20220131_SX_Day. SX AdSec flattening with 4D

12:15 MST RR installed

12:55 Laser light on 4D
  • Full pupil seen
  • quite a few lines
  • Try to use hexalign to improve

1:10 Start Interferometer process and initialize IDL
  • InterferometerCtrl -i interferometerctrl &
  • idl
  • IDL> @startup-onlycom
  • IDL> print, idl_4d_init()
  • IDL> a=get_img4d(/show)
    • 4D phase map is displayed on AdSec machine
  • IDL> help, a
    • A FLOAT = Array[890, 890]
  • Communidation seems to be working properly

1:23 hexalign()
  • hex_1.jpg:
    hex_1.jpg
  • HXP MOVEMENT -0.58075925 1.9944825 0.056776701 -206.92143 -57.782219
  • HXP MOVEMENT 0.032709696 -0.065034414 0.0041135938 6.4184101 2.7392715
  • RMS 0.16*632.8 = 101.248 nm
  • flat_orig.jpg:
    flat_orig.jpg

2:10 Final Good mask 20220131_201913

2:11 Start taking reconstructor data

-- DougMiller - 31 Jan 2022
I Attachment Action Size Date Who Comment
align_1.jpgjpg align_1.jpg manage 18 K 31 Jan 2022 - 20:40 DougMiller  
flat_orig.jpgjpg flat_orig.jpg manage 216 K 31 Jan 2022 - 21:03 DougMiller  
hex_1.jpgjpg hex_1.jpg manage 18 K 31 Jan 2022 - 20:37 DougMiller  
Topic revision: r2 - 08 Feb 2022, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.