Summary

Pupil check for bin3 after observing during night time a pupil errors above threshold at rot 260.

System performance check for bin1, bin3 and bin4. Not done.

My remote connection was very very slow amd it took too long to work with the GUIs. specially the VNCviewer in Argos was very slow.

Details.

WFS check and power ON

[AOeng@soul-sxwfs SOUL]$ w_check
msgdrtdb running

mirrorctrl NOT running
pinger running and connected to Msgd
powerboard running and connected to Msgd
slopecompctrl NOT running
techviewctrl NOT running
ocam2 NOT running
ccd47 NOT running
ttctrl NOT running
adc1 NOT running
adc2 NOT running
rerotator NOT running
filterwheel1 NOT running
filterwheel2 NOT running
mercury NOT running
cubestage NOT running
baysidex NOT running
baysidey NOT running
baysidez NOT running
cameralensctrl NOT running
picctrl NOT running
optloopdiag NOT running
pupilcheckocam2 NOT running
masterdiagnostic NOT running
housekeeperWFS NOT running
dbfiller running and connected to Msgd
wfsarb NOT running
od_gopt NOT running
gopt NOT running

wfs_init

  • Power on WFS.OK
  • Ocam ambient temperature. OK
11:48. Chamber Lights still ON

Setup.

WFS left configured for LBTI. It has to be set to work with the FLAO system.

WfsArbitrator_69009|INF| 51|2021-02-26 19:06:34.887272| WFS > Executing presetAO...
WfsArbitrator_69009|INF| 52|2021-02-26 19:06:34.887386| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LBTI', mode = 'ACE-AO', starxpos = 0.09, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

WfsArbitrator_69009|INF| 53|2021-02-26 19:06:34.888293| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator_69009|INF| 54|2021-02-26 19:06:34.888321| MAIN > LUT Applied: deg deg V V mm mm V V V V
WfsArbitrator_69009|WAR| 55|2021-02-26 19:06:34.888747| WFS > Command returns retryable error : RETRY: No settings for instrument:LBTI - mode:ACE-AO (-20004) WFSARB_ARG_ERROR

WfsArbitrator_69009|INF| 56|2021-02-26 19:06:34.888786| COMMANDHANDLER > Command not executed (status RETRY) (errstr RETRY: No settings for instrument:LBTI - mode:ACE-AO (-20004) WFSARB_ARG_ERROR)
WfsArbitrator_69009|WAR| 57|2021-02-26 19:06:34.888808| MAIN > presetAO: WARNING - RETRY: No settings for instrument:LBTI - mode:ACE-AO (-20004) WFSARB_ARG_ERROR

  • Check for Luci Authorize and do it again.
  • Set in AOSGUI->WFS spec to FLAOWFS
  • PSFGUI. Clear the Active optics. also there are different offsets already applied. Cleaer the offsets.
lights off.

Open the GUI ENV from TCS.

WfsArbitrator _69009|INF| 91|2021-02-26 19:17:01.428401| COMMANDHANDLER > Command CalibrateTVDark (code 2018) successfully completed

WfsArbitrator _69009|INF| 96|2021-02-26 19:23:40.681799| COMMANDHANDLER > Command CalibrateHODark (code 2014) successfully completed

WfsArbitrator _69009|INF| 110|2021-02-26 19:32:42.185732| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 10, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

Center Star ->OK looking at the CCD47.
Not too much light on OCAM.
Cancel preset FLAOGUI

WfsArbitrator _69009|INF| 153|2021-02-26 19:48:18.339370| WFS > Executing presetAO...

WfsArbitrator _69009|INF| 154|2021-02-26 19:48:18.339460| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0,

WfsArbitrator _69009|INF| 165|2021-02-26 19:51:23.591103| WFS > Executing centerStar...

WfsArbitrator _69009|INF| 166|2021-02-26 19:51:23.591348| WFS > Python command: WfsArbScripts.centerStar(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _69009|INF| 167|2021-02-26 19:51:23.592380| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _69009|INF| 168|2021-02-26 19:51:23.592429| MAIN > LUT Applied: deg deg V V mm mm V V V V

WfsArbitrator _69009|INF| 169|2021-02-26 19:51:49.589901| MAIN > LUT Applied: Rotang Ele X_LUT Y_LUT X_Enc Y_Enc X_Enc Y_Enc X_Tot Y_Tot

WfsArbitrator _69009|INF| 170|2021-02-26 19:51:49.589954| MAIN > LUT Applied: deg deg V V mm mm V V V V

WfsArbitrator _69009|INF| 171|2021-02-26 19:51:49.687635| WFS > Command returns success : 0.3431359 2.8404806

WfsArbitrator _69009|INF| 172|2021-02-26 19:51:49.687689| COMMANDHANDLER > Command centerStar (code 2038) successfully completed.

Camera lens positions: posX=0,posY=0

derotAngle = 0, starMag = 8, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

CLOSE LOOP BIN1 USING AUTOMATIC MODE IN THE AOSGUI.
ARGOS SHUTTER 51.WFS MAG 14.2

WfsArbitrator _69009|INF| 364|2021-02-26 20:33:12.835146| WFS > Python command: WfsArbScripts.presetAO(self, sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 14, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
MANUAL.

WfsArbitrator _69009|INF| 392|2021-02-26 20:39:49.185024| WFS > Command returns success .

WfsArbitrator _69009|INF| 393|2021-02-26 20:39:49.185092| COMMANDHANDLER > Command centerPupils (code 2039) successfully completed

WfsArbitrator _69009|INF| 394|2021-02-26 20:39:49.185438| WFS > Executing closeLoop...

WfsArbitrator _69009|INF| 395|2021-02-26 20:39:49.185500| WFS > Python command: WfsArbScripts.closeLoop(self, sensor = 'FLAOWFS', instr = 'LUCIFER')

WfsArbitrator _69009|INF| 396|2021-02-26 20:39:49.189486| WFS > Command returns success .

WfsArbitrator _69009|INF| 397|2021-02-26 20:39:49.189515| MASTDIAGN > Setting master diagnostic in closed loop mode

WfsArbitrator _69009|INF| 398|2021-02-26 20:39:54.088587| MAIN > Camera lens tracking: mov

CAMERA LENS DIDN'T CONVERGE IN 180SEC.
OBSERVED THAT THE POSY WAS AT -100 FIXED. HARDW GUI SHOWS POSX=-100.
MOVE MANUALLY THE X TO MOVE OUT -100.

BIN2 PUPIL CHECK.

sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 14, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
  • Argos 54.
  • mag 14.3
  • OPtimize gain.OK

bin2_pup_doen_up_feb26.png

bin2_puperr_doen_up_feb26.png

BIN3

  • Argos 49.
  • mag 15.3
  • OPGain. Ok
sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 15.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)
sensor = 'FLAOWFS', instr = 'LUCIFER', mode = 'ACE-AO', starxpos = -0, starypos = 0, telElevation = 0, derotAngle = 0, starMag = 15.3, starColor = 0, r0 = 0, badseeing = 0, v0 = 0)

bin3_pupdown_up_feb23.png

bin3_puperr_down_up2.png

bin3_puperr_up_down_3.png

23:12 STOP

-- JuanCarlosGuerra - 26 Feb 2021
Topic revision: r3 - 03 Mar 2021, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.