Summary.

Details.

  • Power ON WFS.
  • Launched the wfs control gui
  • The buttons are not available ("not click"
  • Stop/start wfsarbitrator.
  • Buttons available.
  • Ocam bias->OK
  • Acquire a bias
  • 20210417_155742
Note: the CCD starts with a saturation level of 45100 photons/subap. Chamber lights off.

  • ocam cool.
  • frames 20210417_161028
  • restore binternal2
  • HO dark: 20210417_161344.fits
  • counts starts to going down to 20 phtons/subap
  • _161637
  • Bin1
  • 600
  • 6l/d
  • EMG 1
  • _frame: _161245
  • RFBG power ON.
  • fault when power on.
  • ERROR: cable chain did not become ready in 4 seconds.
  • call mountain.
  • Restore the RFBG.
Running internal pupil position acqusition.

Moving rotator

script wobble_test showed an error.

[AOeng@soul-dxwfs scripts]$ wobble_test.py

ssh LBTO@obs3 "echo \"/lbt/irtc/current/bin/IRCR RotServicePosition 360.000000 right\" | tcsh"

----------------------------------------------

Resmsg[0] = RotServicePosition result status: Error

Resmsg[1] = ERROR rotator is not movable (probably because of cable chain).

Exit 1

Exit 1

Pupil at rotator 210 aligned, but the camera lens in y directions is all the way in negative position about 3.5 pixels differences.

Power Off

-- JuanCarlosGuerra - 17 Apr 2021
Topic revision: r2 - 17 Apr 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.