staff:

LBTO: Juan Carlos

Arcetri: NA

Summary.

It was observed yesterday that the Bin1 could be shifted one pixel up to avoid the saturation in the negative Y direction. The pupil mask shift was applied yesterday and checked today. The Y saturation was almost eliminated with some spike in the rotator position about 240.

System setup for only Bin1 checkout and handover to Guido for checking the TSS after the issues with anemomenter last night.

Details.

  • Setting up the system to go in bin1.
  • Argos light shutter full open.
  • magniture 10.
  • PresetAO from AOS
WfsArbitrator _55611|INF| 51|2021-03-10 21:58:24.939607| WFS > Executing presetAO...
WfsArbitrator_55611|INF| 52|2021-03-10 21:58:24.939726| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator _55611|INF| 88|2021-03-10 22:07:28.398294| WFS > Executing centerPupil...
WfsArbitrator_55611|INF| 89|2021-03-10 22:07:28.398342| WFS > Python command: WfsArbScripts.centerPupils(self, sensor = 'FLAOWFS', instr = 'LUCIFER')
WfsArbitrator_55611|WAR| 90|2021-03-10 22:07:28.400056| WFS > Command returns retryable error : RETRY: Camera lens temperature too low. T= 6.59, minimum allowed is 7.00 (0) NO_ERROR
WfsArbitrator_55611|INF| 91|2021-03-10 22:07:28.400310| COMMANDHANDLER > Command not executed (status RETRY) (errstr RETRY: Camera lens temperature too low. T= 6.59, minimum allowed is 7.00 (0) NO_ERROR)
WfsArbitrator_55611|WAR| 92|2021-03-10 22:07:28.400354| MAIN > centerPupils: WARNING - RETRY: Camera lens temperature too low. T= 6.59, minimum allowed is 7.00 (0) NO_ERROR


WfsArbitrator_55611|INF| 117|2021-03-10 22:09:39.730345| MAIN > Received cmd: centerPupils()
WfsArbitrator_55611|INF| 118|2021-03-10 22:09:39.730442| COMMANDHANDLER > FSM (status: AOSet) has received command 2039 (centerPupils)
WfsArbitrator_55611|INF| 119|2021-03-10 22:09:39.730481| WFS > Executing centerPupil...
WfsArbitrator_55611|INF| 120|2021-03-10 22:09:39.730518| WFS > Python command: WfsArbScripts.centerPupils(self, sensor = 'FLAOWFS', instr = 'LUCIFER')
WfsArbitrator_55611|WAR| 121|2021-03-10 22:09:39.732306| WFS > Command returns retryable error : RETRY: Camera lens temperature too low. T= 6.84, minimum allowed is 7.00 (0) NO_ERROR
WfsArbitrator_55611|INF| 122|2021-03-10 22:09:39.732352| COMMANDHANDLER > Command not executed (status RETRY) (errstr RETRY: Camera lens temperature too low. T= 6.84, minimum allowed is 7.00 (0) NO_ERROR)
WfsArbitrator_55611|WAR| 123|2021-03-10 22:09:39.732380| MAIN > centerPupils: WARNING - RETRY: Camera lens temperature too low. T= 6.84, minimum allowed is 7.00 (0) NO_ERROR

Running wobble_trest_back.

Set rot to 440.

BIN1 PUP ROT CHECK

bin1_pupilcheck_march10.png

march10_bin1_pupilerrcheck.png

bin1_pupilerrcheck_march10_up.png

Open loop.
  • Cancel preset from FLAOgui.
  • Retracted Argos calk unit.
  • turn off the light Argos cal unit.
  • MCS rotator position to 340
  • MCS rotator speed to 5deg/sec
  • enable lookup table and temperature compensation.
  • disable_rr_mode.
  • ocam set safe
  • WFS power OFF.

22:50 STOP