Staff:

LBTO: Juan Carlos

Summary:

Details:

telescope released at 10:50am

11:00 WFS power ON.
  • OCAM WARM.
  • WFS:100HZ, BIN1, EMG=1,10L/D
Alignment of pupils.

LOAD TELESCOPE_INTERNAL_SOURCE_101421

Small adjustment of stagey position.
  • TN: 20211105_184843
  • CL:-56.8,-23.0
  • diff:-0.36,-0.18.
  • rot 341.
pupils_internal_110521.png

rotation 341 to 0

starting 11:00

the circle doen't get the center of the pupils.

Saving tracking numbers to identify center of pupils post-processing . The empy center position is due to a bad computation of the pupils centers.

180 degrees upper left illumination (one pupil)

150 degrees two upper pupil illuminated, some elongation.

90 degrees, stop and moved up... drfiting up....

180 moved up and elongated.

0 all the way down.
rot cx cy cx cy TN  
0 90.07 142.7 42.55 94.71 _190504  
30 89.74 141.9 41.74 93.9 _190633  
60 89.94 141.46 41.88 93.44 _190736  
90 82.5 143.10 34.38 95.17 _190859  
120 82 140.2 34.02 92.19 _191017  
150 83.04 145.2 35 97.59 _191239  
180         _191531  
210         _191707  
240 76.46 142 28.68 94 _191823  
270 79.38 141.5 31.3 93.6 _191938  
300 82.8 140.8 34.8 92.8 _192113  
330 86.3 140.7 38.7 92.69 _192217  
360 90.97 141.4 42.98 93.4 _192358  
390 91.02 140.58 43 92.58 _192500  
420         _192617  
390 91.4 140.5 43.4 92.5 _192739  
360 92.05 142.24 44.05 94.24 _192840  
340 90.63 143.28 42.71 95.42 _193016  
330 90.7 142.19 42.67 94.19 _193118  
300 89.43 142 414 93.99 _193232  
270 83.7 140.74 35.4 92.74 _193350  
240 78.38 143.96 30.38 95.97 _193521  
210         _193621  
180         _193725  
150         _193839  
120 77.36 142.18 29.8 94.1 _194001  
90 82.99 140.9 34.92 92.94 _194120  
60 85 141 37 141 _194231  
30         _194231  
0         _194610 0.15,-1.94
             
             
  • move from 90 to 180-> TN:195034
  • at 45 -> TN:_195344, diffx=1,diffy=-0.59,91.01,141.59/43.01,93.59
  • at 90-> TN: move up..._195606 , diffx=-7.05,-2.84, 82.96,143.83,34.96,95.94.
  • at 340->. diffx=-0.39,-0.43,89.61,141.43,41.62,93.43

POWER ON DX.

1PM
  • THE DX PUPILS ARE NOT EVEN ILLUMINATED. This has been observed before and discussed with Enrico. It seems the alignment was done using the positions of the external pupil (telescope pupil)
  • 100 Hz,
  • cl:65,-100
  • diffx=0.5, diffy=2.54
  • 210
  • tn:20211105_202805
  • load board setup: 09 Jul21 internalcheck
rot cx cy cx cy TN  
0         _  
30 84.45 139 36.45 91 _210011  
60 84.49 138.99 36.49 90.99 _210052  
90 84.63 138.87 36.63 90.87 210208  
120 84.77 138.84 36.61 90.84 _210313  
150 84.68 138.78 36.7 90.78 _210422  
180 84.61 138.69 36.61 90.69 210542  
210 84.6 138.68 36.6 90.68 210649  
240 84.5 138.42 36.5 90.45 _203045 0.5,2.55
270 84.49 138.4 36.5 138 _203249  
300 84.47 138.47 36.47 138.47 _203450 0.47,2.53
330 84.23 138.52 36.23 138.62 _203701 0.21,2.49
360 84.42 138.53 36.42 138.53 _203903 0.42,2.47
390 84.38 138.5 36.38 90.51 _204026  
420 84.44 138.5 36.44 90.55 _204159 0.44,2.45
390 84.27 138.6 36.27 90.62 _204318 0.27,2.38
360 84.27 138.6 36.27 90.62 _204434  
340 84.27 138.63 36.27 90.63 _204551  
330 84.27 138.63 36.27 90.63 _204658  
300 84.26 138.58 36.26 90.51 _204753  
270 84.43 138.51 36.43 90.51 _204858  
240 84.5 138.53 36.46 90.51 _204950  
210 84.52 138.51 36.52 90.51 _205053  
180 84.62 138.51 36.62 90.51 _205158  
150 84.75 138.46 36.75 90.46 _205259  
120 84.71 138.51 36.7 90.51 _205403  
90 84.74 138.82 36.74 90.82 _205522  
60 84.50 139 36.5 90.95 _205626  
30 84.49 138.96 36.49 90.96 _205729  
0 84.5 138.97 36.5 90.97 _205845  
             
           
  • at 60 4 illuminated.
  • at 30 three illuminated.
  • at 0 two right illuminated.
  • at 90 three diagonal left illuminated.
  • at 120 two up illumated.
  • at 150 two left illuminated
  • at 180 two bright left and two faint right.
  • at 210 four illuminated.

HANDOVER.

2pm. soulDX power off. cube translator and rotator parked. light source off.
  • soulSX power off.
  • parked LFBG 341.
  • parked RFBG 210

-- JuanCarlosGuerra - 05 Nov 2021