20200928 Night SX SOUL Commissioning

01:40 WFS Msgdrtb died as did several other processes.
  • I had started an AUXLOOP GUI about the time the processes died. Tried AUX Loop GUI later, no affect.
  • Did a w_stop_all and w_stop_all and all was good.
  • Next Preset and RunAO was successful for OSCO

01:55 Getting many skip frames in bin 1. 8.9 mag
  • GOPT was bouncing around
  • 0.3 -> 0.17 -> 1.0 -> 1.6 -> 0.19 etc
  • Turn off GOPT, Set GOPT to 1.0 and set gains back to ~original
  • 2.5, 0.5, 0.3

02:04 TN
  • 020451
  • 020531
  • 020636

02:25 Closed loop, bin 2, 13.7 mag
  • 024441
  • 025816
  • 032008

03:33 Error when using telserv to make RaDec offset. WFS throws errors below.
  • Loop is Paused
WfsArbitrator_178192|INF|       846|2020-09-29 03:33:20.544481|             MAIN > offsetXYZ(): Focus correction (absolute): 0.000 mm
WfsArbitrator_178192|INF|       847|2020-09-29 03:34:00.604360|             MAIN > Stage movement timeout after 40 seconds. Retry 1 of 3
WfsArbitrator_178192|INF|       848|2020-09-29 03:34:43.701178|             MAIN > Stage movement timeout after 40 seconds. Retry 2 of 3
WfsArbitrator_178192|ERR|       849|2020-09-29 03:35:26.850265|             MAIN > RETRY: Stages position not reached (-5001) TIMEOUT_ERROR
  • Sent absolute 0,0 offset via telserv.
  • bayside stage move and loop is closed
  • All is well and LUCI take data

04:39 Loop stalled.
  • OCAM report not live. Live when loop was opened on WFS Control GUI
  • AdSec in Failure state
  • Recover failure did not work
  • fsm_power_off() at IDL terminal
  • adsc_stop, adsec_start, power on, load, setflatAO. All ok
  • Stop/Start WFS arbitrator
  • Guessing that something powered off the AdSec.

04:46 WFS msgdrtdb died as did several processes
  • w_stop_all
  • w_start_all
  • everything ok
  • Next Preset and RunAO sucessfull

05:30 ADI observation