PUPIL ROTATION CHECK. RUN THE GRANS SLAM.

The pupils positions shows a slight saturation for bin1 and Bin2 moving badkward (440->0) they are good in forward movements (0->400), bin3 is showing a little saturation point in negative Y at 250 (moving backward), but the Y pupils erros are still inside the threshold. A shift positive will move the saturation to positive direction increasing the rotator angle where there is saturation., the movements in forward direction is good. To be follow up.

Bin4 is at midrange and it is good.

The temperature is 3.7 degrees celcius.

Details.

WfsArbitrator _17251|INF| 36|2021-04-16 16:19:05.835244| OPTLOOPDIAG > Executing enableAntiDrift(true)...

WfsArbitrator _17251|INF| 37|2021-04-16 16:19:05.835444| WFS > Executing operate...

WfsArbitrator _17251|INF| 38|2021-04-16 16:19:05.835514| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)
  • Ocam good bias.
  • Acquire bias frames.
WfsArbitrator _17251|INF| 50|2021-04-16 16:28:28.800040| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = '', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_162827/wfs.fits')
  • CCD47 dark
WfsArbitrator _17251|INF| 55|2021-04-16 16:46:12.217640| WFS > Executing calibrateHODark...

WfsArbitrator _17251|INF| 56|2021-04-16 16:46:12.217705| WFS > Python command: WfsArbScripts.calibHODark(self, sensor = 'FLAOWFS', instr = '', num = 100)
  • ocam dropping the counts..to 7phto/subap
  • ocam dark
WfsArbitrator _17251|INF| 68|2021-04-16 16:50:44.665512| WFS > Executing presetAO...

WfsArbitrator _17251|INF| 69|2021-04-16 16:50:44.665620| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Hexapod.
  • 2.8,3.0,-1,-130,-5
  • close loop save data.
WfsArbitrator _17251|INF| 180|2021-04-16 17:02:24.002744| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_170222/wfs.fits')

WfsArbitrator _17251|INF| 303|2021-04-16 17:09:25.001857| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_170923/wfs.fits')

BIN1 PUPIL ROTATION.

  • GOOD POSTIVE DiRECTIONS.
  • SATURATION FROM 40 TO 0 and 410 to 380

BIN1_APRIL16.png

BIN2 PUPIL ROTATION

  • shutter 58
  • mag 13.3
  • NEGATIVE DIRECTION 380 camera lens in satiration but the pupil error is still inside of the threshold.20 to 0 a saturaion and errors just above the threshold.
WfsArbitrator _17251|INF| 1486|2021-04-16 17:42:10.369085| WFS > Executing presetAO...

WfsArbitrator _17251|INF| 1487|2021-04-16 17:42:10.369213| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 13.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • CLOSE LOOP ACQUIRE DATA.
WfsArbitrator _17251|INF| 1551|2021-04-16 17:46:22.002068| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_174620/wfs.fits')
  • FIRST ROT DATA.
WfsArbitrator _17251|INF| 1556|2021-04-16 17:47:10.001480| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_174708/wfs.fits')

BIN2_APRIL16.png
  • backward error
bin2_april16_err.png

BIN3 PUPIL ROTATION

  • SHUTTER 50
  • MAG 14.5.
  • saturation in position backward at 280 to 250 but pupils errors inside threshold with some spikes.
bin3_april16.png

bin3_april16_err.png

BIN4 PUPIL ROTATION.

  • SHUTTER:46
  • MAG: 16.3
  • Pupil in midrange. Good in both directions.

WfsArbitrator _17251|INF| 3004|2021-04-16 18:57:20.389102| WFS > Executing presetAO...

WfsArbitrator _17251|INF| 3005|2021-04-16 18:57:20.389206| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 17, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

bin4_april1.png

Handover to LBTI 12:40 (MST)

POWER OFF

WfsArbitrator _17251|INF| 3751|2021-04-16 20:17:13.895165| OPTLOOPDIAG > Executing enableAntiDrift(false)...

WfsArbitrator _17251|INF| 3752|2021-04-16 20:17:13.895329| WFS > Executing off...

WfsArbitrator _17251|INF| 3753|2021-04-16 20:17:13.895376| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

GRAND SLAM

april16_grandslam_0.png


stop/start processes

POWERON.

W_check-> showed all the process up anc connected to msgd
  • masterdiagnotic process showed to init (yellow)
  • ocam viewer NOTLIVE.
  • Start process
  • Ocam viewer LIVE
  • Check Ocam bias ->OK
WfsArbitrator _193589|INF| 51|2021-04-16 20:36:14.004928| WFS > Executing saveStatus...

WfsArbitrator _193589|INF| 52|2021-04-16 20:36:14.005014| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = '', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_203612/wfs.fits')

frames_20210416_202612.png

POWER OFF

WfsArbitrator _193589|INF| 56|2021-04-16 20:37:52.970164| WFS > Executing off...

WfsArbitrator _193589|INF| 57|2021-04-16 20:37:52.970239| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = '')

WfsArbitrator _193589|INF| 58|2021-04-16 20:39:23.824416| WFS > Command returns success .

wobble_test_backward/forward.py

  • script requires a non LNGain gain vector file. Scripts error: raise e

IOError: [Errno 2] No such file or directory: '/local/aomeas/adsec_calib/M2C/p/LNgains.fits'

[AOeng@soul-sxwfs scripts]$ wobble_test_backward.py

Traceback (most recent call last):

File "/home/AOeng/aoroot/bin/wobble_test_backward.py", line 15, in <module>

g = adsec.get_gains()

File "/home/AOeng/aoroot/lib/python/AdOpt/asm.py", line 71, in get_gains

return fits_lib.readSimpleFits(gainpath)

File "/home/AOeng/aoroot/lib/python/AdOpt/fits_lib.py", line 87, in readSimpleFits

fitsobj = pyfits.open(filename, ignore_missing_end=True)

File "/usr/lib64/python2.6/site-packages/pyfits/core.py", line 10729, in open

ffo = _File(name, mode=mode, memmap=memmap, **parms)

File "/usr/lib64/python2.6/site-packages/pyfits/core.py", line 9268, in __init__

raise e

IOError: [Errno 2] No such file or directory: '/local/aomeas/adsec_calib/M2C/p/LNgains.fits'

  • OCAM SATURATION
WfsArbitrator _17251|INF| 3704|2021-04-16 19:52:09.267308| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_195208/wfs.fits')

  • background taken.
WfsArbitrator _17251|INF| 3742|2021-04-16 20:14:24.805129| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210416/Data_20210416_201423/wfs.fits')

-- JuanCarlosGuerra - 16 Apr 2021
I Attachment Action Size Date Who Comment
BIN1_APRIL16.pngpng BIN1_APRIL16.png manage 77 K 16 Apr 2021 - 17:46 JuanCarlosGuerra  
BIN2_APRIL16.pngpng BIN2_APRIL16.png manage 73 K 16 Apr 2021 - 18:45 JuanCarlosGuerra  
april16_grandslam.pngpng april16_grandslam.png manage 90 K 16 Apr 2021 - 20:46 JuanCarlosGuerra  
april16_grandslam_0.pngpng april16_grandslam_0.png manage 130 K 16 Apr 2021 - 20:52 JuanCarlosGuerra  
bin2_april16_err.pngpng bin2_april16_err.png manage 82 K 16 Apr 2021 - 19:09 JuanCarlosGuerra  
bin3_april16.pngpng bin3_april16.png manage 70 K 16 Apr 2021 - 19:14 JuanCarlosGuerra  
bin3_april16_err.pngpng bin3_april16_err.png manage 82 K 16 Apr 2021 - 19:15 JuanCarlosGuerra  
bin4_april1.pngpng bin4_april1.png manage 74 K 16 Apr 2021 - 20:19 JuanCarlosGuerra  
frames_20210416_202612.pngpng frames_20210416_202612.png manage 274 K 16 Apr 2021 - 21:06 JuanCarlosGuerra  
Topic revision: r1 - 16 Apr 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback