DX Reconstructor Final

17:07
  • Waiting for telescope
17:56
  • Almost there........
18:10
  • Searching for source
  • Going to flatten mirror in bin1 and then measure bin4
  • source was at a much different position. -0.68 1.2 1.5 -420 -155 0
  • Actuator 479 is causing the flat to be applied with only 166 modes, huge bite out of the bottom.
  • idl_140257 |INF| 1384|2019-01-11 18:39:13.602258| MAIN > Maximum mode index used to apply flat -> # 166
  • New Flat: 20190111_400, applies with 641 modes
  • Put 479 back in
  • Only loads with 188, but the huge bite is smaller, maybe we can work with this.
  • New Flat: 20190111_400_2
  • idl_147091 |INF| 670|2019-01-11 19:01:02.897320| MAIN > Maximum mode index used to apply flat -> # 0
  • idl_147091 |INF| 698|2019-01-11 19:01:32.825579| MAIN > Maximum mode index used to apply flat -> # 350
  • New Flat: 20190111_400_3
19:04
  • Starting with Bin 4
  • Rock and Roll!
  • 10 mode recon, open loop.
  • Using doug_400_2l_0.6
  • Disturb: 20190111_190823
  • Intmat: 20190111_191137
  • Recon: Something wrong, I tried it twice, maybe bad disturbance file
  • Disturb: 20190111_191338
  • Intmat: 20190111_191408
  • Recon: 20190111_191537
  • Looks OK, now in closed loop
  • Intmat: 20190111_191758
  • Recon: 20190111_191921
  • 36 Modes
  • Disturb: 20190111_192138
  • Intmat: 20190111_192247
  • Recon: 20190111_192413
  • Looks fine.
  • Long Dataset, 48 iterations
  • Intmat: 20190111_192632
  • Recon: 20190111_193944
  • Looks good.
19:41
  • Back to bin1
  • idl_147091 |INF| 1012|2019-01-11 19:43:14.816111| MAIN > Maximum mode index used to apply flat -> # 343
  • Disturb: 20190111_195418
  • 153 mode flat: 20190111_153_1
  • 10 mode disturb: 20190111_205233 (open Loop)
  • 672_0.7.fits
  • Intmat: 20190111_205647
  • Recon: 20190111_205824
  • Closed Loop
  • Intmat:20190111_205943
  • Recon:20190111_210221
  • Try 36 mode
  • Disturb: 20190111_212350
  • Intmat: 20190111_212517
  • Recon: 20190111_212646
21:33
  • The new flat: 20190111_36_with_turbulence_1
  • /local/aomeas/adsec_calib/CMD/disturb/atm_1.6seeing_1000Hz.fits
  • The big bite is still there, but maaaaaybe a little smaller.
  • Disturb: 20190111_213606
  • Intmat: 20190111_213732
  • REcon: 20190111_213916
  • New Flat: 20190111_66_with_turbulence_1
  • Going to re-do the 66 mode, if this is a better flat
  • It's terrible, putting in the last good 400 mode flat: 20190111_400
  • 66 mode again: It gets some light in the big bite, but not completely, I wonder if there is something wrong with those actuators
  • Intmat: 20190111_214710
  • Recon: 20190111_214827
  • New Flat: 20190111_66_with_turbulence_2. That got a little bit of light into the bite
  • 153 Modes.
  • Disturb: 20190111_215238
  • Intmat: 20190111_215537
  • Recon:20190111_215731
  • Flat: 20190111_153_with_turbulence_1
  • Bite's almost gone:
  • 400
  • Disturb: 20190111_220201
  • Intmat: 20190111_220253
  • Recon: 20190111_220419
  • Flat: 20190111_400_4
  • Intmat:20190111_220809
  • Recon:20190111_220930
  • Unstable

-- GregoryTaylor - 11 Jan 2019
Topic revision: r2 - 11 Jan 2019, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.