Summary

Details.

Power ON

  • Josh started the power on sequence at 2:30.
  • OCAM not live.
  • Power cycle BCU2k two times and the Ocam viewer didn;t show live status. After the binning cycle the ocam viewer shows live status.
  • Discussing with Josh about the troubleshooting.
  • Bias good.
  • saved frames.
WfsArbitrator _117426|INF| 37|2021-04-24 23:45:45.822936| WFS > Executing operate...

WfsArbitrator _117426|INF| 38|2021-04-24 23:45:45.823000| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)

WfsArbitrator _117426|INF| 66|2021-04-25 00:01:18.895410| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = '', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_000117/wfs.fits')
  • frames stddev 7.9
  • frame_20210425_000117.png
2:25 COLLIMATION.
  • AUTHORIZE LUCI.FAILED M3 FAILED.
  • SECOND TRY FAILED.
  • 2:31 HOME AND INITIALIZE M3
  • 2:34 AUTHORIZE.

OSCO

  • Masterdiagnostic Down.
  • OCAMNOT LIVE.
  • Start masterdiagnostic.
WfsArbitrator _117426|INF| 150|2021-04-25 02:43:23.523477| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 151|2021-04-25 02:43:23.523603| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 2.91082, starypos = 2.78393, telElevation = 0, derotAngle = 0, starMag = 7.48, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • pupil check process failed after applying the NCPA.
fsArbitrator_117426|INF| 181|2021-04-25 02:45:55.724458| MAIN > Camera lens tracking: moving by x=-19.6 y=-28.0 (um)

WfsArbitrator _117426|INF| 182|2021-04-25 02:45:57.165831| MAIN > Camera lens tracking: moving by x=-11.5 y=-12.9 (um)

WfsArbitrator _117426|INF| 183|2021-04-25 02:45:58.287256| MAIN > Camera lens tracking: moving by x= -4.9 y= -5.7 (um)

WfsArbitrator _117426|INF| 184|2021-04-25 02:45:59.397879| MAIN > Camera lens tracking: moving by x= -1.9 y= -1.9 (um)

WfsArbitrator _117426|INF| 185|2021-04-25 02:46:00.518764| MAIN > Camera lens tracking: moving by x= -1.8 y= -0.8 (um)

WfsArbitrator _117426|INF| 186|2021-04-25 02:46:15.613975| WFS > Executing stopLoop...

WfsArbitrator _117426|INF| 187|2021-04-25 02:46:15.614114| WFS > Python command: WfsArbScripts.stopLoop(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _117426|INF| 188|2021-04-25 02:46:15.623452| WFS > Command returns success .

WfsArbitrator _117426|INF| 189|2021-04-25 02:46:15.623506| MASTDIAGN > Setting master diagnostic in open loop mode

WfsArbitrator _117426|INF| 190|2021-04-25 02:46:16.475712| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 165.562, zero = 0)

WfsArbitrator _117426|INF| 191|2021-04-25 02:46:26.505118| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 165.619, zero = 0)
  • Shell failure->rest
  • Recovery fail->OK
  • Set the Shell.
  • Not more time for OSCO.
ERROR AFTER SHELL RESTED.:
  • WfsArbitrator _117426|ERR| 201|2021-04-25 02:48:16.758462| MAIN > Cannot update NCPA vector: [AOException] thWaitMsg error(-5001): Timeout error (code -1) Return from search routine (item not found)
  • WfsArbitrator _117426|INF| 202|2021-04-25 02:48:26.768078| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 166.3, zero = 0)
  • WfsArbitrator _117426|ERR| 20

SEQUENCE

fsArbitrator_117426|INF| 168|2021-04-25 02:44:30.407879| WFS > Command returns success : -0.4244941 0.0112588
WfsArbitrator_117426|INF| 169|2021-04-25 02:44:30.407950| COMMANDHANDLER > Command centerStar (code 2038) successfully completed
WfsArbitrator_117426|INF| 170|2021-04-25 02:45:23.620300| MAIN > Received cmd: centerPupils()
WfsArbitrator_117426|INF| 171|2021-04-25 02:45:23.620451| COMMANDHANDLER > FSM (status: AOSet) has received command 2039 (centerPupils)
WfsArbitrator_117426|INF| 172|2021-04-25 02:45:23.620529| WFS > Executing centerPupil...
WfsArbitrator_117426|INF| 173|2021-04-25 02:45:23.620605| WFS > Python command: WfsArbScripts.centerPupils(self, sensor = 'FLAOWFS', instr = 'LUCIFER')
WfsArbitrator_117426|INF| 174|2021-04-25 02:45:44.321843| WFS > Command returns success .
WfsArbitrator_117426|INF| 175|2021-04-25 02:45:44.321927| COMMANDHANDLER > Command centerPupils (code 2039) successfully completed
WfsArbitrator_117426|INF| 176|2021-04-25 02:45:44.322401| WFS > Executing closeLoop...
WfsArbitrator_117426|INF| 177|2021-04-25 02:45:44.322493| WFS > Python command: WfsArbScripts.closeLoop(self, sensor = 'FLAOWFS', instr = 'LUCIFER')
WfsArbitrator_117426|INF| 178|2021-04-25 02:45:44.354995| WFS > Command returns success .
WfsArbitrator_117426|INF| 179|2021-04-25 02:45:44.355049| MASTDIAGN > Setting master diagnostic in closed loop mode
WfsArbitrator_117426|INF| 180|2021-04-25 02:45:45.498281| MAIN > Camera lens tracking: moving by x=-30.3 y=-46.2 (um)
WfsArbitrator_117426|INF| 181|2021-04-25 02:45:55.724458| MAIN > Camera lens tracking: moving by x=-19.6 y=-28.0 (um)
WfsArbitrator_117426|INF| 182|2021-04-25 02:45:57.165831| MAIN > Camera lens tracking: moving by x=-11.5 y=-12.9 (um)
WfsArbitrator_117426|INF| 183|2021-04-25 02:45:58.287256| MAIN > Camera lens tracking: moving by x= -4.9 y= -5.7 (um)
WfsArbitrator_117426|INF| 184|2021-04-25 02:45:59.397879| MAIN > Camera lens tracking: moving by x= -1.9 y= -1.9 (um)
WfsArbitrator_117426|INF| 185|2021-04-25 02:46:00.518764| MAIN > Camera lens tracking: moving by x= -1.8 y= -0.8 (um)
WfsArbitrator_117426|INF| 186|2021-04-25 02:46:15.613975| WFS > Executing stopLoop...
WfsArbitrator_117426|INF| 187|2021-04-25 02:46:15.614114| WFS > Python command: WfsArbScripts.stopLoop(self, sensor = 'FLAOWFS', instr = 'LUCIFER')
WfsArbitrator_117426|INF| 188|2021-04-25 02:46:15.623452| WFS > Command returns success .
WfsArbitrator_117426|INF| 189|2021-04-25 02:46:15.623506| MASTDIAGN > Setting master diagnostic in open loop mode
WfsArbitrator_117426|INF| 190|2021-04-25 02:46:16.475712| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 165.562, zero = 0)
WfsArbitrator_117426|INF| 191|2021-04-25 02:46:26.505118| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 165.619, zero = 0)
WfsArbitrator_117426|INF| 192|2021-04-25 02:46:36.532955| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 165.675, zero = 0)


AdSecArbitrator _23140|INF| 2020|2021-04-25 02:46:15.613836| COMMANDHANDLER > Command TTOffload (code 2029) successfully completed

AdSecArbitrator _23140|INF| 2021|2021-04-25 02:46:16.630033| MAIN > Received cmd: StopAo (1)

COSMO

7:04 DIMM 1.1".

08:10 Authorize Luci1

AO

First object is a try to close the loop in a galaxy. core concentrated but a bit elliptical, estimated magnitude is 14.7 r, but forced to go to Bin4 16.51 from script.

8:17 dimm1.1"

WfsArbitrator _117426|INF| 4206|2021-04-25 08:22:44.619464| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 4207|2021-04-25 08:22:44.619591| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 48.8539, starypos = -0.805501, telElevation = 0, derotAngle = 0, starMag = 14.72, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Auxilliar loops close
  • magnitude 17.2",372Hz
  • B4,gtt=2.7,ho1=1.21,ho2=0.1
  • gopt gain->0.03,0.01,0
  • slope rms->44
ython_127007 |INF| 50|2021-04-25 08:34:37.097764| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.32e-08/2.62e-08 gopt: 0.012

WfsArbitrator _117426|INF| 4379|2021-04-25 08:37:13.002210| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_083710/wfs.fits')

WfsArbitrator _117426|INF| 4380|2021-04-25 08:37:13.027689| WFS > Command returns success .

WfsArbitrator _117426|INF| 4381|2021-04-25 08:37:13.028428| WFS > Executing savePsf...

WfsArbitrator _117426|INF| 4382|2021-04-25 08:37:13.028532| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210425/Data_20210425_083710/psf.fits', num = 10)

WfsArbitrator _117426|INF| 4383|2021-04-25 08:37:13.719329| WFS > Command returns success .

10425/Data_20210425_083906/wfs.fits')

WfsArbitrator _117426|INF| 4406|2021-04-25 08:39:08.028423| WFS > Command returns success .

WfsArbitrator _117426|INF| 4407|2021-04-25 08:39:08.029196| WFS > Executing savePsf...

WfsArbitrator _117426|INF| 4408|2021-04-25 08:39:08.029301| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210425/Data_20210425_083906/psf.fits', num = 10)
  • 8:43 dimm 1"
WfsArbitrator _117426|INF| 4482|2021-04-25 08:45:37.002257| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_084535/wfs.fits')

WfsArbitrator _117426|INF| 4483|2021-04-25 08:45:37.037970| WFS > Command returns success .

WfsArbitrator _117426|INF| 4484|2021-04-25 08:45:37.038786| WFS > Executing savePsf...

WfsArbitrator _117426|INF| 4485|2021-04-25 08:45:37.038925| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210425/Data_20210425_084535/psf.fits', num = 10)

python_127007 |INF| 133|2021-04-25 08:45:47.697396| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 2.59e-08/1.86e-08 gopt: 0.025

python_127007 |INF| 134|2021-04-25 08:45:53.883294| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 8.33e-09/1.83e-08 gopt: 0.021

WfsArbitrator _117426|INF| 4610|2021-04-25 08:58:28.002486| WFS > Executing saveStatus...

WfsArbitrator _117426|INF| 4611|2021-04-25 08:58:28.002562| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_085826/wfs.fits')

8:58 DIMM 1"

Tracknum: 20210425_250000 demodDelta/demodCmd: 2.49e-08/1.52e-08 gopt: 0.023

python_127007 |INF| 368|2021-04-25 09:16:51.832015| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 8.45e-09/1.56e-08 gopt: 0.020

WfsArbitrator _117426|INF| 4856|2021-04-25 09:22:10.148732| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 4857|2021-04-25 09:22:10.148841| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.2091, starypos = -2.32196, telElevation = 0, derotAngle = 0, starMag = 15.42, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

  • preset cancel. Pupil position at the limit rot 280.
  • change rotation to 0.
WfsArbitrator _117426|INF| 5217|2021-04-25 09:35:22.993114| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 5218|2021-04-25 09:35:22.993220| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.226, starypos = -2.31588, telElevation = 0, derotAngle = 0, starMag = 15.42, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator_117426|ERR| 5308|2021-04-25 09:36:44.473612| MAIN > Internal rotator position 389.7 is out of the allowed range

WfsArbitrator_117426|ERR| 5309|2021-04-25 09:36:45.474458| MAIN > Internal rotator position 389.7 is out of the allowed range
  • Changing to bin 4 from luci script.
WfsArbitrator _117426|INF| 5442|2021-04-25 09:38:54.266751| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 5443|2021-04-25 09:38:54.266861| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.2306, starypos = -2.31413, telElevation = 0, derotAngle = 0, starMag = 16.51, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Back rotator position 280
ator_117426|INF| 5519|2021-04-25 09:42:00.559712| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 5520|2021-04-25 09:42:00.559824| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.2347, starypos = -2.31253, telElevation = 0, derotAngle = 0, starMag = 16.51, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • cancel preset.
  • the check flux set back to bin3.
  • almost in a good pupil stage position two tries for close loop and cancel.
WfsArbitrator _117426|INF| 5889|2021-04-25 09:51:55.080541| WFS > Executing presetAO...

WfsArbitrator _117426|INF| 5890|2021-04-25 09:51:55.080648| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.248, starypos = -2.30713, telElevation = 0, derotAngle = 0, starMag = 16.51, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • skip check flux.
  • close all the loops.
  • bin4,491
WfsArbitrator _117426|INF| 5976|2021-04-25 09:58:16.001763| WFS > Executing saveStatus...

WfsArbitrator _117426|INF| 5977|2021-04-25 09:58:16.001916| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_095814/wfs.fits')

WfsArbitrator _117426|INF| 5978|2021-04-25 09:58:16.035823| WFS > Command returns success .

WfsArbitrator _117426|INF| 5979|2021-04-25 09:58:16.036704| WFS > Executing savePsf...

WfsArbitrator _117426|INF| 5980|2021-04-25 09:58:16.036799| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210425/Data_20210425_095814/psf.fits', num = 10)

INF| 49|2021-04-25 09:58:54.548099| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 4.28e-08/3.72e-08 gopt: 0.144
  • 9:59 dimm 1"
WfsArbitrator _117426|INF| 6025|2021-04-25 10:03:28.641584| OPTLOOPDIAG > Executing save...

WfsArbitrator _117426|INF| 6026|2021-04-25 10:03:30.003418| WFS > Executing saveStatus...

WfsArbitrator _117426|INF| 6027|2021-04-25 10:03:30.003513| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_100328/wfs.fits')

python_152112 |INF| 50|2021-04-25 09:58:59.599219| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.16e-08/3.44e-08 gopt: 0.141

python_152112 |INF| 161|2021-04-25 10:16:53.937414| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.28e-08/3.49e-08 gopt: 0.178

python_152112 |INF| 162|2021-04-25 10:17:15.211673| MAIN > Closed loop iterations=1, waiting for the next one

WfsArbitrator _117426|INF| 6169|2021-04-25 10:17:41.003745| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_101739/wfs.fits')
  • 10:38 SR plot from AO telemetry
sr_20210425_1038.png

6259|2021-04-25 10:26:41.003094| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_102639/wfs.fits')
  • Shell state failure.
  • ocam2k went not live
  • Shell rest.
  • recover fail
  • AOloop was still close.
  • Stop loop
  • ocam live again. No action taken to have ocam2k live again


MirrorCtrl_64536 |ERR| 1380|2021-04-25 10:21:46.680094| MAIN > Received BCU command: bcu 5-5, dsp 255-255, opcode 10, address 0, datasize 8, shmid 1515487645, flags 1
MirrorCtrl_64536 |ERR| 1381|2021-04-25 10:29:28.948572| MAIN > ****** FAILED PACKET FROM BCU 2 - CMD TIMEOUT CA ******
MirrorCtrl_64536 |ERR| 1382|2021-04-25 10:29:28.948650| BCUCOMMANDHANDLER_2_P3 > BcuCommandHandler::handleBcuReply: REQUEST COMPLETED but FAILED :-(
MirrorCtrl_64536 |ERR| 1383|2021-04-25 10:29:29.978358| MAIN > *******
FAILED PACKET FROM BCU 2 - CMD TIMEOUT CA *****


fastdiagn_64745 |WAR| 985156|2021-04-25 10:29:30.239306| MAIN > Timeout - Will emergencyReact next loop [FastDiagnostic.cpp:1090]
fastdiagn_64745 |WAR| 985157|2021-04-25 10:29:32.240095| MAIN > Timeout - Will emergencyReact now [FastDiagnostic.cpp:1085]
fastdiagn_64745 |ERR| 985158|2021-04-25 10:29:32.240114| FUNCTEMERGENCYSTOP > emergencyReact = 1 [Funct.cpp:95]
fastdiagn_64745 |INF| 985159|2021-04-25 10:29:32.240150| ADAM-MODBUS > AdamModbus: disabling coils...


python_79954 |ERR| 1520|2021-04-25 10:27:44.573482| MAIN > [AOException] RETRY: Offset outside FoV limits. Target xyz is 4.18 -82.90 52.90 (-5002) VALUE_OUT_OF_RANGE_ERROR
python_79954 |INF| 1521|2021-04-25 10:29:13.495180| MAIN > Request: Pause()
python_79954 |DEB| 1522|2021-04-25 10:29:13.495236| MAIN > Status after command: LoopSuspended
python_79954 |INF| 1523|2021-04-25 10:29:16.255621| MAIN > Request: OffsetXY((-0.10640328377485275, -0.39264097809791565))
python_79954 |DEB| 1524|2021-04-25 10:29:19.464978| MAIN > Status after command: LoopSuspended
python_79954 |INF| 1525|2021-04-25 10:29:28.144952| MAIN > Request: Resume()
python_79954 |DEB| 1526|2021-04-25 10:29:28.241452| MAIN > Status after command: LoopClosed
python_79954 |INF| 1527|2021-04-25 10:34:12.741920| MAIN > Request: Stop(Starting Preset)


AdSecArbitrator_23140|INF| 2941|2021-04-25 10:29:28.241176| COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_23140|ERR| 2942|2021-04-25 10:29:32.256842| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_23140|ERR| 2943|2021-04-25 10:29:32.256890| ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L


MasterDiagnostic_64594|INF| 8713|2021-04-25 10:29:32.269026| MAIN > OFramesHistory::dumpToFile: output file /local/aolog/2021/04/25/ADAPTIVE-SECONDARY_00001619346572_20210425_102932.bin
MasterDiagnostic_64594|INF| 8714|2021-04-25 10:29:32.269215| MAIN > Dumping 1st part of size 6883680...
MasterDiagnostic_64594|INF| 8715|2021-04-25 10:29:32.273634| MAIN > Dumping 2nd part of size 50480320...
MasterDiagnostic_64594|INF| 8716|2021-04-25 10:29:32.312022| ALERT-NOTIFIER > ArbitratorInterface to adsecarb.L succesfully created!
MasterDiagnostic_64594|INF| 8717|2021-04-25 10:29:32.312108| MAIN > Alert for dump to adsec arbitrator successfully sent
MasterDiagnostic_64594|INF| 8718|2021-04-25 10:31:15.073997| MAIN > OFramesHistory::dumpToFile: output file /local/aolog/2021/04/25/ADAPTIVE-SECONDARY_00001619346675_20210425_103115.bin
MasterDiagnostic_64594|INF| 8719|2021-04-25 10:31:15.074172| MAIN > Dumping 1st part of size 57364000...
MasterDiagnostic_64594|INF| 8720|2021-04-25 10:31:15.110005| MAIN > Dumping 2nd part of size 0...
MasterDiagnostic_64594|INF| 8721|2021-04-25 10:31:15.115642| ALERT-NOTIFIER > ArbitratorInterface to adsecarb.L succesfully created!
MasterDiagnostic_64594|INF| 8722|2021-04-25 10:31:15.115754| MAIN > Alert for dump to adsec arbitrator successfully sent
MasterDiagnostic_64594|INF| 8723|2021-04-25 10:33:33.915558| MAIN > OFramesHistory::dumpToFile: output file /local/aolog/2021/04/25/ADAPTIVE-SECONDARY_00001619346813_20210425_103333.bin
MasterDiagnostic_64594|INF| 8724|2021-04-25 10:33:33.915799| MAIN > Dumping 1st part of size 24953340...
MasterDiagnostic_64594|INF| 8725|2021-04-25 10:33:33.932427| MAIN > Dumping 2nd part of size 32410660...
MasterDiagnostic_64594|INF| 8726|2021-04-25 10:33:33.957903| ALERT-NOTIFIER > ArbitratorInterface to adsecarb.L succesfully created!
MasterDiagnostic_64594|INF| 8727|2021-04-25 10:33:33.958031| MAIN > Alert for dump to adsec arbitrator successfully sent



fsArbitrator_117426|INF| 6299|2021-04-25 10:34:21.754010| WFS > Executing presetAO...

WfsArbitrator_117426|INF| 6300|2021-04-25 10:34:21.754125| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 55.3113, starypos = -2.27746, telElevation = 0, derotAngle = 0, starMag = 16.51, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • all loop close correctly.
  • done-> 6363|2021-04-25 10:38:48.759141| WFS > Command returns success
10:40:47.475097| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.79e-08/3.71e-08 gopt: 0.138

python_191325 |INF| 38|2021-04-25 10:40:52.518088| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.16e-08/3.55e-08 gopt: 0.133
  • 10:41 seeing
WfsArbitrator_117426|INF| 6399|2021-04-25 10:41:42.003727| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_104140/wfs.fits')
  • 10:55 dimm 0.9"
  • preset PSF star
WfsArbitrator_117426|INF| 6502|2021-04-25 10:55:17.306541| WFS > Executing presetAO...

WfsArbitrator_117426|INF| 6503|2021-04-25 10:55:17.306659| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = 46.8154, starypos = -16.3228, telElevation = 0, derotAngle = 0, starMag = 16.51, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • close all loops.
  • seeing 0.96"
  • TT 0.18,0.13,0.02
  • slope rms 5
WfsArbitrator_117426|INF| 6589|2021-04-25 11:00:49.001615| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_110047/wfs.fits')

32|2021-04-25 11:01:18.122136| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.59e-08/3.6e-08 gopt: 0.134

python_117149 |INF| 33|2021-04-25 11:01:23.358874| MAIN > Tracknum: 20210425_250000 demodDelta/demodCmd: 3.21e-08/3.51e-08 gopt: 0.130

sArbitrator_117426|INF| 6641|2021-04-25 11:03:25.003272| WFS > Executing saveStatus...

WfsArbitrator_117426|INF| 6642|2021-04-25 11:03:25.003362| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210425/Data_20210425_110323/wfs.fits')

STOP

11:30:24.121995| WFS > Executing stopLoop...

11:30 WARM UP OCAM

tor_117426|INF| 7144|2021-04-25 11:52:16.958386| WFS > Executing off...

WfsArbitrator_117426|INF| 7145|2021-04-25 11:52:16.958453| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator_117426|INF| 7146|2021-04-25 11:52:46.585038| WFS > Command returns success .

WfsArbitrator_117426|INF| 7147|2021-04-25 11:52:46.585241| COMMANDHANDLER > Command Off (code 2005) successfully completed

WfsArbitrator_117426|FAT| 7148|2021-04-25 11:52:51.527027| MAIN > wfsarb.L terminated by idle

-- JuanCarlosGuerra - 25 Apr 2021
Topic revision: r4 - 25 Apr 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback