20200208 SX AO Commissioning

8.95

08:40 Start collection of SX AO Performance data with IRTC
  • 1, 1700, 460
  • 0900644 108 No GOPT
  • 091143, 109
  • 091212, 110
  • 091359, 111
  • Dark, 112

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq emGain gain mod nph/sa/fr Gopt AntiDrift SR SR (from slopes) FWHM (mas) filter exp #frames disturb SN skip notes
20200208_090644 ??? 10.0 272827 4 0.91 0.10 161000 1 503 1700 460 1.3 0.4 1.3 3 14.4 1.00 OFF 68.8 -100.0 -1 H 50 1 ONSKY NO 0 GOPT Off
20200208_091143 ??? 10.0 276317 4 0.95 0.10 161000 1 503 1700 460 0.5 0.2 0.5 3 14.6 0.35 OFF 70.0 66.3 -1 H 50 1 ONSKY NO 0 GOPT On
20200208_091212 ??? 10.0 276654 4 0.95 0.10 161000 1 503 1700 460 0.5 0.2 0.5 3 14.5 0.28 OFF 78.1 54.7 -1 H 50 1 ONSKY NO 0 GOPT On
20200208_091359 ??? 10.0 277897 4 0.94 0.10 161000 1 503 1700 460 0.5 0.2 0.5 3 14.4 0.33 OFF 69.4 67.7 -1 H 50 1 ONSKY NO 0 GOPT On

IDL string: ['20200208_090644','20200208_091143','20200208_091212','20200208_091359']
SR 0.7155 (0.0436372)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 0.938751 (0.0219051)

09:27 DX AdSec went into failed state because of Stop from fastdiagnostics
  • recover fail: fail
  • adsc_stop/start

09:42 DX AdSec went into failed state because of Stop from fastdiagnostics
  • recover fail: fail
  • adsc_stop/start

10:10 DX AdSec went into failed state because of Stop from fastdiagnostics
  • recover fail: fail
  • adsc_stop/start

10:12 Optimize gain completed successfully, but set all gains to 0.0. Curves looked normal. Tracking num 20200208_101201
  • Never mind, this was because Gopt diag was not running and set everything to 0.

R = 9.8

  • TNs -> IRTC num
  • 103930 -> 113
  • 103953 -> 114
  • 104009 -> 115
  • dark -> 116

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq emGain gain mod nph/sa/fr Gopt AntiDrift SR SR (from slopes) FWHM (mas) filter exp #frames disturb SN skip notes
20200208_103930 ??? 10.1 286184 4 0.87 0.10 161000 1 503 1700 514 0.5 0.1 0.5 3 13.1 0.28 OFF 56.8 62.2 -1 H 50 1 ONSKY NO 0 GOPT On
20200208_103953 ??? 10.1 286471 4 0.83 0.10 161000 1 503 1700 514 0.5 0.1 0.5 3 13.2 0.31 OFF 53.5 62.4 -1 H 50 1 ONSKY NO 0 GOPT On
20200208_104009 ??? 10.1 286670 4 0.83 0.10 161000 1 503 1700 514 0.5 0.1 0.5 3 13.1 0.32 OFF 60.3 63.6 -1 H 50 1 ONSKY NO 0 GOPT on

IDL string: [,'20200208_103930','20200208_103953','20200208_104009']
SR 0.568585 (0.0338954)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 0.842644 (0.0232685)
10:50 DX adsec left powered on, but program not loaded.

R =10.5

  • 110047 -> 117
  • 110241 -> 118
  • 110327 -> 119
  • dark -> 120

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq emGain gain mod nph/sa/fr Gopt AntiDrift SR SR (from slopes) FWHM (mas) filter exp #frames disturb SN skip notes
20200208_110047 ??? 11.1 295528 5 1.24 0.10 161000 1 503 1444 600 0.8 0.8 -1.0 3 6.5 0.27 OFF 60.8 53.3 -1 H 100 1 ONSKY NO 0  
20200208_110241 ??? 11.0 294359 5 NaN 0.10 161000 1 503 1444 600 0.8 0.8 -1.0 3 6.8 0.28 OFF 66.8 -1.0 -1 H 100 1 ONSKY NO 0  
20200208_110327 ??? 11.0 293877 4 1.33 0.10 161000 1 503 1444 600 0.8 0.8 -1.0 3 6.7 0.30 OFF 68.1 52.7 -1 H 100 1 ONSKY NO 0  

IDL string: ['20200208_110047','20200208_110241','20200208_110327']
SR 0.65196 (0.0388512)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 1.28586 (0.0603798)

R = 11.3

11:16 GS is too close, vignetting WFS so changing stars. Also SX adsec is spewing dumpfiles.

R = 11.1

  • 113425 -> 121
  • 113550 -> 122
  • 113728 -> 123
  • dark -> 124

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq emGain gain mod nph/sa/fr Gopt AntiDrift SR SR (from slopes) FWHM (mas) filter exp #frames disturb SN skip notes
20200208_113425 ??? 11.6 232931 4 0.77 0.10 161000 1 503 1107 600 0.9 0.9 -1.0 3 5.0 0.36 OFF 54.6 65.8 -1 H 300 1 ONSKY NO 0 -
20200208_113550 ??? 11.6 234004 3 0.82 0.10 161000 1 503 1107 600 0.9 0.9 -1.0 3 5.0 0.34 OFF 62.6 64.4 -1 H 300 1 ONSKY NO 0 -
20200208_113728 ??? 11.6 235244 3 0.79 0.10 161000 1 503 1107 600 0.9 0.9 -1.0 3 5.1 0.35 OFF 57.8 37.7 -1 H 300 1 ONSKY NO 0 -
IDL string: ['20200208_113425','20200208_113550','20200208_113728']
SR 0.583414 (0.0403854)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 0.794212 (0.0265412)
A
SR 0.65196 (0.0388512)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 1.28586 (0.0603798)

R = 12.8

  • 115332 -> 125
  • 115524 -> 126
  • 115709 -> 127
  • dark -> 128

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq emGain gain mod nph/sa/fr Gopt AntiDrift SR SR (from slopes) FWHM (mas) filter exp #frames disturb SN skip notes
20200208_115332 ??? 12.0 274839 5 1.11 0.10 161000 1 503 877 600 0.7 0.1 0.7 3 4.4 0.32 OFF 59.3 53.6 -1 H 300 1 ONSKY NO 0  
20200208_115524 ??? 12.0 276251 5 1.09 0.10 161000 1 503 877 600 0.7 0.1 0.7 3 4.4 0.31 OFF 56.6 52.0 -1 H 300 1 ONSKY NO 0  
20200208_115709 ??? 12.0 277580 4 0.95 0.10 161000 1 503 877 600 0.7 0.1 0.7 3 4.4 0.32 OFF 60.1 57.4 -1 H 300 1 ONSKY NO 0  
IDL string: ['20200208_115332','20200208_115524','20200208_115709']
SR 0.586472 (0.0185114)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 1.04916 (0.0827502)

SR 0 (NaN)
SEEING (WFS) 0.103093 (0)
SEEING (DIMM) 1.04916 (0.0827502)

R = 13.6
  • 120919 -> 129
  • 121025 -> 130
  • 131 loop data not saved
  • 132
  • dark -> 133
  • Problems with background subtraction...backgound has an average higher level compared to data cubes.

12:25 Can't get optical loop data working. Giving up for tonight and handing to Josh for PEPSI testing.

-- DougMiller - 08 Feb 2020
Topic revision: r10 - 09 Feb 2020, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback