Summary.

Details

WFS_INIT

startProcessByName(): starting process /home/AOeng/aoroot/bin/pupilcheck.py -i pupilcheckocam2 ocam2
-- Stage masterdiagnostic housekeeperWFS dbfiller
startProcessByName(): starting process /home/AOeng/aoroot/bin/MasterDiagnostic -i masterdiagnostic
startProcessByName(): starting process /home/AOeng/aoroot/bin/housekeeperWFS -i housekeeperWFS
startProcessByName(): starting process /home/AOeng/aoroot/bin/DBFiller -i dbfiller
Traceback (most recent call last):
File "/home/AOeng/aoroot/bin/w_start", line 256, in <module>
if name == '__main__': main()
File "/home/AOeng/aoroot/bin/w_start", line 144, in main
processControl.waitProcessStart( app, process)
File "/home/AOeng/aoroot/lib/python/AdOpt/processControl.py", line 243, in waitProcessStart
raise AdOptError ( AOErrCode ['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)
AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR
keepUp start/running, process 149691
nohup: appending output to `nohup.out'nohup:
appending output to `nohup.out'
[AOeng@soul-sxwfs SOUL]$ Warning: cannot add process irtcccd
QLayout "unnamed" added to QMainWindow "unnamed", which already has a

wfs_init_april7.png
  • SECOND EXECUTION OF THE COMMAND WFS_INIT same error
  • w_check
  • [AOeng@soul-sxwfs SOUL]$ w_check

    msgdrtdb running
    mirrorctrl running and connected to Msgd
    pinger running and connected to Msgd
    powerboard running and connected to Msgd
    slopecompctrl running and connected to Msgd
    techviewctrl running and connected to Msgd
    ocam2 running and connected to Msgd
    ccd47 running and connected to Msgd
    ttctrl running and connected to Msgd
    adc1 running and connected to Msgd
    adc2 running and connected to Msgd
    rerotator running and connected to Msgd
    filterwheel1 running and connected to Msgd
    filterwheel2 running and connected to Msgd
    mercury running and connected to Msgd
    cubestage running and connected to Msgd
    baysidex running and connected to Msgd
    baysidey running and connected to Msgd
    baysidez running and connected to Msgd
    cameralensctrl running and connected to Msgd
    picctrl running and connected to Msgd
    optloopdiag running and connected to Msgd
    pupilcheckocam2 running and connected to Msgd
    masterdiagnostic running and connected to Msgd
    housekeeperWFS NOT running
    dbfiller running and connected to Msgd
    wfsarb NOT running
    od_gopt NOT running
    gopt NOT running

  • w_stop_all/_start_all
  • [AOeng@soul-sxwfs SOUL]$ w_check

    msgdrtdb running
    mirrorctrl running and connected to Msgd
    pinger running and connected to Msgd
    powerboard running and connected to Msgd
    slopecompctrl running and connected to Msgd
    techviewctrl running and connected to Msgd
    ocam2 running and connected to Msgd
    ccd47 running and connected to Msgd
    ttctrl running and connected to Msgd
    adc1 running and connected to Msgd
    adc2 running and connected to Msgd
    rerotator running and connected to Msgd
    filterwheel1 running and connected to Msgd
    filterwheel2 running and connected to Msgd
    mercury running and connected to Msgd
    cubestage running and connected to Msgd
    baysidex running and connected to Msgd
    baysidey running and connected to Msgd
    baysidez running and connected to Msgd
    cameralensctrl running and connected to Msgd
    picctrl running and connected to Msgd
    optloopdiag running and connected to Msgd
    pupilcheckocam2 running and connected to Msgd
    masterdiagnostic running and connected to Msgd
    housekeeperWFS running and connected to Msgd
    dbfiller running and connected to Msgd
    wfsarb running and connected to Msgd
    od_gopt running and connected to Msgd
    gopt running and connected to Msgd

POWER ON

WfsArbitrator _157194|INF| 37|2021-04-07 19:48:54.768751| WFS > Executing operate...

WfsArbitrator _157194|INF| 38|2021-04-07 19:48:54.768817| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)

WfsArbitrator _157194|INF| 39|2021-04-07 19:49:15.299789| MAIN > Sensor mode: FLAO
  • BIAS OK
  • BIAS FRAMES
WfsArbitrator _157194|INF| 48|2021-04-07 19:57:16.911372| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 49|2021-04-07 19:57:17.997986| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 50|2021-04-07 19:57:17.998077| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = '', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_195716/wfs.fits')

WfsArbitrator _157194|INF| 51|2021-04-07 19:57:18.037012| WFS > Command returns success .

WfsArbitrator _157194|INF| 61|2021-04-07 20:07:55.962401| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 62|2021-04-07 20:07:55.962530| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 63|2021-04-07 20:07:55.963362| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 64|2021-04-07 20:07:55.963401| MAIN > LUT Applied: deg deg V V mm mm V V V V
  • PRESET COMPLETE BUT THE AOSGUI STILL SHOWING "RUNNING" AND THE ADSEC SHOWS LBTI FOCAL STATION.
  • REST THE SHELL
Wed Apr 7 20:11:47.254 2021 aos.info - Updated variable adamhousekeeper.L.COIL_STATUS: 0
Wed Apr 7 20:11:50.558 2021 aos.info - FLAO: No misfunctioning capacitive sensors found.
Wed Apr 7 20:11:50.579 2021 aos.info - FLAO: Adsec 672b-left SAFE position reached.
Wed Apr 7 20:12:10.120 2021 aos.info - Updated variable AOARB.L.AO_READY: 1
Wed Apr 7 20:12:10.120 2021 aos.PresetAO{FLAO}.complete - PresetAO {FLAO}[1587] - dx:nan dy:nan slNull: f1:0.0 f2:1.0 freq:1700 mag:0 nBins:1 nModes:501 r0:nan snMode:0 strehl:0 ttMod:0

WfsArbitrator _157194|INF| 76|2021-04-07 20:14:19.575627| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 0, zero = 1)

WfsArbitrator _157194|INF| 77|2021-04-07 20:14:19.576976| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 78|2021-04-07 20:14:19.577086| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 79|2021-04-07 20:14:19.578037| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 80|2021-04-07 20:14:19.578080| MAIN > LUT Applied: deg deg V V mm mm V V V V

WfsArbitrator _157194|INF| 81|2021-04-07 20:14:20.263343| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 82|2021-04-07 20:14:20.263391| MAIN > LUT Applied: deg deg V V mm mm V V V V

sArbitrator_157194|INF| 206|2021-04-07 20:28:28.190770| WFS > Executing startGopt...

WfsArbitrator _157194|INF| 207|2021-04-07 20:28:28.190865| WFS > Python command: WfsArbScripts.startGopt(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _157194|INF| 208|2021-04-07 20:28:28.815857| MAIN > Camera lens tracking: moving by x= -1.1 y= -0.1 (um)

WfsArbitrator _157194|INF| 213|2021-04-07 20:31:15.699599| ASYNCCOMMANDHANDLER > COMMAND received from optloopdiagngui.L

WfsArbitrator _157194|INF| 214|2021-04-07 20:31:15.699976| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 215|2021-04-07 20:31:17.002218| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 216|2021-04-07 20:31:17.002326| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_203115/wfs.fits')

WfsArbitrator _157194|INF| 217|2021-04-07 20:31:17.045779| WFS > Command returns success .

WfsArbitrator _157194|INF| 220|2021-04-07 20:32:32.296885| ASYNCCOMMANDHANDLER > COMMAND received from optloopdiagngui.L

WfsArbitrator _157194|INF| 221|2021-04-07 20:32:32.297141| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 222|2021-04-07 20:32:34.001972| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 223|2021-04-07 20:32:34.002045| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_203232/wfs.fits')

WfsArbitrator _157194|INF| 224|2021-04-07 20:32:34.045717| WFS > Command returns success .

WfsArbitrator _157194|INF| 225|2021-04-07 20:32:34.046720| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 226|2021-04-07 20:32:34.046875| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_203232/luci.fits', num = 10)

WfsArbitrator _157194|INF| 236|2021-04-07 20:35:01.002472| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 237|2021-04-07 20:35:01.002549| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_203459/wfs.fits')

WfsArbitrator _157194|INF| 238|2021-04-07 20:35:01.140353| WFS > Command returns success .

WfsArbitrator _157194|INF| 239|2021-04-07 20:35:01.141411| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 240|2021-04-07 20:35:01.141507| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_203459/luci.fits', num = 10)
  • luci H
  • 2.75sec
WfsArbitrator _157194|INF| 244|2021-04-07 20:37:15.004178| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 245|2021-04-07 20:37:15.004277| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_203713/wfs.fits')

WfsArbitrator _157194|INF| 246|2021-04-07 20:37:15.041813| WFS > Command returns success .

WfsArbitrator _157194|INF| 247|2021-04-07 20:37:15.042576| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 248|2021-04-07 20:37:15.042661| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_203713/luci.fits', num = 10)

WfsArbitrator _157194|INF| 257|2021-04-07 20:39:27.002231| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_203925/wfs.fits')

WfsArbitrator _157194|INF| 258|2021-04-07 20:39:27.038345| WFS > Command returns success .

WfsArbitrator _157194|INF| 259|2021-04-07 20:39:27.039019| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 260|2021-04-07 20:39:27.039093| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_203925/luci.fits', num = 10)

WfsArbitrator _157194|INF| 261|2021-04-07 20:39:32.797173| MAIN > Camera lens tracking: moving by x= 1.1 y= -0.4 (um)
  • ncpa
  • optimize gain
  • gopt
  • optimize gain...
  • skip frames
  • optmize gain failed
WfsArbitrator _157194|INF| 308|2021-04-07 20:42:52.659538| WFS > Executing optimizeGain...

WfsArbitrator _157194|INF| 309|2021-04-07 20:42:52.659589| WFS > Python command: WfsArbScripts.optimizeGain(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _157194|ERR| 310|2021-04-07 20:42:56.839929| WFS > Command returns error : Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE

WfsArbitrator _157194|INF| 311|2021-04-07 20:42:56.840018| COMMANDHANDLER > Command execution failed: Fsm forced to Failure

WfsArbitrator _157194|INF| 312|2021-04-07 20:42:56.840068| WFS > Executing stopLoop...

WfsArbitrator _157194|INF| 313|2021-04-07 20:42:56.840120| WFS > Python command: WfsArbScripts.stopLoop(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _157194|INF| 314|2021-04-07 20:42:56.844218| WFS > Command returns success .

WfsArbitrator _157194|INF| 315|2021-04-07 20:42:56.844268| MASTDIAGN > Setting master diagnostic in open loop mode

WfsArbitrator _157194|INF| 316|2021-04-07 20:42:56.844445| ADSEC > Executing openLoop...

WfsArbitrator _157194|INF| 317|2021-04-07 20:42:56.844471| ADSEC > Requesting StopAo to AdSecArbitrator...

WfsArbitrator _157194|INF| 318|2021-04-07 20:42:58.201519| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.002, zero = 0)

WfsArbitrator _157194|ERR| 319|2021-04-07 20:43:01.443666| MAIN > OptimizeGain: FATAL - Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE

WfsArbitrator _157194|INF| 320|2021-04-07 20:43:02.293822| COMMANDHANDLER > FSM (status: AOSet) has received command 2023 (EnableDisturb)

WfsArbitrator _157194|INF| 321|2021-04-07 20:43:02.293897| WFS > Executing EnableDisturb...

  • ncpa off
  • optimize gain ok.
  • gopt ok
WfsArbitrator _157194|INF| 384|2021-04-07 20:52:08.373879| OPTLOOPDIAG > Executing enableAntiDrift(true)...

WfsArbitrator _157194|INF| 385|2021-04-07 20:52:08.374020| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 0, zero = 1)

WfsArbitrator _157194|INF| 386|2021-04-07 20:52:08.384805| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 387|2021-04-07 20:52:08.384943| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 388|2021-04-07 20:52:08.385860| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 389|2021-04-07 20:52:08.385903| MAIN > LUT Applied: deg deg V V mm mm V V V V

WfsArbitrator _157194|INF| 390|2021-04-07 20:52:09.217367| MAIN > LUT Applied: Rotang El

  • disturbace s1.6
WfsArbitrator _157194|INF| 462|2021-04-07 20:58:37.601047| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 463|2021-04-07 20:58:39.002810| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 464|2021-04-07 20:58:39.002914| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_205837/wfs.fits')

WfsArbitrator _157194|INF| 465|2021-04-07 20:58:39.041444| WFS > Command returns success .

WfsArbitrator _157194|INF| 466|2021-04-07 20:58:39.042251| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 467|2021-04-07 20:58:39.042449| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_205837/luci.fits', num = 10)

WfsArbitrator _157194|INF| 471|2021-04-07 20:59:55.012049| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 472|2021-04-07 20:59:55.012121| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_205953/wfs.fits')

WfsArbitrator _157194|INF| 473|2021-04-07 20:59:55.113328| WFS > Command returns success .

WfsArbitrator _157194|INF| 474|2021-04-07 20:59:55.114066| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 475|2021-04-07 20:59:55.114143| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_205953/luci.fits', num = 10)

WfsArbitrator _157194|INF| 479|2021-04-07 21:01:25.002498| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 480|2021-04-07 21:01:25.002578| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_210123/wfs.fits')

WfsArbitrator _157194|INF| 481|2021-04-07 21:01:25.030241| WFS > Command returns success .

WfsArbitrator _157194|INF| 482|2021-04-07 21:01:25.030999| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 483|2021-04-07 21:01:25.031099| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_210123/luci.fits', num = 10)

WfsArbitrator _157194|INF| 484|2021-04-07 21:02:23.840137| WFS > Command returns success .
  • Tracknum | 20210407_205953 |
  • SR 58%
psf_20210407_205953.png

modalplot_20210407_205953.png
  • HEXAPOD ERROR.
  • CALL THE MOUNTAIN->M.WAGNER.
  • Greg T. gave the input for get back the hexapod
  • HEXAPOD_ERROR.png

  • solved... include the steps..
  • procedure to troubleshoot.
    • open the OSS GUI
    • SECONDARY ICON
    • more button
Display more informaiton about the status of the legs.

OSS_3.png
    • OPEN BREAKS
    • INIT
    • HOME
oss_4.png

  • After the HOMED.
oss_5.png

WfsArbitrator _157194|INF| 566|2021-04-07 21:36:58.180570| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 0, zero = 1)

WfsArbitrator _157194|INF| 567|2021-04-07 21:36:58.192076| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 568|2021-04-07 21:36:58.192196| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 569|2021-04-07 21:36:58.193131| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot
  • ncpa on
  • optimize gain
WfsArbitrator _157194|INF| 655|2021-04-07 21:43:45.655922| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 656|2021-04-07 21:43:47.003217| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 657|2021-04-07 21:43:47.003364| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_214345/wfs.fits')

WfsArbitrator _157194|INF| 658|2021-04-07 21:43:47.029417| WFS > Command returns success .

WfsArbitrator _157194|INF| 659|2021-04-07 21:43:47.030284| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 660|2021-04-07 21:43:47.030372| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_214345/luci.fits', num = 10)

WfsArbitrator _157194|INF| 661|2021-04-07 21:43:47.908049| MAIN > Camera lens tracking: moving by x= -0.1 y= -1.0 (um)

WfsArbitrator _157194|INF| 662|2021-04-07 21:43:49.091480| MAIN > Camera lens tracking: moving by x= 0.7 y= 1.7 (um)

WfsArbitrator _157194|INF| 663|2021-04-07 21:43:51.578163| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)

WfsArbitrator _157194|INF| 664|2021-04-07 21:43:52.643431| MAIN > Camera lens tracking: moving by x= -0.3 y= -1.2 (um)
  • gopt

WfsArbitrator _157194|INF| 689|2021-04-07 21:46:05.773493| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 690|2021-04-07 21:46:07.001137| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 691|2021-04-07 21:46:07.001214| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_214605/wfs.fits')

WfsArbitrator _157194|INF| 692|2021-04-07 21:46:07.027754| WFS > Command returns success .

WfsArbitrator _157194|INF| 693|2021-04-07 21:46:07.028447| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 694|2021-04-07 21:46:07.028529| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_214605/luci.fits', num = 10)

WfsArbitrator _157194|INF| 695|2021-04-07 21:46:07.028564| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 696|2021-04-07 21:46:07.028638| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_214605/psf.fits', num = 2)

WfsArbitrator _157194|INF| 697|2021-04-07 21:46:07.718841| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)

WfsArbitrator _157194|INF| 698|2021-04-07 21:46:08.216298| MAIN > Camera lens tracking: moving by x= 1.0 y= 0.2 (um)

WfsArbitrator _157194|INF| 699|2021-04-07 21:46:14.054468| MAIN > Camera lens tracking: moving by x= -1.1 y= -0.6 (um)
  • Tracknum | 20210407_214605 |
  • SR 68%

psf_20210407_214605.png

modalplot_20210407_214605.png
  • after offset bayside stage.
  • PAUSA->RESUME
  • GOPT SEEMS NOT LOCK AGAIN..??

WfsArbitrator _157194|INF| 795|2021-04-07 21:51:45.462543| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 796|2021-04-07 21:51:47.001778| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 797|2021-04-07 21:51:47.001840| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_215145/wfs.fits')

WfsArbitrator _157194|INF| 798|2021-04-07 21:51:47.032814| WFS > Command returns success .

WfsArbitrator _157194|INF| 799|2021-04-07 21:51:47.033424| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 800|2021-04-07 21:51:47.033432| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 801|2021-04-07 21:51:47.033500| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_215145/luci.fits', num = 10)

WfsArbitrator _157194|INF| 802|2021-04-07 21:51:47.033536| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_215145/psf.fits', num = 2)

WfsArbitrator _157194|INF| 803|2021-04-07 21:51:47.663226| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)

WfsArbitrator _157194|INF| 804|2021-04-07 21:51:55.161415| MAIN > Camera lens tracking: moving by x= 0.7 y= 1.0 (um)

WfsArbitrator _157194|INF| 805|2021-04-07 21:51:59.851375| MAIN > Camera lens tracking: moving by x= -1.2 y= -0.6 (um)

WfsArbitrator _157194|INF| 806|2021-04-07 21:52:00.987334| MAIN > Camera lens tracking: moving by x= -1.0 y= -0.4 (um)

WfsArbitrator _157194|INF| 807|2021-04-07 21:52:02.162810| MAIN > Camera lens tracking: moving by x= 1.6 y= 0.8 (um)

  • SHTTER 56
  • MAG 13.3
WfsArbitrator _157194|INF| 832|2021-04-07 21:55:29.246760| COMMANDHANDLER > FSM (status: AOSet) has received command 2041 (PresetAO)

WfsArbitrator _157194|INF| 833|2021-04-07 21:55:29.246888| OPTLOOPDIAG > Executing enableAntiDrift(true)...

WfsArbitrator _157194|INF| 834|2021-04-07 21:55:29.247041| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 0, zero = 1)

WfsArbitrator _157194|INF| 835|2021-04-07 21:55:29.257749| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 836|2021-04-07 21:55:29.257870| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 13.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • cancel preset.
  • set dist s1.5

WfsArbitrator _157194|INF| 922|2021-04-07 22:00:27.141481| OPTLOOPDIAG > Executing enableAntiDrift(true)...

WfsArbitrator _157194|INF| 923|2021-04-07 22:00:27.141627| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 0, zero = 1)

WfsArbitrator _157194|INF| 924|2021-04-07 22:00:27.152096| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 925|2021-04-07 22:00:27.152251| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 13.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 926|2021-04-07 22:00:27.153141| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 927|2021-04-07 22:00:27.153218| MAIN > LUT Applied: deg

WfsArbitrator _157194|INF| 837|2021-04-07 21:55:29.258701| MAIN > LUT Applied: Rotang Ele X_LUT

WfsArbitrator _157194|INF| 998|2021-04-07 22:08:39.203656| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 999|2021-04-07 22:08:41.001871| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 1000|2021-04-07 22:08:41.001973| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_220839/wfs.fits')

WfsArbitrator _157194|INF| 1001|2021-04-07 22:08:41.046492| WFS > Command returns success .

WfsArbitrator _157194|INF| 1002|2021-04-07 22:08:41.047271| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 1003|2021-04-07 22:08:41.047274| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 1004|2021-04-07 22:08:41.047365| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_220839/luci.fits', num = 10)

WfsArbitrator _157194|INF| 1005|2021-04-07 22:08:41.047450| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_220839/psf.fits', num = 2)

WfsArbitrator _157194|INF| 1006|2021-04-07 22:08:45.119479| MAIN > Camera lens tracking: moving by x= 1.6 y= 2.4 (um)

WfsArbitrator _157194|INF| 1007|2021-04-07 22:09:44.766888| WFS > Command returns success .

WfsArbitrator _157194|INF| 1008|2021-04-07 22:09:44.879205| WFS > Command returns success .

WfsArbitrator _157194|INF| 1009|2021-04-07 22:10:36.467075| ASYNCCOMMANDHANDLER > COMMAND received from optloopdiagngui.L

WfsArbitrator _157194|INF| 1010|2021-04-07 22:10:36.467550| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 1011|2021-04-07 22:10:38.001776| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 1012|2021-04-07 22:10:38.001870| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_221036/wfs.fits')

WfsArbitrator _157194|INF| 1013|2021-04-07 22:10:38.034701| WFS > Command returns success .

WfsArbitrator _157194|INF| 1014|2021-04-07 22:10:38.035480| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 1015|2021-04-07 22:10:38.035487| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 1016|2021-04-07 22:10:38.035578| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221036/luci.fits', num = 10)

WfsArbitrator _157194|INF| 1017|2021-04-07 22:10:38.035618| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221036/psf.fits', num = 2)

WfsArbitrator _157194|INF| 1018|2021-04-07 22:11:11.818209| MAIN > Camera lens tracking: moving by x= 6.2 y= 5.5 (um)

WfsArbitrator _157194|INF| 1019|2021-04-07 22:11:24.442587| MAIN > Camera lens tracking: moving by x= -0.0 y= -1.9 (um)

  • gopt on
  • bin2_april7_modalplot.png
WfsArbitrator _157194|INF| 1029|2021-04-07 22:14:15.001714| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_221413/wfs.fits')

WfsArbitrator _157194|INF| 1030|2021-04-07 22:14:15.028838| WFS > Command returns success .

WfsArbitrator _157194|INF| 1031|2021-04-07 22:14:15.029636| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 1032|2021-04-07 22:14:15.029637| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 1033|2021-04-07 22:14:15.029723| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221413/luci.fits', num = 10)

WfsArbitrator _157194|INF| 1034|2021-04-07 22:14:15.029759| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221413/psf.fits', num = 2)

WfsArbitrator _157194|INF| 1037|2021-04-07 22:16:49.424788| ASYNCCOMMANDHANDLER > COMMAND received from optloopdiagngui.L

WfsArbitrator _157194|INF| 1038|2021-04-07 22:16:49.425099| OPTLOOPDIAG > Executing save...

WfsArbitrator _157194|INF| 1039|2021-04-07 22:16:51.002081| WFS > Executing saveStatus...

WfsArbitrator _157194|INF| 1040|2021-04-07 22:16:51.002165| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec_data/20210407/Data_20210407_221649/wfs.fits')

WfsArbitrator _157194|INF| 1041|2021-04-07 22:16:51.038154| WFS > Command returns success .

WfsArbitrator _157194|INF| 1042|2021-04-07 22:16:51.039319| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 1043|2021-04-07 22:16:51.039330| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 1044|2021-04-07 22:16:51.039416| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221649/luci.fits', num = 10)

WfsArbitrator _157194|INF| 1045|2021-04-07 22:16:51.039443| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221649/psf.fits', num = 2)
  • Tracknumber 20210407_221649
  • SR= 45%

psf_20210407_221649.png

modalplot_20210407_221649.png
  • ncpa on
WfsArbitrator _157194|INF| 1054|2021-04-07 22:19:41.001614| WFS > Executing saveStatus...

WfsArbitrator 157194|INF| 1055|2021-04-07 22:19:41.001666| WFS > Python command: WfsArbScripts.saveStatus(self, sensor = 'FLAOWFS', instr = 'LUCIFER', outputFile = '/local/aomeas/adsec data/20210407/Data_20210407_221939/wfs.fits')

WfsArbitrator _157194|INF| 1056|2021-04-07 22:19:41.032537| WFS > Command returns success .

WfsArbitrator _157194|INF| 1057|2021-04-07 22:19:41.033287| WFS > Executing saveLuci...

WfsArbitrator _157194|INF| 1058|2021-04-07 22:19:41.033304| WFS > Executing savePsf...

WfsArbitrator _157194|INF| 1059|2021-04-07 22:19:41.033351| WFS > Python command: WfsArbScripts.saveLuci(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221939/luci.fits', num = 10)

WfsArbitrator _157194|INF| 1060|2021-04-07 22:19:41.033402| WFS > Python command: WfsArbScripts.savePsf(self, sensor = 'FLAOWFS', instr = 'LUCIFER', filename = '/local/aomeas/adsec_data/20210407/Data_20210407_221939/psf.fits', num = 2)

WfsArbitrator _157194|INF| 1061|2021-04-07 22:19:43.340691| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)
  • 20210407_221939
  • SR 56%

psf_20210407_221939.png

modalplot_20210407_221939.png
  • dark
WfsArbitrator _157194|INF| 1074|2021-04-07 22:22:53.364734| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)

WfsArbitrator _157194|INF| 1075|2021-04-07 22:23:00.583726| COMMANDHANDLER > FSM (status: LoopClosed) has received command 2036 (CalibratePISCESDark)

WfsArbitrator _157194|INF| 1076|2021-04-07 22:23:00.583819| WFS > Executing calibratePISCESDark...

WfsArbitrator _157194|INF| 1077|2021-04-07 22:23:00.583924| WFS > Python command: WfsArbScripts.calibIRTCDark(self, sensor = 'FLAOWFS', instr = 'LUCIFER', num = 10, apply = False, ccd = 'luci')

WfsArbitrator _157194|INF| 1078|2021-04-07 22:23:03.391644| WFS > Python command: WfsArbScripts.updateNcpa(self, sensor = 'FLAOWFS', instr = 'LUCIFER', rotAngle = 341.004, zero = 0)

BIN2 pupil check.

  • TEMPERATURE 5.C

bin2_april7.png

BIN2 SHIFT

WfsArbitrator _157194|INF| 1312|2021-04-07 22:57:22.591710| WFS > Executing presetAO...

WfsArbitrator _157194|INF| 1313|2021-04-07 22:57:22.591855| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 13.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _157194|INF| 1314|2021-04-07 22:57:22.592769| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _157194|INF| 1315|2021-04-07 22:57:22.592818| MAIN > LUT Applied: deg deg V V mm mm V V V V

bin2_april7_shift.png
  • script stop at 200

0.65 0.25 0.0999994

0.65 0.05 0.0499997

Setting rotator position 220

ssh LBTO@obs3 "echo \"/lbt/irtc/current/bin/irc RotServicePosition 220.000000 left\" | tcsh"

Resmsg[0] = RotServicePosition result status: OK

Resmsg[1] = OK


0.65 0.25 0.0999994

0.65 0.05 0.0499997

Setting rotator position 200

ssh LBTO@obs3 "echo \"/lbt/irtc/current/bin/irc RotServicePosition 200.000000 left\" | tcsh"

Resmsg[0] = RotServicePosition result status: OK

Resmsg[1] = OK


0.65 0.25 0.0999994

STOP

WfsArbitrator _157194|INF| 1791|2021-04-07 23:23:59.911741| WFS > Executing stopLoop...

THRDTEST cmd: quit
[AOeng@sxadsec ~]$ disable_rr_mode.sh

THRDTEST: 6.15 64b [TH](Built: Mar 11 2021 20:27:08) - Dbg lev.:0, quiet, Line edit & history:Yes

Trying to connect to MsgD @ 127.0.0.1:9752

THRDTEST cmd: write AOARB.L.RR_ENABLED i 0

THRDTEST cmd: quit
[AOeng@sxadsec ~]$

-- JuanCarlosGuerra - 07 Apr 2021
I Attachment Action Size Date Who Comment
 OSS_3.pngpng OSS_3.png manage 31 K 08 Apr 2021 - 00:11 JuanCarlosGuerra  
HEXAPOD_ERROR.pngpng HEXAPOD_ERROR.png manage 104 K 07 Apr 2021 - 21:25 JuanCarlosGuerra  
bin2_april7.pngpng bin2_april7.png manage 54 K 07 Apr 2021 - 22:50 JuanCarlosGuerra  
bin2_april7_modalplot.pngpng bin2_april7_modalplot.png manage 164 K 07 Apr 2021 - 22:20 JuanCarlosGuerra  
bin2_april7_shift.pngpng bin2_april7_shift.png manage 44 K 08 Apr 2021 - 00:04 JuanCarlosGuerra  
modalplot_20210407_205953.pngpng modalplot_20210407_205953.png manage 65 K 13 Apr 2021 - 03:49 JuanCarlosGuerra  
modalplot_20210407_214605.pngpng modalplot_20210407_214605.png manage 63 K 13 Apr 2021 - 03:54 JuanCarlosGuerra  
modalplot_20210407_221649.pngpng modalplot_20210407_221649.png manage 63 K 13 Apr 2021 - 03:59 JuanCarlosGuerra  
modalplot_20210407_221939.pngpng modalplot_20210407_221939.png manage 64 K 13 Apr 2021 - 04:04 JuanCarlosGuerra  
oss_4.pngpng oss_4.png manage 123 K 08 Apr 2021 - 00:12 JuanCarlosGuerra  
oss_5.pngpng oss_5.png manage 128 K 08 Apr 2021 - 00:12 JuanCarlosGuerra  
psf_20210407_205953.pngpng psf_20210407_205953.png manage 111 K 13 Apr 2021 - 03:48 JuanCarlosGuerra  
psf_20210407_214605.pngpng psf_20210407_214605.png manage 117 K 13 Apr 2021 - 03:53 JuanCarlosGuerra  
psf_20210407_221649.pngpng psf_20210407_221649.png manage 106 K 13 Apr 2021 - 03:58 JuanCarlosGuerra  
psf_20210407_221939.pngpng psf_20210407_221939.png manage 104 K 13 Apr 2021 - 04:04 JuanCarlosGuerra  
wfs_init_april7.pngpng wfs_init_april7.png manage 63 K 07 Apr 2021 - 19:44 JuanCarlosGuerra  
Topic revision: r2 - 13 Apr 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback