20190828 DX AdSec 10 iteration FF Matrix

10 Iteration FF Matrix

06:10 stop/start DX adsec
  • enable_lab_mode
  • Turn off AdSec Arbitrator process
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • AdOpt> tn=tracknum() & print, tn
  • AdOpt> print, ff_acquire_v2(tn)
    • ~33 minutes
  • IDL> print, reduce_ff_data('20190829_061602', /DO_PLOT)
    • Normalized Chi^2 ff_matrix: 165.45942 DOF: 3779136
    • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_061602/ff_+6.00e-05_3.40e-06m.sav
    • 4 images ff_10_1
    • act 101. 231 have slightly higher rms (2.88e-3 -vs- 2.11e-3)
  • ff_10_1_noise.jpg:
    ff_10_1_noise.jpg
  • ff_10_1_cov.jpg:
    ff_10_1_cov.jpg
  • ff_10_1_diff.jpg:
    ff_10_1_diff.jpg
  • ff_10_1_res.jpg:
    ff_10_1_res.jpg

07:15
  • enable_lab_mode
  • Turn off AdSec Arbitrator process
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Large current at act #4
  • print, fsm_rip()
    • Found misfunctioning capacitive sensors
    • 004:00835308um;
    • -10063
  • Mirror in failure
  • Click Recover Fail. In Use actuator is not working anymore
  • adsc_start/stop
    • same, act #4
    • Put in no position list

07:22 stop/start check. All OK
  • set_plot, "X"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_073035
    • ~ 50 minutes

08:00 lost communiation to AdSec
  • no fastdiagn (not updating)
  • No shapes on mirror
  • no log output

08:03 adsc_stop/start

08:10
  • set_plot, "X"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_080941
  • print, ff_acquire_v2(tn)
    • ~ 43 minutes

08:26 lost communiation to AdSec
  • no fastdiagn (not updating)
  • No shapes on mirror
  • no log output

08:29 adsc_stop/start

08:33
  • NO => set_plot, "X"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_083429
  • print, ff_acquire_v2(tn)
    • ~34 minutes
  • IDL> print, reduce_ff_data('20190829_083429', /DO_PLOT)
    • Normalized Chi^2 ff_matrix: 168.77971 DOF: 3767481
    • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_083429/ff_+6.00e-05_3.40e-06m.sav
    • 5 images ff_10_2

09:34
  • NO => set_plot, "X"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_093517
  • print, ff_acquire_v2(tn)
    • ~34 minutes
  • IDL> print, reduce_ff_data('20190829_093517', /DO_PLOT)
    • Normalized Chi^2 ff_matrix: 180.84846 DOF: 3767481
    • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_093517/ff_+6.00e-05_3.40e-06m.sav
    • 4 images ff_10_3

10:34
  • NO => set_plot, "X"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_103316
  • print, ff_acquire_v2(tn)
    • ~34 minutes
  • IDL> print, reduce_ff_data('20190829_103316', /DO_PLOT)
    • Normalized Chi^2 ff_matrix: 128.55017 DOF: 3767481
    • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_103316/ff_+6.00e-05_3.40e-06m.sav
    • 4 images ff_10_4

11:55 in confcalib cp data data.20190820 (date FF matrix was collect by Guido in the lab)
  • cd /local/aomeas/adsec_calib/ff_matrix/20190829_103316
  • cp *.fits /home/aoeng/confcalib/calib/adsec/672a/data/.
  • in confcalib cp data data.20190820 (date FF matrix was on telescope by Doug)

11:57 adsc_stop/start

12:00 New on-telescope 10 iteration ff Matrix
  • set_plot, "Z"
  • enable_lab_mode
  • AdOpt> print, fsm_set_flat(/FF,/NO_FLAT)
  • Turn off AdSec Arbitrator process
  • AdOpt> tn=tracknum() & print, tn
    • 20190829_120214
  • print, ff_acquire_v2(tn)
    • ~34 minutes
  • IDL> print, reduce_ff_data('20190829_120213', /DO_PLOT)
    • Normalized Chi^2 ff_matrix: 202.25952 DOF: 3767481
    • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_120214/ff_+6.00e-05_3.40e-06m.sav
    • 4 images ff_10_5

12:58 20 iteration
  • 20190829_125817
  • Normalized Chi^2 ff_matrix: 202.25952 DOF: 3767481
  • Data saved in /local/aomeas/adsec_calib/ff_matrix/20190829_120214/ff_+6.00e-05_3.40e-06m.sav
  • This FF data have not been processed yet
  • 4 images ff_25

-- DougMiller - 29 Aug 2019
I Attachment Action Size Date Who Comment
ff_10_1_cov.jpgjpg ff_10_1_cov.jpg manage 125 K 29 Aug 2019 - 07:11 DougMiller  
ff_10_1_diff.jpgjpg ff_10_1_diff.jpg manage 96 K 29 Aug 2019 - 07:11 DougMiller  
ff_10_1_noise.jpgjpg ff_10_1_noise.jpg manage 252 K 29 Aug 2019 - 07:10 DougMiller  
ff_10_1_res.jpgjpg ff_10_1_res.jpg manage 257 K 29 Aug 2019 - 07:12 DougMiller  
Topic revision: r5 - 29 Aug 2019, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback