W_check

[AOeng@soul-dxwfs SOUL]$ w_check

msgdrtdb NOT running

mirrorctrl NOT running

pinger running but UNKOWN connection state to Msgd

powerboard NOT running

slopecompctrl NOT running

techviewctrl NOT running

ocam2 NOT running

ccd47 NOT running

ttctrl NOT running

adc1 NOT running

adc2 NOT running

rerotator NOT running

filterwheel1 NOT running

filterwheel2 NOT running

mercury NOT running

cubestage NOT running

baysidex NOT running

baysidey NOT running

baysidez NOT running

cameralensctrl NOT running

picctrl NOT running

optloopdiag NOT running

pupilcheckocam2 NOT running

masterdiagnostic NOT running

housekeeperWFS NOT running

dbfiller NOT running

wfsarb NOT running

gopt NOT running

Error: w_check script UNABLE to connect to RTDB!

The exception wasthStart error(-2010): Connection closed at the other end
  • w_stop
  • w_start
[AOeng@soul-dxwfs SOUL]$ w_start

Side: R

Subsystem: WFS

startProcessByName(): starting process /home/AOeng/aoroot/bin/msgdrtdb

clean_ts8 script is disabled

-- Stage mirrorctrl pinger powerboard slopecompctrl techviewctrl ocam2 ccd47 ttctrl adc1 adc2 rerotator filterwheel1 filterwheel2 mercury cubestage baysidex baysidey baysidez cameralensctrl picctrl optloopdiag pupilcheckocam2

startProcessByName(): starting process /home/AOeng/aoroot/bin/MirrorCtrl -i mirrorctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/Pinger -i pinger

startProcessByName(): starting process /home/AOeng/aoroot/bin/RelayCtrl -i powerboard

startProcessByName(): starting process /home/AOeng/aoroot/bin/SlopeCompCtrl2K -i slopecompctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/FrameGrabberCtrl -i techviewctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/OCAM2Ctrl -i ocam2

startProcessByName(): starting process /home/AOeng/aoroot/bin/JoeCtrl -i ccd47

startProcessByName(): starting process /home/AOeng/aoroot/bin/TTCtrl -i ttctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc1

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc2

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i rerotator

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel1

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel2

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i mercury

startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i cubestage

startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidex

startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidey

startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidez

startProcessByName(): starting process /home/AOeng/aoroot/bin/CameraLensCtrl -i cameralensctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/PicCtrlW2 -i picctrl

startProcessByName(): starting process /home/AOeng/aoroot/bin/OptLoopDiagnostic -i optloopdiag

startProcessByName(): starting process /home/AOeng/aoroot/bin/pupilcheck.py -i pupilcheckocam2 ocam2

-- Stage masterdiagnostic housekeeperWFS dbfiller

startProcessByName(): starting process /home/AOeng/aoroot/bin/MasterDiagnostic -i masterdiagnostic

startProcessByName(): starting process /home/AOeng/aoroot/bin/housekeeperWFS -i housekeeperWFS

startProcessByName(): starting process /home/AOeng/aoroot/bin/DBFiller -i dbfiller

Traceback (most recent call last):

File "/home/AOeng/aoroot/bin/w_start", line 256, in <module>

if name == '__main__': main()

File "/home/AOeng/aoroot/bin/w_start", line 144, in main

processControl.waitProcessStart( app, process)

File "/home/AOeng/aoroot/lib/python/AdOpt/processControl.py", line 240, in waitProcessStart

raise AdOptError ( AOErrCode ['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)

AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR

[AOeng@soul-dxwfs SOUL]$

[AOeng@soul-dxwfs SOUL]$ w_start
  • Side: R
  • Subsystem: WFS
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/msgdrtdb
  • clean_ts8 script is disabled
  • -- Stage mirrorctrl pinger powerboard slopecompctrl techviewctrl ocam2 ccd47 ttctrl adc1 adc2 rerotator filterwheel1 filterwheel2 mercury cubestage baysidex baysidey baysidez cameralensctrl picctrl optloopdiag pupilcheckocam2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/MirrorCtrl -i mirrorctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/Pinger -i pinger
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/RelayCtrl -i powerboard
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SlopeCompCtrl2K -i slopecompctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/FrameGrabberCtrl -i techviewctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/OCAM2Ctrl -i ocam2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/JoeCtrl -i ccd47
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/TTCtrl -i ttctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc1
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i rerotator
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel1
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i mercury
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i cubestage
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidex
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidey
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidez
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CameraLensCtrl -i cameralensctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/PicCtrlW2 -i picctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/OptLoopDiagnostic -i optloopdiag
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/pupilcheck.py -i pupilcheckocam2 ocam2
  • -- Stage masterdiagnostic housekeeperWFS dbfiller
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/MasterDiagnostic -i masterdiagnostic
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/housekeeperWFS -i housekeeperWFS
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/DBFiller -i dbfiller
  • Traceback (most recent call last):
  • File "/home/AOeng/aoroot/bin/w_start", line 256, in <module>
  • if name == '__main__': main()
  • File "/home/AOeng/aoroot/bin/w_start", line 144, in main
  • processControl.waitProcessStart( app, process)
  • File "/home/AOeng/aoroot/lib/python/AdOpt/processControl.py", line 240, in waitProcessStart
  • raise AdOptError ( AOErrCode ['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)
  • AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR
[AOeng@soul-dxwfs SOUL]$

[AOeng@soul-dxwfs SOUL]$ w_start
  • Side: R
  • Subsystem: WFS
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/msgdrtdb
  • clean_ts8 script is disabled
  • -- Stage mirrorctrl pinger powerboard slopecompctrl techviewctrl ocam2 ccd47 ttctrl adc1 adc2 rerotator filterwheel1 filterwheel2 mercury cubestage baysidex baysidey baysidez cameralensctrl picctrl optloopdiag pupilcheckocam2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/MirrorCtrl -i mirrorctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/Pinger -i pinger
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/RelayCtrl -i powerboard
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SlopeCompCtrl2K -i slopecompctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/FrameGrabberCtrl -i techviewctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/OCAM2Ctrl -i ocam2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/JoeCtrl -i ccd47
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/TTCtrl -i ttctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc1
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i adc2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i rerotator
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel1
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i filterwheel2
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i mercury
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/SimpleMotorCtrl -i cubestage
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidex
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidey
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CopleyCtrl -i baysidez
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/CameraLensCtrl -i cameralensctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/PicCtrlW2 -i picctrl
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/OptLoopDiagnostic -i optloopdiag
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/pupilcheck.py -i pupilcheckocam2 ocam2
  • -- Stage masterdiagnostic housekeeperWFS dbfiller
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/MasterDiagnostic -i masterdiagnostic
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/housekeeperWFS -i housekeeperWFS
  • startProcessByName(): starting process /home/AOeng/aoroot/bin/DBFiller -i dbfiller
  • Traceback (most recent call last):
  • File "/home/AOeng/aoroot/bin/w_start", line 256, in <module>
  • if name == '__main__': main()
  • File "/home/AOeng/aoroot/bin/w_start", line 144, in main
  • processControl.waitProcessStart( app, process)
  • File "/home/AOeng/aoroot/lib/python/AdOpt/processControl.py", line 240, in waitProcessStart
  • raise AdOptError ( AOErrCode ['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)
  • AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR
  • [AOeng@soul-dxwfs SOUL]$ w_check
  • msgdrtdb running
  • mirrorctrl running and connected to Msgd
  • pinger running and connected to Msgd
  • powerboard running and connected to Msgd
  • slopecompctrl running and connected to Msgd
  • techviewctrl running and connected to Msgd
  • ocam2 running and connected to Msgd
  • ccd47 running and connected to Msgd
  • ttctrl running and connected to Msgd
  • adc1 running and connected to Msgd
  • adc2 running and connected to Msgd
  • rerotator running and connected to Msgd
  • filterwheel1 running and connected to Msgd
  • filterwheel2 running and connected to Msgd
  • mercury running and connected to Msgd
  • cubestage running and connected to Msgd
  • baysidex running and connected to Msgd
  • baysidey running and connected to Msgd
  • baysidez running and connected to Msgd
  • cameralensctrl running and connected to Msgd
  • picctrl running and connected to Msgd
  • optloopdiag running and connected to Msgd
  • pupilcheckocam2 running and connected to Msgd
  • masterdiagnostic running and connected to Msgd
  • housekeeperWFS NOT running
  • dbfiller running and connected to Msgd
  • wfsarb NOT running
  • gopt NOT running
[AOeng@soul-dxwfs SOUL]$ wfseng &

starting arbitrator and housekeper process manually from the processes gui.

Temperature Monitoring.

soul_dx_monitoring_20210709.png

board setup.internal2.

devices move.

ocam not live.
  • Power cycle bcu2k.
  • select bin 1x1 cropped.
  • camera live
  • save data: _190130
lamp on
  • image one pupil.
  • _190309
  • center pupil moving bayside stage.
  • modulating 10l/d,200Hz,em1
  • modulator operating.
  • _191219
  • recenter bayside stage.
  • _191329
  • ho dark:20210709_191405
cool down ocam.

ok.

Lost house connection.

reconnect.

Arbitrator was down. Rstart from processes didn;t work. w_stop/start.

aligned image. saved board setup: internalcheck

12:56MST warm up ocam.

9 minutes to have at temperature

POWER OFF

-- JuanCarlosGuerra - 09 Jul 2021
Topic revision: r1 - 09 Jul 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback