Staff:

LBTO: Juan Carlos,Alessandro

Arcetri: NA

Summary.

System up to monitor the pupil shift. The power on of the WFS gave an error during the execution of the POWERON. The cube stage/rotator and lamp showed to be offline.

A second power OFF/ON executed with the same outcome. Talking with Alesandro the powerofff/on the AuxUnit gave the cube back online and the power on completes without poroblems; tested two power off/on after and it didn't report any failure. The ocam2K comeup with the pattern and the powercycle the BCU2k/change binning solved the issue.
Not too much time left to perform the rotation test and AO power off and handover to night time. Contacted Josh.

Details

> Executing operate...
WfsArbitrator_153059|INF| 38|2021-03-15 22:13:13.283655| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)
WfsArbitrator_153059|INF| 39|2021-03-15 22:13:33.853272| MAIN > Sensor mode: FLAO
WfsArbitrator_153059|INF| 40|2021-03-15 22:15:49.315856| WFS > Python command: WfsArbScripts.updateModulation(self, sensor = 'FLAOWFS', inst r = '', freq = 2000)
WfsArbitrator_153059|ERR| 41|2021-03-15 22:19:19.844340| WFS > Command returns error : Timeout waiting for picctrl.L.STATUS (-5001) TIMEOUT_ ERROR
WfsArbitrator_153059|INF| 42|2021-03-15 22:19:19.844441| COMMANDHANDLER > Command execution failed: Fsm forced to Failure
WfsArbitrator_153059|ERR| 43|2021-03-15 22:19:19.844493| MAIN > Operate: FATAL - Timeout waiting for picctrl.L.STATUS (-5001) TIMEOUT_ERROR
WfsArbitrator_153059|INF| 44|2021-03-15 22:19:19.846262| WFS > Python command: WfsArbScripts.updateModulation(self, sensor = 'FLAOWFS', inst r = '', freq = 1000)
WfsArbitrator_153059|INF| 45|2021-03-15 22:19:19.848056| WFS

POWER ON FAILED.

WfsArbitrator _153059|INF| 55|2021-03-15 22:26:58.567653| COMMANDHANDLER > FSM (status: PowerOff) has received command 2005 (Off)
WfsArbitrator_153059|INF| 56|2021-03-15 22:26:58.567931| OPTLOOPDIAG > Executing enableAntiDrift(false)...
WfsArbitrator_153059|INF| 57|2021-03-15 22:26:58.568023| WFS > Executing off...
WfsArbitrator_153059|INF| 58|2021-03-15 22:26:58.568084| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = '')
WfsArbitrator_153059|ERR| 59|2021-03-15 22:29:14.452221| WFS > Command returns error : Controller in wrong state for requested operation. (-11005) CTRL_WRONG_STATE_ERROR
WfsArbitrator_153059|INF| 60|2021-03-15 22:29:14.452319| COMMANDHANDLER > Command execution failed: Fsm forced to Failure
WfsArbitrator_153059|ERR| 61|2021-03-15 22:29:14.452366| MAIN > Off: FATAL - Controller in wrong state for requested operation. (-11005) CTRL_WRONG_STATE_ERROR
  • w_stop_all
  • wfs_init
    Arbitrator process didn't come up.
  • wfs_init
WfsArbitrator _191683|INF| 37|2021-03-15 22:36:10.895992| COMMANDHANDLER > FSM (status: Operating) has received command 2004 (Operate)
WfsArbitrator_191683|INF| 38|2021-03-15 22:36:10.896091| OPTLOOPDIAG > Executing enableAntiDrift(true)...
WfsArbitrator_191683|INF| 39|2021-03-15 22:36:10.896304| WFS > Executing operate...
WfsArbitrator_191683|INF| 40|2021-03-15 22:36:10.896401| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)

POWER ON CORRECTLY.

THE OCAM2K SHOWS A NOISE PATTERN.

march15_poweron_2.png

POWER OFF

WfsArbitrator _38831|INF| 66|2021-03-15 23:06:58.997016| OPTLOOPDIAG > Executing enableAntiDrift(false)...
WfsArbitrator_38831|INF| 67|2021-03-15 23:06:58.997106| WFS > Executing off...
WfsArbitrator_38831|INF| 68|2021-03-15 23:06:58.997163| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _83093|INF| 35|2021-03-15 23:10:28.390985| COMMANDHANDLER > FSM (status: PowerOff) has received command 2004 (Operate)
WfsArbitrator_83093|INF| 36|2021-03-15 23:10:28.391061| OPTLOOPDIAG > Executing enableAntiDrift(true)...
WfsArbitrator_83093|INF| 37|2021-03-15 23:10:28.391318| WFS > Executing operate...
WfsArbitrator_83093|INF| 38|2021-03-15 23:10:28.391384| WFS > Python command: WfsArbScripts.setOperating(self, sensor = 'FLAOWFS', instr = '', config = 'WFS_complete_(with_ccd47)', boardSetup = '', opticalSetup = False)
WfsArbitrator_83093|INF| 39|2021-03-15 23:10:46.350646| MAIN > Sensor mode: FLAO

fsArbitrator_83093|INF| 42|2021-03-15 23:18:17.741099| COMMANDHANDLER > Command Operate (code 2004) successfully completed

POWER CYCLE BCU2K.
GOOD.

Adsec control gui. system skipping frame in a rest position.

Focal station set for the Bent gregorian.. This is because the presetAO from the AOS sent without the sell SET.   The AOS doesn't report back any skip frame information. discussing with Guyido and Xinayu if a Flat could be applied. To discuss

march15_poweron_3.png

VNC argos to another VNViewer.
  • vncviewer 192.168.27.148:2,3
WfsArbitrator _83093|INF| 87|2021-03-15 23:39:01.007059| WFS > Python command: WfsArbScripts.setPowerOff(self, sensor = 'FLAOWFS', instr = 'LUCIFER').

23:39 Stop

-- JuanCarlosGuerra - 15 Mar 2021
Topic revision: r1 - 16 Mar 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback