Details.

Bin4

Sat Nov 27 15:32:13.762 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[110] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

w_stop_all

Sat Nov 27 15:45:05.754 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[111] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

mag17

shutter 48

rot->179.749

interesting camera lens at the limit.

loaded pupil 20211124_171903 his has one pixel x more to the center.

unstable close loop.

re-home pup rotator.

cancel preset.

unstable loop.

Sat Nov 27 16:04:56.230 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[113] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupil 20211117_191303 calibrated for rotation range.

cancel preset. failure to load the reconstructor.

check the AO table-> rec and filtering fine.

restart the adsec arbitrator.

preset

adsec failure.

rec not loaded.

adsc_sstop/start

bin4 pupil rotation.

bin4_20211127_pupil_20211119_191303.png

loaded pupil 20211124_171903

shift x=0, y=2 from 202190616_173857

bin4_20211124_171903.png

11011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20190916_173857/pup1.fits'
>>> makePupils.movePupils('20200916_173857',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_174326/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_174326/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_174326
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_174326"
'20211127_174326'
>>>

loaded '20211127_174326.

bin4_pupil20211127_174326.png

stop at 0

reload the flat.

>>> makePupils.movePupils('20200916_173857',-1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_175502/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_175502/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_175502
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_175502"
'20211127_175502'

loaded '20211127_175502.

the pup check red circle shows to be in the upper left corner.

process pupil check stop/start and it shows the same

the real time pupils not shown.

CAncel preset.

Sat Nov 27 18:15:53.072 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[120] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

cancel preset.

The live pupils not shown.

Change to bin1.

Sat Nov 27 18:19:51.757 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[122] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

display the live pu[il circles.

preset mag 17

it still not displaying the real time pupils..

w_stop/start_all

Sat Nov 27 18:34:56.007 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[124] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

not shown the live pupils...

Bin1

Change to bin1 performnaces..close loop at different angles: 0,190,
  • rot angle 0
  • TN:20211127_184825
  • ok
bin1_20211127_rot0.png
  • taken luci dark
  • rot 190
  • rerot 255.592
  • optimize gain

  • fringies after the optimize gain...
  • stop rerot tracking
  • delta movements of the rerot->there is an offset in the rerotator.
  • rot 255.092 no fringes
  • homing.
  • activate tracking.
  • rot->255.649
  • close loop 10modi
  • increase the HO1
  • no firnges.
  • optimize gain
  • no fringes at the end of the process.
  • TN:20211127_190758

psf_rot190_bin1.png
  • open loop
  • rot 341
  • rerot->180.262
  • optimize gain
  • tn:20211127_191801
  • up firng: rot 180.862
  • low:176
  • there is about 4.7 degrees of travel.=>2.35 offset
lower_rot.png upper_rot.png
  • apply Doug offset.
RotatorOffsetBin1 float 349.6 # Should 350.75 to match with REC 20181215_201757. DLM Adjusted 21211122 after changing homing speed

Sat Nov 27 19:29:38.247 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[131] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • rerot->179.10
  • optimize gain
  • TN:20211127_193418

shift to bin4

  • apply Doug offset.
RotatorOffsetBin4 float 348.75 #DLM Adjusted 21211122 after changing homing speed

Sat Nov 27 19:39:09.567 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[133] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupil 20211117_192657 default in the AO table.

j band

tn:20211127_194615

distort PSF

open loop

loaded pup 20200916_173857

error:

difx=1.38

diff y=0.2

cl x=100, y=-100

tn:20211127_195117

bad psf.

stop

>>> makePupils.movePupils('20200916_173857',0,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_195647/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_195647/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_195647
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_195647"
'20211127_195647'
>>>

keep the X and move in Y by two pixels.

dif x=-0.21, diffy=0.04

cl x=9.36 y=-97.82

tn:20211127_195954

luci.0060

change AO table with 20211127_195647

Sat Nov 27 20:04:44.647 2021 aos.PresetAO{FLAO}.complete - PresetAO {FLAO}[137] - dx:nan dy:nan slNull: f1:0.0 f2:1.0 freq:312 mag:0 nBins:4 nModes:54 r0:nan snMode:0 strehl:0 ttMod:0

bin4-pupilcenering_20211127_195647.png

TN:20211127_200800

lngain

manualgain

20211127_201130

stop rerot

tn:20211127_201718

luci.0090

added the pupdata from last pupil in the directory 20200916_173857

at Nov 27 20:22:13.635 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[139] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupilcentering...modify.png

close loop

lngain

looks better pupils

20211127_202540

better PSF

202710

optimize gain

optimizegain_modified....png

  • rerot ->179.450
  • loaded flat 20211123
  • close loop
  • 203851
  • manula gain 0.55,0.15,0.05
  • START TO BE ELONGATED...
  • LUCI.0140
  • optimize gain->failed
  • tt->0.15 ho1=0.2
  • 204413
  • luci.0150
  • tt->0.85 ho1=0.2

clear PSF GUI

load 20211123.sav

nModes:54 r0:nan snMode:0 strehl:0 ttMod:0
Sat Nov 27 20:48:16.540 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[141] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

center pupil.

center_pupil_modified__2.png

close loop

close_loop_modify__2.png
  • tn:20211127_205146
  • good psf
  • luci.0170
  • optimize failed
  • tt->0.4,0.05,0
  • 205529
  • not bad.
  • luci.0180
  • tt->0.4,ho1=0.15,0
  • 205707
  • luci.0190
  • elongated.
  • 0.4,0.15,0.10
  • 205847
  • 0.7,0.2,0.10
  • 210021
  • 0.7,0.10,0.10
  • 210033
  • luci.0220
  • round

update OAO table for ACE and TTM.

27 21:13 table_LUCIFER_TTM-AO.txt -> table_LUCIFER_TTM-AO.txt.20211127

table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO.txt.20211123

preset TTM

Sat Nov 27 21:14:30.503 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[143] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • manula gain:0.5,0.15,0
  • 211942
at Nov 27 21:21:23.851 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[145] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • shutter 45
  • tn 212718
  • luci.0240

>>> makePupils.movePupils('20200916_173857',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_213036/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_213036/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_213036
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_213036"
'20211127_213036'
>>>

loaded the pupil in the AOtable for ACE. ->20211127_213036

difx=-0.45,dify=0

cl=x=-100, y=-26

tn:213926

luci0250

>>> makePupils.movePupils('20200916_173857',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_214340/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_214340/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_214340
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_214340"
'20211127_214340'
>>>
  • loaded pupil in the AO table.
  • close loop;lngain
  • tn215752
  • lici.0260
  • good

  • optimize
  • ok
  • optimizegain_20211127_214340.png
  • 215035
  • elongated...
check rotaiton x requires move up one pixel.

pupil_20211127_214340.png

>>> makePupils.movePupils('20200916_173857',2,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_220558/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_220558/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_220558
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211127_220558"
'20211127_220558
  • loaded in the AO table 20211127_220558
check rotation from 0 to 360

pupil_20211127_220558.png

pupil_rot_20211127_220558.png
  • good range.
  • close loop
  • tn20211127_222231
  • good PSF
  • luci.0280
  • optimize gain
optimize_20211127_220558.png
  • tn:_222448
  • tn_222626
  • luci.300
  • PSF ?

Handover

3:20PM

-- JuanCarlosGuerra - 27 Nov 2021
I Attachment Action Size Date Who Comment
bin1_20211127_rot0.pngpng bin1_20211127_rot0.png manage 1 MB 27 Nov 2021 - 18:51 JuanCarlosGuerra  
bin4-pupilcenering_20211127_195647.pngpng bin4-pupilcenering_20211127_195647.png manage 88 K 27 Nov 2021 - 20:06 JuanCarlosGuerra  
bin4_20211124_171903.pngpng bin4_20211124_171903.png manage 69 K 27 Nov 2021 - 17:51 JuanCarlosGuerra  
bin4_20211127_pupil_20211119_191303.pngpng bin4_20211127_pupil_20211119_191303.png manage 67 K 27 Nov 2021 - 17:33 JuanCarlosGuerra  
bin4_pupil20211127_174326.pngpng bin4_pupil20211127_174326.png manage 55 K 27 Nov 2021 - 18:22 JuanCarlosGuerra  
center_pupil_modified__2.pngpng center_pupil_modified__2.png manage 102 K 27 Nov 2021 - 20:50 JuanCarlosGuerra  
close_loop_modify__2.pngpng close_loop_modify__2.png manage 236 K 27 Nov 2021 - 20:52 JuanCarlosGuerra  
lower_rot.pngpng lower_rot.png manage 351 K 27 Nov 2021 - 23:07 JuanCarlosGuerra  
optimize_20211127_220558.pngpng optimize_20211127_220558.png manage 202 K 27 Nov 2021 - 22:25 JuanCarlosGuerra  
optimizegain_20211127_214340.pngpng optimizegain_20211127_214340.png manage 268 K 27 Nov 2021 - 22:02 JuanCarlosGuerra  
optimizegain_modified....pngpng optimizegain_modified....png manage 282 K 27 Nov 2021 - 20:30 JuanCarlosGuerra  
psf_rot190_bin1.pngpng psf_rot190_bin1.png manage 1 MB 27 Nov 2021 - 19:12 JuanCarlosGuerra  
pupil_20211127_214340.pngpng pupil_20211127_214340.png manage 62 K 27 Nov 2021 - 22:12 JuanCarlosGuerra  
pupil_20211127_220508.pngpng pupil_20211127_220508.png manage 62 K 27 Nov 2021 - 22:49 JuanCarlosGuerra  
pupil_20211127_220558.pngpng pupil_20211127_220558.png manage 86 K 27 Nov 2021 - 22:18 JuanCarlosGuerra  
pupil_rot_20211127_220558.pngpng pupil_rot_20211127_220558.png manage 62 K 27 Nov 2021 - 22:50 JuanCarlosGuerra  
pupilcentering...modify.pngpng pupilcentering...modify.png manage 95 K 27 Nov 2021 - 20:25 JuanCarlosGuerra  
upper_rot.pngpng upper_rot.png manage 412 K 27 Nov 2021 - 23:07 JuanCarlosGuerra  
Topic revision: r2 - 27 Nov 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback