Power up WFS

8:30am

OCAM NOT LIVE.
  • POWER CYCLE BCU2K AND SET 1X1CROPPED.
  • OK
How the pupils looks like.

bin4 to bin3 stack the BCU2K
  • Change to Bin 2 and ocam was live again.
  • Change to bin3 ocam not live.
From the WFS arbitrator works better.

First from WFSarb set bin3 and after the change_ocam_mode.py can be executed for Bin3.

Thu Dec 2 16:08:55.430 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[482] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Dec 2 16:0

Chage to pupreIntegratorCoul

skip frame with 0.1_10modi applied

shutter 44->mag 17.5

freq 345 shutter 50->15.3

pureintegratorSoul

becomes unstable....

loaded fogettingfactorBIN4v2SOUL
  • SKIP FRAMES.
  • LOOP UNSTABLE

LOADED THE PUPIL 20200916_173857.

DIFFX=1.47

DIFFY=2.05

CLX=95,Y=50

bin4_20211202_pupil20200917_173857.png

luci images open loop

luci.20211202.5102 loaded the flat.

luci1.20211202.5100 after bin4 close loop with default pupils.

luci1.20211202.5101 after close loop bin4 default pupils.

flat20211123.sav.png

preset bin1

Thu Dec 2 16:41:03.126 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[485] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:11,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

after center pupil

luci1.503.

close loop 0.1modi10

luci.504

close loop HO1=0.15 and HO2=0.1

luci.505

open loop

luci.5106

dark.luci.5107

move to H

luci.5106

subs_openloopH_bin1.png

close loop

l/d=20

luci.5109

change to bin4 from the wfsarbitrator

freq=200,4l/d/496

loaded pupil->20211201_223347

0.1modi10

difx=1.6

dify=0.4

luci.5110

loaded pupil

WARNING: MasterDiagnostic is not working properly. Impossible to notify it if frame size changes !
ttctrl interface loaded
Lamp interface loaded
Camera lens interface loaded
Calling setHObinningOrCcdMode with mode=5
Setting pupil: 20211201_181140

gain manual:

luci1.5111

psf distort

put the pupil 20211201_181140 in the AOtable

table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO.txt.20211202

Thu Dec 2 17:21:53.782 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[487] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

The flat is not good...

close loop.

The poupils are not fully illuminated.using pupil -223347

Thu Dec 2 17:29:23.471 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[489] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupil rotator->178.27

homing.

close loop.. still showing not full illuminated pupils.

bin3

Thu Dec 2 17:43:37.525 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[491] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Dec 2 17:43:37.546 2021 aos.info - WFS source set to: FLAOWFS
Thu Dec 2 17:43:38.3

shutter 50

>>> makePupils.movePupils('20210506_225353',0,1,240,4)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_174912/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_174912/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 148
out_of_pupil_slopes 2552
lines_to_add 11
residual_pixels_to_add 38
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_174912
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_174912"
'20211202_174912'
>>>

original pupil: 20200210_000000

loaded the pupil

close pupil

gain0.1modi10

luci.5112

>>> makePupils.movePupils('20200210_000000',1,1,240,4)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_175957/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_175957/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 148
out_of_pupil_slopes 2552
lines_to_add 11
residual_pixels_to_add 38
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_175957
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_175957"
'20211202_175957'

The resd circle doen;t show in the ocam viewer.

Brandon checked, the process is running in the right place and the ocam viewer too.

load pupil in the ao table.

Thu Dec 2 18:08:28.785 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[493] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

pupil centering looks good.

the starting flat it seems not good.

close loop LNGain

luci.5113

optimize gain

bin3_optimize.png
  • ok.
  • no skip frame
  • stable
  • TN:20211202_181917
slope rms->0.58

looks good.

loaded disturbance.

Thu Dec 2 18:21:25.942 2021 aos.info - Stages offset applied - dx: -1.23819; dy: -0.230182
Thu Dec 2 18:21:25.960 2021 aos.AcquireRef{FLAO}.complete - AcquireRef{FLAO}[496] - dx:-1.23819 dy:-0.230182 slNull: f1:0.0 f2:1.0 freq:469 mag:0 nBins:3 nModes:90 r0:nan snMode:0 strehl:0 ttMod:0
Thu Dec 2 18:22:59.115 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[497] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Dec 2 18:22:59.128 2021 aos.info - WFS source set to: FLAOWFS
Thu Dec 2 18:22:59.186 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.0000_sd3892.fits
Thu Dec 2 18:23:00.018 2021 aos.info - FLAO: WFS: loading setup for LUCIFER
Thu Dec 2 18:23:00.032 2021 aos.info - FLAO: WFS: configuring CCDs

pupils.

pupils_bin3.png

optimize gain.

bin3_gopt_dist.png
  • luci.5115
  • tn:20211202_182918
  • loop pause
  • lights off
  • luci dark 5 frames
  • shutter 50
  • | SR SE | 0.284713
psf_bin3_20211202.png

bin3_20211202_elab.png

>>> makePupils.movePupils('20200210_000000',2,1,240,4)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_185044/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_185044/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 148
out_of_pupil_slopes 2552
lines_to_add 11
residual_pixels_to_add 38
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_185044
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode4/20211202_185044"
'20211202_185044'
>>>044

Loaded pupil: 20211202_185044

Pupil rotation check.

bin3_20211202.png

working with Enrico.

Thu Dec 2 19:42:34.944 2021 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[505] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Dec 2 19:42:34.969 2021 aos.info - WFS source set to: FLAOWFS
Thu Dec 2 19:42:35.034 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits
Thu Dec 2 19:42:36.032 2021 aos.info - FLAO: WFS: loading setup for LUCIFER
Thu Dec 2 19:42:36.051 2021 aos.info - FLAO: WFS: configuring CCDs

loaded defaul pupil 20200916_173857

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:17,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Dec 2 19:54:28.087 2021 aos.info - WFS source set to: FLAOWFS
Thu Dec 2 19:54:28.145 2021 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

  • Taking data for TT nd HO1 corrections.
  • pupil 20211201_181140
  • shifted x=1,y=2
  • no dist
  • TT=0.1
  • TN20211202_193721.
  • gain0.1_10modi
  • TN:_194003
  • distON
  • _194715

>>> makepupils.movePupils('20200916_173857_LBTI',1,2,240,5)
Traceback (most recent call last):
File "<console>", line 1, in <module>
NameError: name 'makepupils' is not defined
>>> makePupils.movePupils('20200916_173857_LBTI',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837"
'20211202_202837'
>>>

Shifted.

0.1_10modi

b4_shifted_0p110modi.png

LnGain

b4_lngain_pupshifted.png

Disturbance ON:

TN:20211202_194715

HO1=0.05

TN:_194235

loaded pupil 20200916_173857_LBTI same as 20200916_173857

No dis

b4_lngain_centerpupil.png

+ disturbance

B4_manualCL_LBTI.png

0.1modi10

B4_CL_0p1modi10.png

  • LNGain
  • TN:195836
  • 0.1modi10
  • _195942
  • tt=0.4,ho1=0.15
B4_manualCL_LBTI.png
  • _200601
Stop at 1:00PM give back for work in the upper tree house with chamber lights on.

Handover.

Analysis data.

Just lernt from the images headers.

Headers are different because the origin files..The contents are OK.

>>> f1 = fits.open("20211202_211114/pup1.fits")
>>> f2 = fits.open("20211202_202837/pup1.fits")
>>> f1[0].header
SIMPLE = T / conforms to FITS standard
BITPIX = 32 / array data type
NAXIS = 1 / number of array dimensions
NAXIS1 = 80
EXTEND = T
PUPFROM = '20200916_173857'
SHIFT_X = 1
SHIFT_Y = 2
STARTPIX= 0
>>> f2[0].header
SIMPLE = T / conforms to FITS standard
BITPIX = 32 / array data type
NAXIS = 1 / number of array dimensions
NAXIS1 = 80
EXTEND = T
PUPFROM = '20200916_173857_LBTI'
SHIFT_X = 1
SHIFT_Y = 2
STARTPIX= 0

> makePupils.movePupils('20200916_173857_LBTI',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_202837"
'20211202_202837'
>>> makePupils.movePupils('20200916_173857',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_211114/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_211114/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_211114
/bin/mv -f "/ot/calib/wfs/cuhome/AOeng/pause_resume_test_20211011/aororrent/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_211114"

'20211202_211114'
>>>

Differences between the two files shifted from the original on are different.

drwxrwxrwx. 2 AOeng AOeng 4096 Dec 2 20:28 20211202_202837/
drwxrwxrwx. 2 AOeng AOeng 4096 Dec 2 21:11 20211202_211114/
[AOeng@soul-sxwfs mode5]$ diff -r 20211202_211114 20211202_202837
Binary files 20211202_211114/pup1.fits and 20211202_202837/pup1.fits differ
Binary files 20211202_211114/pup2.fits and 20211202_202837/pup2.fits differ
Binary files 20211202_211114/pup3.fits and 20211202_202837/pup3.fits differ
Binary files 20211202_211114/pup4.fits and 20211202_202837/pup4.fits differ

Differences between the the two original directories are none.

[AOeng@soul-sxwfs mode5]$ diff -r 20200916_173857 20200916_173857_LBTI
[AOeng@soul-sxwfs mode5]$

>>> makePupils.move.Pupils('20200916_173857',1,2,240,5)
Traceback (most recent call last):
File "<console>", line 1, in <module>
AttributeError: 'module' object has no attribute 'move'
>>> makePupils.movePupils('20200916_173857',1,2,240,5)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_213235/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_213235/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 80
out_of_pupil_slopes 2688
lines_to_add 12
residual_pixels_to_add 12
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_213235
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode5/20211202_213235"
'20211202_213235'
>>>

[AOeng@soul-sxwfs mode5]$ diff -r 20211202_213235 20211202_202837
Binary files 20211202_213235/pup1.fits and 20211202_202837/pup1.fits differ
Binary files 20211202_213235/pup2.fits and 20211202_202837/pup2.fits differ
Binary files 20211202_213235/pup3.fits and 20211202_202837/pup3.fits differ
Binary files 20211202_213235/pup4.fits and 20211202_202837/pup4.fits differ
[AOeng@soul-sxwfs mode5]$

same test for Bin1

original pupil: 20181103_001410

>>> makePupils.movePupils('20181103_001410',1,1,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214106/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214106/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214106
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214106"
'20211202_214106'
>>>

>>> makePupils.movePupils('20181103_001410',1,1,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214238/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214238/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214238
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214238"
'20211202_214238'
>>>

diff -r 20211202_214106 20211202_214238
[AOeng@soul-sxwfs mode2]$

>>> makePupils.movePupils('20181103_001410',1,2,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214459/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214459/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214459
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214459"
'20211202_214459'
>>>

>>> makePupils.movePupils('20181103_001410',1,2,240,2)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214546/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214546/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 1256
out_of_pupil_slopes 336
lines_to_add 2
residual_pixels_to_add 24
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214546
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20211202_214546"
'20211202_214546'
>>>

diff -r 20211202_214459 20211202_214546
[AOeng@soul-sxwfs mode2]$

no differences..

Same for Bin2

>>> makePupils.movePupils('20211125_011619',1,1,240,3)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222128/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222128/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 328
out_of_pupil_slopes 2192
lines_to_add 10
residual_pixels_to_add 8
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222128
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222128"
'20211202_222128'
>>>

>>> makePupils.movePupils('20211125_011619',1,1,240,3)
makePupils_fromMasks(): moving from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pupdata.txt to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222250/pupdata.txt
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pupdata.txt" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222250/pupdata.txt"
num_slopes 2848
len(pupi_list[0]) 328
out_of_pupil_slopes 2192
lines_to_add 10
residual_pixels_to_add 8
Entering main loop
Out of main loop
makePupils_common: moving fits file from /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3 to /home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222250
/bin/mv -f "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/pup1.fits" "/home/AOeng/pause_resume_test_20211011/aoroot/calib/wfs/current/ocam2/LUTs/mode3/20211202_222250"
'20211202_222250'
>>>

$ diff -r 20211202_222128 20211202_222250
[AOeng@soul-sxwfs mode3]$

-- JuanCarlosGuerra - 02 Dec 2021
I Attachment Action Size Date Who Comment
B4_CL_0p1modi10.pngpng B4_CL_0p1modi10.png manage 912 K 02 Dec 2021 - 23:57 JuanCarlosGuerra  
B4_manualCL_LBTI.pngpng B4_manualCL_LBTI.png manage 871 K 02 Dec 2021 - 23:55 JuanCarlosGuerra  
b4_lngain_centerpupil.pngpng b4_lngain_centerpupil.png manage 918 K 02 Dec 2021 - 23:58 JuanCarlosGuerra  
b4_lngain_pupshifted.pngpng b4_lngain_pupshifted.png manage 911 K 03 Dec 2021 - 00:00 JuanCarlosGuerra  
b4_shifted_0p110modi.pngpng b4_shifted_0p110modi.png manage 899 K 03 Dec 2021 - 00:03 JuanCarlosGuerra  
bin3_20211202.pngpng bin3_20211202.png manage 106 K 02 Dec 2021 - 21:06 JuanCarlosGuerra  
bin3_20211202_elab.pngpng bin3_20211202_elab.png manage 51 K 02 Dec 2021 - 23:26 JuanCarlosGuerra  
bin3_cl_lngain20211202.pngpng bin3_cl_lngain20211202.png manage 496 K 02 Dec 2021 - 18:16 JuanCarlosGuerra  
bin3_gopt_dist.pngpng bin3_gopt_dist.png manage 136 K 02 Dec 2021 - 18:29 JuanCarlosGuerra  
bin3_optimize.pngpng bin3_optimize.png manage 455 K 02 Dec 2021 - 18:24 JuanCarlosGuerra  
bin4_20211202_pupil20200917_173857.pngpng bin4_20211202_pupil20200917_173857.png manage 291 K 02 Dec 2021 - 16:33 JuanCarlosGuerra  
flat20211123.sav.pngpng flat20211123.sav.png manage 1 MB 02 Dec 2021 - 16:41 JuanCarlosGuerra  
psf_bin3_20211202.pngpng psf_bin3_20211202.png manage 46 K 02 Dec 2021 - 18:34 JuanCarlosGuerra  
pupils_bin3.pngpng pupils_bin3.png manage 106 K 02 Dec 2021 - 18:25 JuanCarlosGuerra  
subs_openloopH_bin1.pngpng subs_openloopH_bin1.png manage 1 MB 02 Dec 2021 - 16:54 JuanCarlosGuerra  
Topic revision: r1 - 03 Dec 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback