SX SOUL Commissioning

https://wiki.lbto.org/Commissioning/Com20201001

fastdiagn_24391 |WAR| 3774049|2020-10-01 01:27:07.513432| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774052|2020-10-01 01:27:08.781597| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774056|2020-10-01 01:27:11.050777| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774057|2020-10-01 01:27:11.188270| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774060|2020-10-01 01:27:12.806019| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774061|2020-10-01 01:27:13.577696| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774063|2020-10-01 01:27:13.799645| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774064|2020-10-01 01:27:14.592188| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774067|2020-10-01 01:27:15.810071| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774070|2020-10-01 01:27:17.936590| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16
fastdiagn_24391 |WAR| 3774071|2020-10-01 01:27:18.134954| FUNCTWARNING > FunctWarning CHDISTRMS-0638 -2.95555e-16

No problems are seen on the mirror GUIs for this actuator, I'm leaving it in.

"mostra, #" didn't work when I brought up a terminal from obs4, and ssh'ed to the adsec, but it worked from obs3....

Starting for tonight

01:35
  • Getting ready for OSCO.
02:33
  • pupilcheck ocam didn't come back up after changing the binning, so I did it manually.
  • Bin 2,
  • 898 Hz
  • Catalog mag 12.06
  • WFS 13.3
  • GOPT 0.121
  • g_tt_1.94_ho1_0.14_ho2_0.11
03:13
  • M52
  • Catalog 12.35
  • WFS 13.7
  • Safe skip and the loop opened.
  • bin2
  • 827 Hz
  • GOPT 0.111
  • g_tt_1.64_ho1_0.09_ho2_0.10
04:11
  • High Current on actuator #164
  • The loop keeps closing before we are 100 percent back from a long offset
  • Epic one liner, first, in the shell:
  • export bob=0
  • Then:
  • tail -F housekeeper.L_TELEMETRY.202010010002.tel | perl -ne '$bob = $ENV{bob}; if(/SafeSkipFrameCnt\s\>\s(.+)$/){if($1 \!= $bob){print $_; $bob = $1; $ENV{bob} = $bob}}'
  • This gives us:
  • housekeeper.L |INF| 1713056|2020-10-01 04:06:30.544548| SwitchSafeSkipFrameCnt > 4019
    housekeeper.L |INF| 1716356|2020-10-01 04:11:59.093939| SwitchSafeSkipFrameCnt > 4034
  • It only prints a line when the "SafeSkipFrameCnt" changes
04:30
  • Moving to GRB thingy
  • Catalog mag 15.3
  • WFS 16.2
  • Bin3
  • 310 Hz
05:04
  • We had a huge seeing bubble.
  • We re-closed and are now in bin4, 319 Hz.
  • Sows mag 16.2, partially moonlight, I think.
05:41
  • all CCD viewers have frozen and we can't get them up again.
  • Preparing to cycle the power and the processes for the WFS.
  • Finally came back.
06:33
  • We've just had a power bump

-- GregoryTaylor - 01 Oct 2020