48:48.495314| COMMANDHANDLER > Command Save status (code 2024) successfully completed
    20220802 09:48:49.7078 Alert: HighOrderOffload: HighOrderOffload vector ready
    AdSecArbitrator_140457|WAR| 2739|2022-02-08 09:49:00.044255| MAIN > Skip frame detected!
    AdSecArbitrator_140457|INF| 2740|2022-02-08 09:49:03.144783| COMMANDHANDLER > FSM (status: AORunning) has received command 2024 (Save status)
    AdSecArbitrator_140457|INF| 2741|2022-02-08 09:49:03.241681| COMMANDHANDLER > Command Save status (code 2024) successfully completed
    AdSecArbitrator_140457|WAR| 2742|2022-02-08 09:49:04.139191| MAIN > Skip frame detected!
    20220802 09:49:09.7709 Alert: HighOrderOffload: HighOrderOffload vector ready
    AdSecArbitrator_140457|INF| 2748|2022-02-08 09:49:18.315983| COMMANDHANDLER > FSM (status: AORunning) has received command 2024 (Save status)
    Last message repeated
    AdSecArbitrator_140457|INF| 2749|2022-02-08 09:49:18.396772| COMMANDHANDLER > Command Save status (code 2024) successfully completed
    AdSecArbitrator_140457|ERR| 2750|2022-02-08 09:49:18.967695| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2751|2022-02-08 09:49:18.967740| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:18.9751 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|INF| 2752|2022-02-08 09:49:19.025891| ALERTHANDLER > Received from masterdiagnostic.L a DumpSaved alert with filename /local/aolog/2022/02/08/ADAPTIVE-SECONDARY_00001644313758_20220208_094918.bin
    AdSecArbitrator_140457|INF| 2753|2022-02-08 09:49:19.025913| ALERTHANDLER > Processdump is skipped
    20220802 09:49:19.0264 Alert: DumpSaved: DumpSaved vector ready
    AdSecArbitrator_140457|ERR| 2754|2022-02-08 09:49:23.022751| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2755|2022-02-08 09:49:23.022799| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:23.0233 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2756|2022-02-08 09:49:27.066105| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2757|2022-02-08 09:49:27.066162| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:27.0666 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2758|2022-02-08 09:49:31.115137| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2759|2022-02-08 09:49:31.115183| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:31.1157 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2760|2022-02-08 09:49:35.158339| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2761|2022-02-08 09:49:35.158413| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:35.1589 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2762|2022-02-08 09:49:39.203571| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2763|2022-02-08 09:49:39.203617| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:39.2041 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2764|2022-02-08 09:49:43.247568| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2765|2022-02-08 09:49:43.247614| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:43.2481 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2766|2022-02-08 09:49:47.290751| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2767|2022-02-08 09:49:47.290799| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:47.2913 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2768|2022-02-08 09:49:51.333709| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2769|2022-02-08 09:49:51.333762| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:51.3342 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2770|2022-02-08 09:49:55.074591| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2771|2022-02-08 09:49:55.074641| ALERTHANDLER > Alert message: COILS DISABLED from housekeeper.L
    20220802 09:49:55.0752 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|ERR| 2772|2022-02-08 09:49:55.386709| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
    AdSecArbitrator_140457|ERR| 2773|2022-02-08 09:49:55.386755| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
    20220802 09:49:55.3872 Alert: Error: COILS DISABLED
    AdSecArbitrator_140457|WAR| 2774|2022-02-08 09:49:55.927712| MAIN > Skip frame detected!
    AdSecArbitrator_140457|INF| 2775|2022-02-08 09:49:56.407574| ALERTHANDLER > Received from masterdiagnostic.L a DumpSaved alert with filename /local/aolog/2022/02/08/ADAPTIVE-SECONDARY_00001644313796_20220208_094956.bin
    AdSecArbitrator_140457|INF| 2776|2022-02-08 09:49:56.407596| ALERTHANDLER > Processdump is skipped
    20220802 09:49:56.4165 Alert: DumpSaved: DumpSaved
10:08
  • Same target.
  • 20220208_100805
    20220208_100819
    20220208_100835
    20220208_100849
    20220208_100902
    20220208_100921
  • Loop opened while I wasn't looking. I think that the gains might have been to high, because the seeing improved dramatically, but I re-applied the saved gain.
  • 20220208_101255
    20220208_101323
    20220208_101344
    20220208_101413
    20220208_101428
  • Still trying to roll the edge some times.
  • OOPS! No NCPA!
  • 20220208_101650
    20220208_101711
    20220208_101725
    20220208_101738
    20220208_101757
    20220208_101811
10:20
  • We had an offset failure. It looks to have been on the TCS side. Looks like the star was on the edge of the patrol field, then we did an offset and it went to TRACK mode.
10:27
  • Random actuator jumping, I've excluded those that are in the "Do Not Use" Lists
  • fastdiagn_100114 |WAR| 36778|2022-02-08 10:27:03.694378| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0667 -0.662558
    fastdiagn_100114 |WAR| 36779|2022-02-08 10:27:03.697270| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0667 -0.655537
    fastdiagn_100114 |WAR| 36780|2022-02-08 10:27:03.700086| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0667 -0.679493
    fastdiagn_100114 |WAR| 36781|2022-02-08 10:27:03.702973| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0667 -0.675703
    fastdiagn_100114 |WAR| 36782|2022-02-08 10:27:03.705891| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0667 -0.613031
    fastdiagn_100114 |WAR| 36783|2022-02-08 10:27:03.919853| FUNCTWARNING > FunctWarning CHFFPURECURRENT-0621 -0.605078
10:30
  • Going to a new, fainter Bin1 target, now that the seeing is stable.
  • 20220208_104249
    20220208_105345
  • The last one should have ccd47 frames.
  • It was a close double and that was fouling us up.
10:54
  • RequestAOStatus fstat=self._fsm.getState().getName().split('.') File "/home/aoeng/live/aoroot/lib/python/statemap.py", line 104, in getState raise StateUndefinedException StateUndefinedException
  • Sent a new preset while the old preset was still running, or something.^^^^^^
  • This cascaded to some weird errors on the telescope side.
  • David G. is fixing it.
  • Heading to a new object.
  • Headin' on to Eden, yea, brother!
  • Need more chocolate, brain slipping into Space Hippie mode (If that's not a thing, it is now).
11:05
  • We had the issue where the pupil finding routine needed to be helped by moving the camera lens.
  • It's 12.8. I skipped check flux, just to give us a super tough bin1 test, or something like that.
  • housekeeper.L |INF| 48032|2022-02-08 11:13:56.245689| SwitchSafeSkipFrameCnt > 40405
    housekeeper.L |INF| 48068|2022-02-08 11:13:58.443254| SwitchSafeSkipFrameCnt > 40425
    housekeeper.L |INF| 48080|2022-02-08 11:13:59.523979| SwitchSafeSkipFrameCnt > 40433
    housekeeper.L |INF| 48088|2022-02-08 11:14:00.592490| SwitchSafeSkipFrameCnt > 40438
    housekeeper.L |INF| 48096|2022-02-08 11:14:01.668946| SwitchSafeSkipFrameCnt > 40445
  • That level of safe skipping isn't showing up in the AdSec Control GUI.
  • 20220208_111541
    20220208_111555
    20220208_111608
    20220208_111622
    20220208_111640
  • 20220208_111832
  • Several Safe Skips.
11:22
  • Going to go and take some nice images.
  • Guide star was showing up as mag 14, going to a different one.
  • New guide star showing 13.1
  • 20220208_114640
    20220208_114653
    20220208_114704
    20220208_114720
    20220208_114731
11:51
  • Going to try Bin3.
12:00
  • Optimize gain worked without problems.
  • Shows 16.4
  • 20220208_120107
    20220208_120138
    20220208_120208
    20220208_120228
    20220208_120250
12:25
  • Going to try another bin3
  • Mag 15.8
  • 20220208_123602
    20220208_123615
    20220208_123629
    20220208_123641
    20220208_123655
13:00
  • Trying one last Bin1
  • Doing a pointing check.
  • Bin1
  • 11.7
  • NCPA On
  • There was a strange jump, hopefully, we capture it in one of these TNs
  • 20220208_131213
    20220208_131229
    20220208_131249
    20220208_131305
    20220208_131319

-- GregoryTaylor - 08 Feb 2022
Topic revision: r1 - 08 Feb 2022, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.