Staff.

Summary.

Check pupil positions. Running Bin1 and there is 10um since december 7th shift. Close loop bin4 at 0.8 arcsec seeing, it shows good performances, J band.

Close loop Bin1 and check a pupil built fom the original. The pupil from the oringal shows to have a better performances in the modes residuals.... Analysis of the data on going...

Details.

Power up

8am

test preset.

Fri Dec 10 15:08:56.493 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[877] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

5811| MAIN > arbFunc doCenterPupils (, 'FLAOWFS', 'LUCIFER')
WfsArbitrator_34622|ERR| 232|2021-12-10 15:13:04.266768| MAIN > RETRY: Camera lens temperature too low. T= 6.05, minimum allowed is 7.00 (0) NO_ERROR

8:14am

bcu47 temperature 6.36C

8:21am

bcu47 temp 7.05C

bin1

center pupil.
  • Camera lens: x=-5, y=45 @341
December 7th CL,x=,7 y=59 @341

wobble check.

OK.

bin1_20211210.png

one pixel positive in X can be applied.

bin2

rotation check

shutter 60

Fri Dec 10 16:05:03.090 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[879] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:14,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

CL: x=7 ,y=13 @0

december 7; x=15 y=20 @0

ok

bin2_20211210.png

pupils_bin2-20211210.png

lower chamber lights off.

Bin4

Fri Dec 10 16:19:37.726 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[881] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:14,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

shutter 45

CL; x=30, y=47 @341

December7th. CL x=50 y=35 @341

bin4_pupil_20211210.png

close loop

ln gain
  • 20211210_165507
  • luci.20211210.5460
optimize
  • 0.05,0.30,0.10
  • tn:165854
  • luci1.5461
SHUTTER 50

LOADED DISTURBANCE.

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 17:01:28.628 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 17:01:28.689 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

CLOSE LOOP

LNGAIN
  • tn:_170502
  • luci.5462
optimize
  • 1.28,1.06,0.10
  • tn:_170754
moving a bit the source to have in a different place in the Luci field

i Dec 10 17:09:48.705 2021 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[888] - dx:-1.93182 dy:0.433524 slNull: f1:0.0 f2:1.0 freq:312 mag:0 nBins:4 nModes:54 r0:nan snMode:0 strehl:0 ttMod:0
Fri Dec 10 17:09:54.716 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[889] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 17:09:54.729 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 17:09:54.788 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

mag 15.5

close loop

lngain
  • tn:_171224
  • luci:5463
optimize
  • 1.26,1.13,0.10
  • tn:_171545
  • luci.5465
gopt:
  • 0.26,0.23,0.02
  • tn:_171911
  • 5466
  • tn:172039
  • 5467
bin4_171545.png

pause
  • light off
  • luci dark
  • light on
resume
  • tn:172331
  • luci.5469
after-resume_bin4172331.png

after the tn done
  • tn:_172459
  • luci.5470
PSF_172459.png
  • SR SE 0.066137
  • shutter 45
  • mag 16.5
  • _172802
  • luci.5471
  • _172915
  • luci.5472
  • _173043
  • luci.5473
loop open ..?17:32:00

reclose

optimize.
  • 1.18,0.78,0.10
gopt
2021-12-10 17:34:27.062822| -13006 - [AOException] RETRY: Cannot read current optical gain value (-5001) TIMEOUT_ERROR

optimize again
  • tn:173637
  • luci.5474
gopt
  • 0.25,0.23,0.02
  • tn:_173847
  • 5475
  • _174500
  • 5476
A DROP IN THE GOPT..?
  • tn:_175019
  • luci.5477
gopt seems stall..

and drop to zero...?17:52:50

gopt_droptozero.png
  • shuttter to 50
  • tn:_175342
  • luci.5478
  • tn:175446
  • luci.5479
  • optimize
  • 1.21,1.16,0.10
  • tn:-175715
  • luci.5480
  • gopt
  • 0.25,0.24,0.02
  • tn:_175943
  • 5481
  • _180108
  • 5482
move rot to 341

close loop

optimize
  • 1.23,1.01,0.10
  • TN:_181224
  • luci.5483
gopt
  • 0.25,0.20,0.02
  • tn:_181535
  • luci.5484
psf_181535.png
  • tn:_181733
  • luci.5485

TTM

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 18:22:43.663 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 18:22:43.744 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

shutter 50

luci camera to 3.75

cloe loop

lngain

luci.5486

tt 0.8,0.05,0

_183028

luci.5487

shutter open

mag11

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 18:34:48.234 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 18:34:48.311 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

loop 400Hz.

close loop.

tt=0.8,0.05

rmslope=0.77

luci.5488

filter H

_184402

luci.5489

light off

luci dark

luci.5490

2 frames

luci.5491


test under----

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 18:51:32.104 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 18:51:33.615 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.0000_sd3892.fits

luci open loop

luci.5492

tn:_185545 disturbance on

luci.5493

tt 0.95,0.05,0

_185714

luci.5494

stop 12:00

make a shift pupil for bin1 for 1 pixel up in X axis

>>> makePupils.movePupils('20211125_004613',1,0,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_191322/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_191322/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_191322
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_191322"
'20211210_191322'
>>>

loaded the pupil 20211210_191322'

Fri Dec 10 19:16:25.196 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[904] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • camera lens: x=56.81, y=23.7 @360
check downward.

124

close loop check.

optimize not disturbance.

bin1_cl_pupil20211210_191322.png
  • tn:20211210_193610
  • luci.20211211.5001
loaded disturbance

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 19:38:39.085 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 19:38:39.153 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits
Fri Dec 10 19:38:39.995 2021 aos.info - FLAO: WFS: loading setup for LUCIFER

close loop angle 0

optimize

high order skip frames but process complete
  • gain 3.48,0.67,0.50
gopt
  • 1.48,0.28,0.22
  • tn:194429
  • luci.20211211.5002
optimize
  • 3.55,0.67,0.44
gopt
  • 1.53,0.29,0.20
  • _194914
AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 19:51:18.838 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 19:51:18.896 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits
Fri Dec 10 19:51:19.719 2021 aos.info - FLAO: WFS: loading setup for LUCIFER

optimize

3.06,0.74,0.41

_202120

maybe a bit of rotation...

open loop

cancel preset.

loaded the default pupil:20211125_004613

ec 10 20:08:05.155 2021 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[911] - dx:-3.10956 dy:-0.266864 slNull: f1:0.0 f2:1.0 freq:900 mag:0 nBins:1 nModes:501 r0:nan snMode:0 strehl:0 ttMod:0
Fri Dec 10 20:08:10.431 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[912] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 20:08:10.446 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 20:08:10.508 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits
Fri Dec 10 20:08:11.332 2021 aos.info - FLAO: WFS: loading setup for LUCIFER

optimize gain..

same skip frames...

2.89,0.71,0.36

gopt
  • tn:_201457
  • luci.5005
before_homing.png

loop pause

homing rotator.

tracking rotators

resume

after_homing.png
  • tn:202141
  • luci.5006
optimize gain.
  • 2.85,0.69,0.52
optimize-afterhoming.png

gopt.
  • tn:_202551
psf_202551.png

luci5007

move to 341

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 20:29:23.829 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 20:29:23.889 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits

close loop.

optimize
  • 2.41.0.71,0.54
optimize_341_.png

gopt
  • 1.12,0.33,0.26
  • _203538
psf_203538.png
  • luci5008
  • luci dark. 5009
loaded new pupil.

ln -fs table_LUCIFER_ACE-AO.txt.20211210 table_LUCIFER_ACE-AO.txt

Fri Dec 10 20:41:02.772 2021 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[921] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 20:41:02.789 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 20:41:02.863 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits
Fri Dec 10 20:41:03.572 2021 aos.info - FLAO: WFS: loading setup for LUCIF

close loop.

optimize gain

3.18,0.69,0.54

gopt
  • 2-10 20:49:57.941183| -13006 - [AOException] RETRY: Cannot read current optical gain value (-5001) TIMEOUT_ERROR
  • retry
  • 1.47,0.32,0.26
  • 20211210_205008

check pupil #102

REsting pupil #102

optimize gain

skip frames with ho2

loaded pupil 102 in the AO table

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Dec 10 21:16:10.834 2021 aos.info - WFS source set to: FLAOWFS
Fri Dec 10 21:16:10.898 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits
Fri Dec 10 21:16:11.646 2021 aos.info - FLAO: WFS: loading setup for LUCIFER
  • cl: x=59, y=26 @341
optimize

finished first t and it didn;t display the ho1

retry
  • tt2.04,0.69,0.48
  • tn:_212204
  • luci:5011
modalplot_212204.png

psf_212004.png
  • 212422
  • luci.5012
wobble backward.

ok.

psf_210322.png

Power off.

tor_34622|INF| 3803|2021-12-10 21:50:43.805276| MAIN > Turning WFS off
WfsArbitrator_34622|INF| 3804|2021-12-10 21:50:43.806633| MAIN > Setting tiptilt to amp= 0.00, freq=1000.00, offx= 0.00, offy= 0.00
WfsArbitrator_34622|DEB| 3805|2021-12-10 21:50:44.025691| MAIN > ('set framerate: 1000.000000 Hz',)
WfsArbitrator _34622|ERR| 3806|2021-12-10 21:50:44.228816| MAIN > Controller in wrong state for requested operation. (-11005) CTRL_WRONG_STATE_ERROR

troubleshoot:

  • Operate
  • ok
  • power off
  • ok

Handover.

2:40PM

Analysis.

bin1 orig pupil: 20181103_001410

>>> makePupils.movePupils('20181103_001410',6,10,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210117/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210117/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210117
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210117"
'20211210_210117'

#101

>>> makePupils.movePupils('20181103_001410',7,10,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210322/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210322/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210322
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211210_210322"
'20211210_210322

#102
Update the pupil 20211210_210322 in the table_LUCIFER_ACE-AO.txt.20211210

_170502.

LNGain

psf_170502.png

c->modalplot

tn:_170754

PSF_172459.png

modalplot-170754.png

175715

sr=4%

psf_175715.png

modalplot_175715.png

jitter-175715.png

-- JuanCarlosGuerra - 10 Dec 2021
I Attachment Action Size Date Who Comment
PSF_172459.pngpng PSF_172459.png manage 48 K 10 Dec 2021 - 17:48 JuanCarlosGuerra  
after-resume_bin4172331.pngpng after-resume_bin4172331.png manage 175 K 10 Dec 2021 - 17:25 JuanCarlosGuerra  
after_homing.pngpng after_homing.png manage 180 K 10 Dec 2021 - 20:22 JuanCarlosGuerra  
before_homing.pngpng before_homing.png manage 178 K 10 Dec 2021 - 20:23 JuanCarlosGuerra  
bin1_20211210.pngpng bin1_20211210.png manage 56 K 10 Dec 2021 - 16:22 JuanCarlosGuerra  
bin1_cl_pupil20211210_191322.pngpng bin1_cl_pupil20211210_191322.png manage 259 K 10 Dec 2021 - 19:41 JuanCarlosGuerra  
bin1_pupil_20211210_210322.pngpng bin1_pupil_20211210_210322.png manage 76 K 10 Dec 2021 - 21:59 JuanCarlosGuerra  
bin1_shiftonepix_x-20211210.pngpng bin1_shiftonepix_x-20211210.png manage 67 K 10 Dec 2021 - 19:36 JuanCarlosGuerra  
bin2_20211210.pngpng bin2_20211210.png manage 79 K 10 Dec 2021 - 16:25 JuanCarlosGuerra  
bin4_171545.pngpng bin4_171545.png manage 408 K 10 Dec 2021 - 17:17 JuanCarlosGuerra  
bin4_pupil_20211210.pngpng bin4_pupil_20211210.png manage 94 K 10 Dec 2021 - 16:38 JuanCarlosGuerra  
gopt_droptozero.pngpng gopt_droptozero.png manage 966 K 10 Dec 2021 - 17:59 JuanCarlosGuerra  
jitter-175715.pngpng jitter-175715.png manage 13 K 11 Dec 2021 - 06:02 JuanCarlosGuerra  
modalplot-170754.pngpng modalplot-170754.png manage 18 K 11 Dec 2021 - 00:58 JuanCarlosGuerra  
modalplot_175715.pngpng modalplot_175715.png manage 49 K 11 Dec 2021 - 06:02 JuanCarlosGuerra  
modalplot_212204.pngpng modalplot_212204.png manage 21 K 10 Dec 2021 - 22:00 JuanCarlosGuerra  
optimize-afterhoming.pngpng optimize-afterhoming.png manage 507 K 10 Dec 2021 - 20:25 JuanCarlosGuerra  
optimize_341_.pngpng optimize_341_.png manage 594 K 10 Dec 2021 - 20:35 JuanCarlosGuerra  
psf_170502.pngpng psf_170502.png manage 131 K 11 Dec 2021 - 00:57 JuanCarlosGuerra  
psf_175715.pngpng psf_175715.png manage 53 K 11 Dec 2021 - 06:02 JuanCarlosGuerra  
psf_181535.pngpng psf_181535.png manage 119 K 10 Dec 2021 - 22:18 JuanCarlosGuerra  
psf_202551.pngpng psf_202551.png manage 47 K 10 Dec 2021 - 20:33 JuanCarlosGuerra  
psf_203538.pngpng psf_203538.png manage 47 K 10 Dec 2021 - 20:39 JuanCarlosGuerra  
psf_210322.pngpng psf_210322.png manage 47 K 10 Dec 2021 - 21:59 JuanCarlosGuerra  
psf_212004.pngpng psf_212004.png manage 48 K 10 Dec 2021 - 22:03 JuanCarlosGuerra  
pupils_bin2-20211210.pngpng pupils_bin2-20211210.png manage 104 K 10 Dec 2021 - 16:22 JuanCarlosGuerra  
Topic revision: r4 - 13 Dec 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback