20200708 First Half: OSCO, ESM Science, NSO, ESM

Closed Dome test

02:15 PresetAO from AOS GUI
  • Sucess
  • On-axis bayside -76.5, -50.29

02:37 NS Preset to Anik
  • Bayside -76.41, -50.19
  • Preset failed because was ACE_ACTIVE
  • Preset delta x,y 0.0, -0.1

02:44 NS Preset to Anik
  • Bayside -76.41, -50.19
  • Preset was AC_TRACK
  • Preset delta x,y 0.0, -0.1
  • Canceled AcquireRef
  • offset in preset script 0.0

02:47
  • offset 5,5 in LUCI preset
  • xy offset in PresetAO is 2.99, 3.10
  • bayside -79.414, -47.209

02:50
  • offset -5, -5 in LUCI preset
  • xy offset in PresetAO -3.0, -2.98
  • bayside -73,4, -53.164

QED!

03:20ish OSCO
  • OCAM went not live when centering pupil
  • restarted wfsarbitrator, no change
  • WFS Master diagnostics died
  • camera lens was not moving
  • Cancel preset and move to ESM science field

03:38 ESM Preset
  • Same problem as OSCO
  • OCAM go not live when pupil center start
  • AdSec go into failure when Loop open
  • restart AdSec
  • Stop/start wfsarbitrator

AdSecArbitrator_141169|INF|       405|2020-07-09 04:20:50.024969|  COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_141169|INF|       406|2020-07-09 04:20:50.025336|            MAIN > Received cmd: PauseAo()
AdSecArbitrator_141169|INF|       407|2020-07-09 04:20:50.025406|  COMMANDHANDLER > FSM (status: AORunning) has received command 2027 (Pause Ao)
AdSecArbitrator_141169|INF|       408|2020-07-09 04:20:50.032730|  COMMANDHANDLER > Command Pause Ao (code 2027) successfully completed
AdSecArbitrator_141169|INF|       409|2020-07-09 04:21:01.732333|            MAIN > Received cmd: ResumeAo()
AdSecArbitrator_141169|INF|       410|2020-07-09 04:21:01.732440|  COMMANDHANDLER > FSM (status: AOPause) has received command 2028 (ResumeAo Ao)
AdSecArbitrator_141169|INF|       411|2020-07-09 04:21:01.736813|  COMMANDHANDLER > Command ResumeAo Ao (code 2028) successfully completed
AdSecArbitrator_141169|INF|       412|2020-07-09 04:21:02.126922|            MAIN > Received cmd: Offloading(1)
AdSecArbitrator_141169|INF|       413|2020-07-09 04:21:02.126999|  COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_141169|INF|       414|2020-07-09 04:21:02.134748|  COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_141169|ERR|       415|2020-07-09 04:21:02.469970|    ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_141169|ERR|       416|2020-07-09 04:21:02.470006|    ALERTHANDLER > Alert message: COILS DISABLED from housekeeper.L
20200907 04:21:02.4708 Alert: Error: COILS DISABLED
AdSecArbitrator_141169|INF|       417|2020-07-09 04:21:03.816119|    ALERTHANDLER > Received from masterdiagnostic.L a DumpSaved alert with filename /local/aolog/2020/07/09/ADAPTIVE-SECONDARY_00001594268463_20200709_042103.bin
AdSecArbitrator_141169|INF|       418|2020-07-09 04:21:03.816150|    ALERTHANDLER > Processdump is skipped
20200907 04:21:03.8170 Alert: DumpSaved: DumpSaved vector ready
AdSecArbitrator_141169|ERR|       419|2020-07-09 04:21:06.155357|    ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_141169|ERR|       420|2020-07-09 04:21:06.155392|    ALERTHANDLER > Alert message: COILS DISABLED from fastdiagn.L
20200907 04:21:06.1561 Alert: Error: COILS DISABLED

housekeeper_25836  |ERR|    163474|2020-07-09 04:21:02.409332| FUNCTEMERGENCYSTOP > BCUVOLTAGEVSSP-0004 = 0.49163  [Funct.cpp:95]
housekeeper_25836  |INF|    163475|2020-07-09 04:21:02.409369|      ADAM-MODBUS > AdamModbus: disabling coils...

  • Restarted AdSec
  • Stop/start WFS Arbitrator
  • SX Loop opened and then LUCI script failed on the next offset
  • Sent PresetAO from AOS Command GUI
  • Sent AcquireRef from AOS Command GUI
  • Close Loop with FLAO GUI
  • Set Gains manually. Basically the same values as before, 1.0, 0.6
  • StartAO on AOS Command GUI
  • IIF See the loop is closed
  • David canceled DX Offset that was waiting for SX Offset
  • Andrew start LUCI data collection script at place it died before. All is Good.

05:05 Preset to star near Anik to collimate

05:15 NS Preset
  • +IE 30 moves up
  • +CA 30 left
  • -CA 60 to upper left
  • -CA 3 move diagonal ul to lr
  • -IE 5 move ll to ur
  • +IE 10 move down
  • +CA 5 spot on
  • 05:40

05:43 NS Preset

05:49 AO Loop Closed. Off to Andrew

05:58 Subframe setup and start collecting data. LUCI1 33

-- DougMiller - 09 Jul 2020
Topic revision: r3 - 09 Jul 2020, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback