20220114_SX_Day_NCPA Check and calibration of LUCI1 NCPA Coefficients at one (frozen) rotator angle

12:53 MST Take over from Juan Carlos check of closed loop
  • JC and Gustavo working in parallel on ARGOS internal checks

12:56 Change symbolic link to LUCIFER NCPA coefficeints table
lrwxrwxrwx. 1 AOeng AOeng    21 Nov 19 23:07 LUCIFER -> ncpa_fit_20200830.dat
-rw-r--r--. 1 AOeng AOeng   960 Jan 14 19:30 ncpa_measured_20211124.dat
-rw-r--r--. 1 AOeng AOeng 29241 Jan 14 19:31 ncpa_fit_20211124.dat
[AOeng@soul-sxwfs ncpa]$ pwd
/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ncpa
[AOeng@soul-sxwfs ncpa]$ rm LUCIFER
[AOeng@soul-sxwfs ncpa]$ ln -s ncpa_fit_20211124.dat LUCIFER
[AOeng@soul-sxwfs ncpa]$ ls -l LUCIFER
lrwxrwxrwx. 1 AOeng AOeng 21 Jan 14 19:56 LUCIFER -> ncpa_fit_20211124.dat

1:02 LUCI1 instrument rotator at 317.6 degrees

1:39 Closed Loop
  • Opt Gain 1.96 0.23 0.22
  • GOPT 0.583
  • NCPA on -15.8 6.1 19.9 0.9 9.4 -9.4 1.9 -5.1
  • from table 317 -15.7 6.48 19.5 0.922 9.52 -9.35 2.22 -5.29

2:00 LUCI1 set in 128x128 subframe
  • Last image luci1.20220115.0010.fits
  • First frame H 0.567 sec 10 image cube luci1.20220115.0011.fits
  • Dark frame H 0.567 sec 10 image cube luci1.20220115.0012.fits

2:08 Started script for Z5,6
[AOeng@soul-sxwfs ncpa_collect]$ python ncpa_collect_frozen.py 

   => No command line arguments: using default values

   => NCPA NCPA Collect Parameters:
        Rotation Parmeters:  Angles 0 -> 0  steps of 30
        Zernike Data to Collect: Modes [5, 6], 2 steps (+-) of 30
        Test Mode: False
        Set Name: 20220114_56
        Log List: 20220114_56/list.log

   => Are these NCPA Collect parameters correct?
       Yes: press carriage return
       No: press any character and carriage return
   =>
   => Yes

Starting mode 5
-> Mode 5:  Amplitude 6.480000 -54.000000 67.000000 30.000000

-> Mode 5:  Amplitude -54
    Set  0   Tracknum  20220114_211113   LuciName  luci1.20220115.5013.fits
-> Mode 5:  Amplitude -24
    Set  0   Tracknum  20220114_211140   LuciName  luci1.20220115.5014.fits
-> Mode 5:  Amplitude 6
    Set  0   Tracknum  20220114_211203   LuciName  luci1.20220115.5015.fits
-> Mode 5:  Amplitude 36
    Set  0   Tracknum  20220114_211226   LuciName  luci1.20220115.5016.fits
-> Mode 5:  Amplitude 66
    Set  0   Tracknum  20220114_211248   LuciName  luci1.20220115.5017.fits
   -> wrote file 20220114_56/0/rot_0_mode_5.log

   -> Setting NCPA Coefficients to original values

Starting mode 6
-> Mode 6:  Amplitude 19.500000 -41.000000 80.000000 30.000000

-> Mode 6:  Amplitude -41
    Set  0   Tracknum  20220114_211311   LuciName  luci1.20220115.5018.fits
-> Mode 6:  Amplitude -11
    Set  0   Tracknum  20220114_211333   LuciName  luci1.20220115.5019.fits
-> Mode 6:  Amplitude 19
    Set  0   Tracknum  20220114_211355   LuciName  luci1.20220115.5020.fits
-> Mode 6:  Amplitude 49
    Set  0   Tracknum  20220114_211420   LuciName  luci1.20220115.5021.fits
-> Mode 6:  Amplitude 79
    Set  0   Tracknum  20220114_211442   LuciName  luci1.20220115.5022.fits
   -> wrote file 20220114_56/0/rot_0_mode_6.log

   -> Setting NCPA Coefficients to original values

List of log names written to 20220114_56/list.log

2:15 Start Z7,8,9,10
[AOeng@soul-sxwfs ncpa_collect]$ python ncpa_collect_frozen.py 

   => No command line arguments: using default values

   => NCPA NCPA Collect Parameters:
        Rotation Parmeters:  Angles 0 -> 0  steps of 30
        Zernike Data to Collect: Modes [7, 8, 9, 10], 2 steps (+-) of 20
        Test Mode: False
        Set Name: 20220114_891011
        Log List: 20220114_891011/list.log

   => Are these NCPA Collect parameters correct?
       Yes: press carriage return
       No: press any character and carriage return
   =>
   => Yes

Starting mode 7
-> Mode 7:  Amplitude 0.922000 -40.000000 41.000000 20.000000

-> Mode 7:  Amplitude -40
    Set  0   Tracknum  20220114_211802   LuciName  luci1.20220115.5023.fits
-> Mode 7:  Amplitude -20
    Set  0   Tracknum  20220114_211832   LuciName  luci1.20220115.5024.fits
-> Mode 7:  Amplitude 0
    Set  0   Tracknum  20220114_211854   LuciName  luci1.20220115.5025.fits
-> Mode 7:  Amplitude 20
    Set  0   Tracknum  20220114_211918   LuciName  luci1.20220115.5026.fits
-> Mode 7:  Amplitude 40
    Set  0   Tracknum  20220114_211940   LuciName  luci1.20220115.5027.fits
   -> wrote file 20220114_891011/0/rot_0_mode_7.log

   -> Setting NCPA Coefficients to original values

Starting mode 8
-> Mode 8:  Amplitude 9.520000 -31.000000 50.000000 20.000000

-> Mode 8:  Amplitude -31
    Set  0   Tracknum  20220114_212002   LuciName  luci1.20220115.5028.fits
-> Mode 8:  Amplitude -11
    Set  0   Tracknum  20220114_212027   LuciName  luci1.20220115.5029.fits
-> Mode 8:  Amplitude 9
    Set  0   Tracknum  20220114_212049   LuciName  luci1.20220115.5030.fits
-> Mode 8:  Amplitude 29
    Set  0   Tracknum  20220114_212112   LuciName  luci1.20220115.5031.fits
-> Mode 8:  Amplitude 49
    Set  0   Tracknum  20220114_212134   LuciName  luci1.20220115.5032.fits
   -> wrote file 20220114_891011/0/rot_0_mode_8.log

   -> Setting NCPA Coefficients to original values

Starting mode 9
-> Mode 9:  Amplitude -9.350000 -49.000000 32.000000 20.000000

-> Mode 9:  Amplitude -49
    Set  0   Tracknum  20220114_212156   LuciName  luci1.20220115.5033.fits
-> Mode 9:  Amplitude -29
    Set  0   Tracknum  20220114_212219   LuciName  luci1.20220115.5034.fits
-> Mode 9:  Amplitude -9
    Set  0   Tracknum  20220114_212242   LuciName  luci1.20220115.5035.fits
-> Mode 9:  Amplitude 11
    Set  0   Tracknum  20220114_212304   LuciName  luci1.20220115.5036.fits
-> Mode 9:  Amplitude 31
    Set  0   Tracknum  20220114_212328   LuciName  luci1.20220115.5037.fits
   -> wrote file 20220114_891011/0/rot_0_mode_9.log

   -> Setting NCPA Coefficients to original values

Starting mode 10
-> Mode 10:  Amplitude 2.220000 -38.000000 43.000000 20.000000

-> Mode 10:  Amplitude -38
    Set  0   Tracknum  20220114_212350   LuciName  luci1.20220115.5038.fits
-> Mode 10:  Amplitude -18
    Set  0   Tracknum  20220114_212413   LuciName  luci1.20220115.5039.fits
-> Mode 10:  Amplitude 2
    Set  0   Tracknum  20220114_212436   LuciName  luci1.20220115.5040.fits
-> Mode 10:  Amplitude 22
    Set  0   Tracknum  20220114_212459   LuciName  luci1.20220115.5041.fits
-> Mode 10:  Amplitude 42
    Set  0   Tracknum  20220114_212525   LuciName  luci1.20220115.5042.fits
   -> wrote file 20220114_891011/0/rot_0_mode_10.log

   -> Setting NCPA Coefficients to original values

List of log names written to 20220114_891011/list.log

2:28 Start Z4,11
  • 4 ok
  • 11 only first three amp then luci script stuck
  • Killed script then took dark from WFS Control GUI (with light on and CL) 0052
[AOeng@soul-sxwfs ncpa_collect]$ python ncpa_collect_frozen.py 

   => No command line arguments: using default values

   => NCPA NCPA Collect Parameters:
        Rotation Parmeters:  Angles 0 -> 0  steps of 30
        Zernike Data to Collect: Modes [4, 11], 2 steps (+-) of 5
        Test Mode: False
        Set Name: 20220114_411
        Log List: 20220114_411/list.log

   => Are these NCPA Collect parameters correct?
       Yes: press carriage return
       No: press any character and carriage return
   =>
   => Yes

Starting mode 4
-> Mode 4:  Amplitude -15.700000 -25.000000 -4.000000 5.000000

-> Mode 4:  Amplitude -25
    Set  0   Tracknum  20220114_213034   LuciName  luci1.20220115.5043.fits
-> Mode 4:  Amplitude -20
    Set  0   Tracknum  20220114_213102   LuciName  luci1.20220115.5044.fits
-> Mode 4:  Amplitude -15
    Set  0   Tracknum  20220114_213125   LuciName  luci1.20220115.5045.fits
-> Mode 4:  Amplitude -10
    Set  0   Tracknum  20220114_213147   LuciName  luci1.20220115.5046.fits
-> Mode 4:  Amplitude -5
    Set  0   Tracknum  20220114_213210   LuciName  luci1.20220115.5047.fits
   -> wrote file 20220114_411/0/rot_0_mode_4.log

   -> Setting NCPA Coefficients to original values

Starting mode 11
-> Mode 11:  Amplitude -5.290000 -15.000000 6.000000 5.000000

-> Mode 11:  Amplitude -15
    Set  0   Tracknum  20220114_213233   LuciName  luci1.20220115.5048.fits
-> Mode 11:  Amplitude -10
    Set  0   Tracknum  20220114_213255   LuciName  luci1.20220115.5049.fits
-> Mode 11:  Amplitude -5
    Set  0   Tracknum  20220114_213318   LuciName  luci1.20220115.5050.fits
-> Mode 11:  Amplitude 0
    Set  0   Tracknum     LuciName  unknown
-> Mode 11:  Amplitude 5
^C^C
^Z
[2]+  Stopped                 python ncpa_collect_frozen.py
[AOeng@soul-sxwfs ncpa_collect]$ jpbs
-bash: jpbs: command not found
[AOeng@soul-sxwfs ncpa_collect]$ jobs
[1]-  Running                 emacs ncpa_collect_frozen.py &
[2]+  Stopped                 python ncpa_collect_frozen.py
[AOeng@soul-sxwfs ncpa_collect]$ kill -9 %2

[2]+  Stopped                 python ncpa_collect_frozen.py
[AOeng@soul-sxwfs ncpa_collect]$ jobs
[1]-  Running                 emacs ncpa_collect_frozen.py &
[2]+  Killed                  python ncpa_collect_frozen.py

2:39 set steps for 4,11 to 10 nm
  • Wrong directory so killed job
  • Luci readout stuck.
  • Dark frames again. OK 0053

2:43 4,11 with steps of 10 again
  • still stuck
  • try New teminal
  • Nope, still not working.
  • Try stop start of WFSARB
  • Loop open, had to send new preset
  • All ok
  • Moved bayside stages to center
  • Collect dark as test. 0054

3:09 4,11 with steps of 10 again
  • ok data being collected
[AOeng@soul-sxwfs ncpa_collect]$ python ncpa_collect_frozen.py 

   => No command line arguments: using default values

   => NCPA NCPA Collect Parameters:
        Rotation Parmeters:  Angles 0 -> 0  steps of 30
        Zernike Data to Collect: Modes [4, 11], 2 steps (+-) of 10
        Test Mode: False
        Set Name: 20220114_411_10
        Log List: 20220114_411_10/list.log

   => Are these NCPA Collect parameters correct?
       Yes: press carriage return
       No: press any character and carriage return
   =>
   => Yes

Starting mode 4
-> Mode 4:  Amplitude -15.700000 -35.000000 6.000000 10.000000

-> Mode 4:  Amplitude -35
    Set  0   Tracknum  20220114_221014   LuciName  luci1.20220115.5055.fits
-> Mode 4:  Amplitude -25
    Set  0   Tracknum  20220114_221041   LuciName  luci1.20220115.5056.fits
-> Mode 4:  Amplitude -15
    Set  0   Tracknum  20220114_221103   LuciName  luci1.20220115.5057.fits
-> Mode 4:  Amplitude -5
    Set  0   Tracknum  20220114_221128   LuciName  luci1.20220115.5058.fits
-> Mode 4:  Amplitude 5
    Set  0   Tracknum  20220114_221150   LuciName  luci1.20220115.5059.fits
   -> wrote file 20220114_411_10/0/rot_0_mode_4.log

   -> Setting NCPA Coefficients to original values

Starting mode 11
-> Mode 11:  Amplitude -5.290000 -25.000000 16.000000 10.000000

-> Mode 11:  Amplitude -25
    Set  0   Tracknum  20220114_221213   LuciName  luci1.20220115.5060.fits
-> Mode 11:  Amplitude -15
    Set  0   Tracknum  20220114_221235   LuciName  luci1.20220115.5061.fits
-> Mode 11:  Amplitude -5
    Set  0   Tracknum  20220114_221258   LuciName  luci1.20220115.5062.fits
-> Mode 11:  Amplitude 5
    Set  0   Tracknum  20220114_221322   LuciName  luci1.20220115.5063.fits
-> Mode 11:  Amplitude 15
    Set  0   Tracknum  20220114_221344   LuciName  luci1.20220115.5064.fits
   -> wrote file 20220114_411_10/0/rot_0_mode_11.log

   -> Setting NCPA Coefficients to original values

List of log names written to 20220114_411_10/list.log

3:15 another REAL LUCI dark 0065

3:16 Done

-- DougMiller - 14 Jan 2022