20210222 LBTB Science OSCO + Bin 4

02:12 Preset to 15.5 mag
  • see 16.5

  • OCAM Parameters 4, 495, 3, 600
  • The OCAM has bright corners LL and UR. See image. Not near pupils, but may be extra noise in pupils
  • bin4_bright corners.jpg:
    bin4_bright corners.jpg
  • Optimize Gain failes on second pass
  • Open loop, Reset OCAM to bin 1. Looks good.
  • Reset OCAM to bn 4, 495 Hz Looks better, no bright corners
  • Optimize gain fail on second pass. Can barely see pupil
  • Open loop, set loop to .
  • TN 023514, 023823

02:41 AdSec Failure.

RR|       743|2021-02-23 02:41:19.515891|    ALERTHANDLER > Alert message: COILS DISABLED from housekeeper.L
20212302 02:41:19.5171 Alert: Error: COILS DISABLED
AdSecArbitrator_112795|INF|       744|2021-02-23 02:41:20.872597|    ALERTHANDLER > Received from masterdiagnostic.L a DumpSaved alert with filename /local/aolog/2021/02/23/ADAPTIVE-SECONDARY_00001614048080_20210223_024120.bin
AdSecArbitrator_112795|INF|       745|2021-02-23 02:41:20.872624|    ALERTHANDLER > Processdump is skipped
20212302 02:41:20.8737 Alert: DumpSaved: DumpSaved vector ready
AdSecArbitrator_112795|ERR|       746|2021-02-23 02:41:21.903769|    ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_112795|ERR|       747|2021-02-23 02:41:21.903843|    ALERTHANDLER > Alert message: COILS DISABLED from housekeeper.L
20212302 02:41:21.9049 Alert: Error: COILS DISABLED
AdSecArbitrator_112795|ERR|       748|2021-02-23 02:41:22.119671|    ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_112795|ERR|       749|2021-02-23 02:41:22.119715|    ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
20212302 02:41:22.1206 Alert: Error: COILS DISABLED
AdSecArbitrator_112795|INF|       750|2021-02-23 02:41:22.175842|    ALERTHANDLER > Received from masterdiagnostic.L a DumpSaved alert with filename /local/aolog/2021/02/23/ADAPTIVE-SECONDARY_00001614048082_20210223_024122.bin
AdSecArbitrator_112795|INF

Adding housekeeper log message:

housekeeper_112700 |INF| 85365|2021-02-23 02:41:18.428532| FAST > Processed 1 cycles of 22443 vars in 1.07 s ( 0.939 Hz). HW frame rate 0.939 [DiagnApp.cpp:832]
housekeeper_112700 |ERR| 85366|2021-02-23 02:41:18.449446| MAIN > Error reading BCUs: [RequestFaultException] The request has failed [HouseKeeper.cpp:2645]
housekeeper_112700 |ERR| 85367|2021-02-23 02:41:19.473814| FUNCTEMERGENCYSTOP > BCUVOLTAGEVCCP-0001 = -0.0411356 [Funct.cpp:95]
housekeeper_112700 |INF| 85368|2021-02-23 02:41:19.488965| ADAM-MODBUS > AdamModbus: disabling coils..

Updated IT#8304.

  • IDL> print. fsm_power_off()
  • Josh move the telescope to 65 el. Wind at adsec is 0.2 m/s
  • adsc_stop/start
  • Josh press E on AOS gui
  • Will need to set elevation limit to 26
  • Power On
  • Load program
  • Set

02:46 Some W process die while AdSec was being restarted

[AOeng@soul-sxwfs SOUL]$ w_check
msgdrtdb             NOT running
mirrorctrl           running but UNKOWN connection state to Msgd
pinger               running but UNKOWN connection state to Msgd
powerboard           running but UNKOWN connection state to Msgd
slopecompctrl        running but UNKOWN connection state to Msgd
techviewctrl         running but UNKOWN connection state to Msgd
ocam2                running but UNKOWN connection state to Msgd
ccd47                NOT running
ttctrl               NOT running
adc1                 running but UNKOWN connection state to Msgd
adc2                 running but UNKOWN connection state to Msgd
rerotator            running but UNKOWN connection state to Msgd
filterwheel1         running but UNKOWN connection state to Msgd
filterwheel2         running but UNKOWN connection state to Msgd
mercury              running but UNKOWN connection state to Msgd
cubestage            running but UNKOWN connection state to Msgd
baysidex             running but UNKOWN connection state to Msgd
baysidey             running but UNKOWN connection state to Msgd
baysidez             running but UNKOWN connection state to Msgd
cameralensctrl       NOT running
picctrl              running but UNKOWN connection state to Msgd
optloopdiag          NOT running
pupilcheckocam2      running but UNKOWN connection state to Msgd
masterdiagnostic     NOT running
housekeeperWFS       NOT running
dbfiller             running but UNKOWN connection state to Msgd
wfsarb               running but UNKOWN connection state to Msgd
od_gopt              NOT running
gopt                 running but UNKOWN connection state to Msgd
Error: w_check script UNABLE to connect to RTDB!
The exception wasthStart error(-2010): Connection closed at the other end

  • w_stop * w_start: failed, housekeeper did not start
tartProcessByName(): starting process /home/AOeng/aoroot/bin/DBFiller -i dbfiller
Traceback (most recent call last):
  File "/home/AOeng/aoroot/bin/w_start", line 256, in 
    if __name__ == '__main__': main()
  File "/home/AOeng/aoroot/bin/w_start", line 144, in main
    processControl.waitProcessStart( app, process)
  File "/home/AOeng/aoroot/lib/python/AdOpt/processControl.py", line 243, in waitProcessStart
    raise AOExcept.AdOptError( AOConstants.AOErrCode['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)
AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR
  • w_stop/start
  • All processed back up.
  • Click chilled
  • OCAM not live
  • Reset to bin 2. Live now

03:01 Pupil check ocam2 not running. Not circles. Start process, center pupil working and complete
  • pupil check ocam2 down again when closed loop, bin 3
  • bin3_middle stripe.jpg:
    bin3_middle stripe.jpg
  • change to bin 4, 200 Hz
  • optimize gain Good
  • Luci taking data
  • 24 minutes of down time, according to Josh
  • TN 031351,031528, 032311, 032933

03:54 Done with AO tonight

-- DougMiller - 23 Feb 2021