Summary.

Checking pupil rotations for all bins, the bin1 and bin3 a shift in positive X directions can be applied.

Soul.latest was deployed and it showed issues during the close loop, the high order modes were not corrected. bin1 was executed using the soul.latest and after it was changed to Soul old version to continue with the pupil rotation check.

With soul old the loop was stable and it didnt; show problems.

Details.

Power ON.

  • Power on ->OK.
  • OCAM bias good.

WfsArbitrator _191870|INF| 38|2021-05-03 16:27:27.213046| WFS > Executing operate...

WfsArbitrator _191870|INF| 39|2021-05-03 16:27:27.213111| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)

2021-05-03 16:33:58.444427| COMMANDHANDLER > Command Operate (code 2004) successfully completed

16:37:11.595640| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = '', outputFile = '/local/aomeas/adsec_data/20210503/Data_20210503_163710/wfs.fits')
  • cancel preset.
  • NOframes to the Adsec. Looking at the mirror controller. Also the loop indicate Close but not corrections applied.
WfsArbitrator _191870|INF| 166|2021-05-03 17:06:44.404589| WFS > Executing presetAO...

WfsArbitrator _191870|INF| 167|2021-05-03 17:06:44.404762| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Stop/start BCU2K
  • NOframes to the Adsec. Looking at the mirror controller. Also the loop indicate Close but not corrections applied.
  • cancel preset.
  • Check pupils soul.latest
drwxrwxr-x. 23 AOeng AOeng 4096 Mar 13 2018 mode1/

drwxrwxr-x. 2 AOeng AOeng 4096 Jun 6 2018 20180605-172546/

drwxrwxr-x. 2 AOeng AOeng 4096 Jun 6 2018 20180605-172620/

-rw-rw-r--. 1 AOeng AOeng 80 Dec 14 2019 pupdata.txt

-rw-rw-r--. 1 AOeng AOeng 5760 Dec 14 2019 pup1.fits

-rw-rw-r--. 1 AOeng AOeng 5760 Dec 14 2019 pup2.fits

-rw-rw-r--. 1 AOeng AOeng 5760 Dec 14 2019 pup3.fits

-rw-rw-r--. 1 AOeng AOeng 5760 Dec 14 2019 pup4.fits

drwxrwxr-x. 7 AOeng AOeng 4096 Feb 10 2020 mode4wrong/

drwxrwxr-x. 19 AOeng AOeng 4096 Sep 16 2020 mode5/

drwxrwxr-x. 7 AOeng AOeng 4096 Jan 6 17:27 m4shifts/

drwxrwxr-x. 30 AOeng AOeng 4096 Jan 27 19:06 mode3/

drwxrwxr-x. 24 AOeng AOeng 4096 Feb 19 22:19 mode4/

drwxrwxr-x. 52 AOeng AOeng 4096 Mar 9 23:04 mode2/

lrwxrwxrwx. 1 AOeng AOeng 69 Apr 28 11:18 current -> /home/AOeng/aoroot/calib/wfs/current/ocam2/LUTs/mode2/20210210_205151/


  • wfs_stop_all.
  • The system close the loop, frames are travelling.
WfsArbitrator _130145|INF| 152|2021-05-03 18:20:27.220379| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Close loop check.->OK

Bin1 Pup Rotation.

  • Shift of one pixel in X direction positive

bin1_may3.png

bin1_may3_puperr.png
  • ADSC_STOP/START
  • WFS_STOP/START
  • STILL SHOWING THAT THERE IS NOT CORRECTIONS FOR HIGHER MODES
WfsArbitrator _171114|INF| 42|2021-05-03 19:05:16.801673| WFS > Executing presetAO...

WfsArbitrator _171114|INF| 43|2021-05-03 19:05:16.801778| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 14, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Discussed with Greg.
  • CHANGE FROM NEW TO OLD SOUL VERSION.
  • THE WFS_INIT IS NOT IN THE BIN
  • higher order seems better corrected.
  • pupil check process was continous restarting.
Arbitrator_99079|INF| 204|2021-05-03 19:45:31.329994| COMMANDHANDLER > Command StopLoopAdsec (code 2033) successfully completed

WfsArbitrator _99079|INF| 205|2021-05-03 19:45:31.544373| MAIN > Camera lens tracking: waiting for data from pupilcheck process

WfsArbitrator _99079|INF| 206|2021-05-03 19:45:34.128979| MAIN > Camera lens tracking: waiting for data from pupilcheck process

WfsArbitrator _99079|INF| 207|2021-05-03 19:45:36.756871| MAIN > Camera lens tracking: waiting for data from pupilcheck process

WfsArbitrator _99079|INF| 208|2021-05-03 19:45:39.357909| MAIN > Camera lens tracking: waiting for data from pupilcheck process
  • wfs_stop_all/start
  • check pupil the process is still reseting.
  • Open loop.

adsec deamon down.

[AOeng@sxadsec ~]$ [AOeng@sxadsec ~]$ adsc_check

-bash: [AOeng@sxadsec: command not found

[AOeng@sxadsec ~]$

[AOeng@sxadsec ~]$ ^C

-bash: :s^C: substitution failed

[AOeng@sxadsec ~]$ Traceback (most recent call last):

-bash: syntax error near unexpected token `most'

[AOeng@sxadsec ~]$ File "/home/aoeng/soul/soul.aoroot/bin/adsc_check", line 46, in <module>

-bash: syntax error near unexpected token `newline'

[AOeng@sxadsec ~]$ test = thAOApp('check')

-bash: syntax error near unexpected token `('

[AOeng@sxadsec ~]$ File "/home/aoeng/soul/soul.aoroot/lib/python/AdOpt/thAOApp.py", line 175, in init

  • restart adsec_stop/start
  • Xianyu called for restarting adsec.
  • restarting adsec and wfs
  • BIN2 PRESET fine.
  • close loop ->ok

BIN2 PUPIL ROTATION.

  • shift one pixel y direction negative..?

bin2_may3.png

bin2_may3_puperr.png

BIN3 PUPIL ROTATION.

  • SHIFT ONE PIXEL UP IN x direction
bin3_may3.png

bin3_may3_puperr.png

BIN4 PUPIL ROTATION.

  • Still fordward direction showing good positions.
bin4_may3.png

bin4_may3_puperr.png

GRAND SLAM

grandslam_may3.png

-- JuanCarlosGuerra - 03 May 2021
I Attachment Action Size Date Who Comment
bin1_may3.pngpng bin1_may3.png manage 317 K 03 May 2021 - 22:11 JuanCarlosGuerra  
bin1_may3_puperr.pngpng bin1_may3_puperr.png manage 552 K 03 May 2021 - 23:26 JuanCarlosGuerra  
bin2_may3.pngpng bin2_may3.png manage 329 K 03 May 2021 - 22:48 JuanCarlosGuerra  
bin2_may3_puperr.pngpng bin2_may3_puperr.png manage 428 K 03 May 2021 - 23:05 JuanCarlosGuerra  
bin3_may3.pngpng bin3_may3.png manage 317 K 03 May 2021 - 23:04 JuanCarlosGuerra  
bin3_may3_puperr.pngpng bin3_may3_puperr.png manage 415 K 03 May 2021 - 23:04 JuanCarlosGuerra  
bin4_may3.pngpng bin4_may3.png manage 330 K 03 May 2021 - 23:54 JuanCarlosGuerra  
bin4_may3_puperr.pngpng bin4_may3_puperr.png manage 476 K 03 May 2021 - 23:55 JuanCarlosGuerra  
grandslam_may3.pngpng grandslam_may3.png manage 114 K 04 May 2021 - 18:32 JuanCarlosGuerra  
Topic revision: r4 - 04 May 2021, JuanCarlosGuerra