20191118 SX SOUL Testing, specifically ESM

Binocular LUCI: AGw4 calibratoin on DX, ESM testing on SX

AO Table

00:30 The ESM (TTM_AO) Table I had created has been removed
  • found in on my laptop
  • copied it to AOeng@soul-sxwfs:~/aoroot/calib/wfs/current/ao/table_LUCIFER_TTM-AO.txt.20190407
  • made symbolic link from table_LUCIFER_TTM-AO.txt -> table_LUCIFER_TTM-AO.txt.20190407
  • Had to add GOpt->mode, amp and freq columns

01:30 The file aoroot/calib/wfs/current/setups/LUCIFER_TTM-AO had a tt table interpolation call which gave an error.
  • Copied LUCIFER_ACE-AO to LUCIFER_TTM-AO
  • Success

01:55 send ACE preset to BS9188 on-axis.
  • 12.4 mag
  • WFS see 14.2 mag.
    • I think off-axis probe is vignetting the AO FOV
  • Choose different off-axis guide star

02:02 sent ACE Preset
  • 12.4 mag
  • see 13.9 mag
  • change to bin 2
  • Optimize gain skip frame on HO2 setting. Twice.
  • Set gain to 4.0, 2.0, 2.0
  • TN
    • 022603 (no LUCI), no GOPT (will not run)
    • 023334, 023527, 023605 Gopt On, Gopt=0.07, no LUCI
  • Optimize Gain again

-- DougMiller - 18 Nov 2019