20201028 Closed Dome check NCPA

04:00 System Running

04:24 Turned on LUCI1 Rotator. See more vibrations in AO Loop
  • TN 042443
  • LUCI1 10

04:46 Take dark
  • LUCI 256x256
  • Dark 20201028_044652, luci 11
Moving to rotator angle  360
    Set  0   Tracknum  20201028_051435   LuciName  20201028/luci1.20201028.0012.fits
    Set  1   Tracknum  20201028_051605   LuciName  20201028/luci1.20201028.0013.fits
    Set  2   Tracknum  20201028_051725   LuciName  20201028/luci1.20201028.0014.fits
  
Moving to rotator angle  330
    Set  0   Tracknum  20201028_051902   LuciName  20201028/luci1.20201028.0015.fits
Moving to rotator angle  330
    Set  0   Tracknum  20201028_054505   LuciName  20201028/luci1.20201028.0018.fits
    Set  1   Tracknum  20201028_054628   LuciName  20201028/luci1.20201028.0019.fits
    Set  2   Tracknum  20201028_054748   LuciName  20201028/luci1.20201028.0020.fits

...

Moving to rotator angle  0
    Set  0   Tracknum  20201028_063336   LuciName  20201028/luci1.20201028.0051.fits
    Set  1   Tracknum  20201028_063504   LuciName  20201028/luci1.20201028.0052.fits
    Set  2   Tracknum  20201028_063623   LuciName  20201028/luci1.20201028.0053.fits

List of log names written to check/check.log

It appears that the NCPA loop is not applying the correct values

08:15 Manually enter NCPA values from LUT in GOP GUI
  • Rotator at 330
  • Manually applied LUT values on GOPT GUI
    • 330 -18.70 -10.00 30.00 -5.00 5.00 -10.00 -5.00 -4.10
  • TN 20200828_081939
  • LUCI image looks good
  • ncpa_manual.jpg:
    ncpa_manual.jpg
  • ncpa_manual_psf.jpg:
    ncpa_manual_psf.jpg

08:28 Turn on NCPA loop
  • LUT values
    • 330 -18.70 -10.00 30.00 -5.00 5.00 -10.00 -5.00 -4.10
  • Applied by loop as seen on GOPT GUI
    • 330 -18.2, -30.5, 12.7, -0.2, -7.1, 2.5, -11.4, -3.3
  • TN 20200828_083038
  • LUCI image looks bad
  • ncpa_loop.jpg:
    ncpa_loop.jpg
  • ncpa_loop_psf.jpg:
    ncpa_loop_psf.jpg

-- DougMiller - 28 Oct 2020

I Attachment Action Size Date Who Comment
ncpa_loop.jpgjpg ncpa_loop.jpg manage 29 K 28 Oct 2020 - 08:37 DougMiller  
ncpa_loop_psf.jpgjpg ncpa_loop_psf.jpg manage 8 K 28 Oct 2020 - 08:37 DougMiller  
ncpa_manual.jpgjpg ncpa_manual.jpg manage 30 K 28 Oct 2020 - 08:36 DougMiller  
ncpa_manual_psf.jpgjpg ncpa_manual_psf.jpg manage 8 K 28 Oct 2020 - 08:36 DougMiller  