Staff.

Summary

Details.

Mon Nov 29 16:19:16.740 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[487] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

THERE IS A SHIFT TOWARD THE NEGAIVE X of about 2 pixels.

bin1_pupilshift_20211129.png

>>> makePupils.movePupils('20211125_004613',2,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_164335/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_164335/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_164335
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_164335"
'20211129_164335'
>>>

loaded pupil '20211129_164335'.

from 0 to 360

>>> makePupils.movePupils('20211125_004613',-2,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_171656/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_171656/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_171656
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_171656"
'20211129_171656'

loaded 20211129_171656'

'20211129_171656'
>>> makePupils.movePupils('20211125_004613',-4,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173441/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173441/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173441
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173441"
'20211129_173441'
>>>

too far.....

>>> makePupils.movePupils('20211125_004613',-3,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718"
'20211129_173718'
>>>


loading different pupils looks like there are not displacements in the pupil check.......

making apupil from the origin used for the one used in night time and it worked..

>>> makePupils.movePupils('20211125_004613',-3,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_173718"
'20211129_173718'
>>>

loaded 20211129_173718.

close loop cl x=-43..

requiires to move positive in x

build another pupil

>>> makePupils.movePupils('20211117_180043',-3,1,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_181459/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_181459/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_181459
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_181459"
'20211129_181459'

update AO table with the pupil:20211129_181459'

generate a link in the AO table.

table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO.txt.20211129

Mon Nov 29 18:23:57.072 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[489] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupil check in the range of x->-53

>>> makePupils.movePupils('20211117_180043',2,1,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_182826/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_182826/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_182826
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_182826"
'20211129_182826

center pupils -> x=-52

check with a large shift..

>>> makePupils.movePupils('20211117_180043',5,1,240,2)
makePupils_fromMasks(): moving from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_183515/pupdata.txt
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_183515/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_183515
/bin/mv -f "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/live/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211129_183515"
'20211129_183515'
>>>

the pupil cehck runs in a different environment.

Check with Brandon...

AOeng@soul-sxwfs source]$ which_w_procs.sh | gre
p latest
14633,OptLoopDiagnost,AOeng ,OptLoopDiagnostic
-i od_gopt,/home/AOeng/aoroot.latest/bin
71807,python ,AOeng ,python pupilcheck
.py -i pupilcheckocam2 ocam2,/home/AOeng/aoroot.l
atest/bin

  • cancel preset
  • w_stop_all.

Mon Nov 29 19:07:16.653 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[491] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

the pupil check keeps running with the soul.latest.

This could bring the error in the pupil check..

Brandon is checking.

back to the working AO table.

table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO.txt.20211127

w_stop_all

w_start

to avoid. the keep up process

nModes:501 r0:nan snMode:0 strehl:0 ttMod:0
Mon Nov 29 19:20:59.628 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[493] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

Mon Nov 29 19:23:02.549 2021 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[494] - [AOException] presetAO: WARNING - RETRY: No settings for instrument:LUCIFER - mode:ACE-AO (-20004) WFSARB_ARG_ERROR

w_stop_all

The w_start shows to have an issue with the msgdbrtd.. The elevation is reporting an error>

20 minutes pause.

12:40PM

Mon Nov 29 20:11:06.327 2021 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[495] - [AOException] presetAO: WARNING - RETRY: No settings for instrument:LUCIFER - mode:ACE-AO (-20004) WFSARB_ARG_ERROR
  • restart AOS.
  • set flat->638
  • Brandon.looking at the pupil check.
  • pupil check still running with the Soul.latest env.

back to pause_resume...to test if the system is consistent..

AOeng@soul-sxwfs.mountain.lbto.org's password:
Last login: Mon Nov 29 20:28:04 2021 from obs4.mountain.lbto.org
ADOPT_ROOT=/home/AOeng/pause_resume_test_20211011/aoroot
ADOPT_SOURCE=/home/AOeng/pause_resume_test_20211011/ao-supervisor
[AOeng@soul-sxwfs ao-supervisor]$
[AOeng@soul-sxwfs ao-supervisor]$
[AOeng@soul-sxwfs ao-supervisor]$

Mon Nov 29 20:31:56.837 2021 aos.info - WFS source set to: FLAOWFS
Mon Nov 29 20:32:06.863 2021 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[2] - [AOException] presetAO: WARNING - RETRY: No settings for instrument:LUCIFER - mode:ACE-AO (-20004) WFSARB_ARG_ERROR

msgdaemon restarted....

issues with the message deamon and the adsec ...

message_daemon.png

adsc_stop/start

1:49 PM

The Adsec is looking at the spare-soul machine and it can not connect to the soul mountain machine.

Gustavo is talking with Fabio to shutdown the spare in a safe mode...

spare_conf..png

spare wfs arbitrator stop.

adsc_stop/start seems still connected to the spare.

spare--warm up ocam.

startup WFS

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Nov 29 21:39:15.840 2021 aos.info - WFS source set to: FLAOWFS
Mon Nov 29 21:39:16.754 2021 aos.info - FLAO: WFS: loading setup for LUCIFER

wobble bin 1 good.

bin1_nov29.png

starting bin2 from zero

bin2_check20211129.png

3:02

Handover.

3:20PM

-- JuanCarlosGuerra - 29 Nov 2021