DX On Sky

1:12
  • Preparing for preset.
  • Optimize gain failed
  • Manual gain setting failed because I clicked the button too quickly. I then stopped the loop incorrectly, I should have used the variable inspector.
  • I can't get the loop to re-close because of a slot-not found error. That usually requires a restart of the WFS.
2:01
  • Everything worked well this time.
  • Telemetry. GOPT and NCPA correction on.
  • 20171029_020903, 021018, 021055, 021125
2:19
  • Jochen is doing his first preset.
  • Successful
  • Dataset: 023047, 023106, 023135, 023155
2:48
  • Doug is working on SX.
  • Sun Oct 29 02:53:30.995 2017 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[90] - [AOException] presetAO: FATAL - Reading shared buffer masterdiagnostic.R:OPTLOOPBUF (-5001) TIMEOUT_ERROR
  • Outgoing.cpp:493: Ice::UnknownException: unknown exception: Traceback (most recent call last): File "aoarbitrator.py", line 965, in RequestAOStatus fstat=self._fsm.getState().getName().split('.') File "/home/aoeng/int4d/lib/python/statemap.py", line 104, in getState raise StateUndefinedException StateUndefinedException
  • Outgoing.cpp:493: Ice::UnknownException: unknown exception: Traceback (most recent call last): File "aoarbitrator.py", line 965, in RequestAOStatus fstat=self._fsm.getState().getName().split('.') File "/home/aoeng/int4d/lib/python/statemap.py", line 104, in getState raise StateUndefinedException StateUndefinedException
03:02
  • Successful preset on both sides.
  • We are still having trouble with SX.
  • The pupils are shifting.(on SX)
  • It seems that since we had to do some moving of the stages to center the same star as on DX, the offset of 0,0 moved the pupils.
03:32
  • Correcting pointing.
  • SX still went to the wrong star.
03:56
  • New preset to M2
  • Doug and I seem to have found different stars... I'm in binning 2, mag 12.6, he is in bin 3.
  • Seeing has worsened.
  • Datasets: 040748, 040811, 040839, 040857
  • GCS seems to be pointing the two different sides toward different off-axis guide stars.
  • Since we are in a globular cluster, the AO systems found guide stars anyway.
04:29
  • Had to re-close loop due to seeing.
  • Actuator 118 has given a warning, probably spurious.
  • There is a problem with vibration. Probably.
  • All of the LUCI images are elongated, on both LUCIs.
  • 1 CCD 47 image 044548
  • 5 CCD 47 image 044637
  • 10 CCD 47 Images 044659
05:14
  • Dataset to compare with vibration from before.
  • 10 CCD 47 frames: 051510
  • 5 CCD 47 frames: 051538
  • 1 CCD 47 frame: 051604
05:28
  • New preset
  • Bin 4, indicated magnitude is 15.1, Vs 14.4 in the script.
  • Gains: 3.40 1.18 0.90
  • Datasets. 053923, 054010, 054041
  • NCPA correction and GOPT on.
  • That caused the loop to become unstable.
06:06
  • We've redone the preset.
  • it's stable.
  • Bin 4, still, mag 15.
06:53
  • Loop Opened
  • The real time pupil system is having troubles for some reason.
  • The pupils as calculated by the CCD 39 software were wandering all over the place.
07:15
  • Had to restart the W-unit, the entire unit was unresponsive. I'm not sure why.
  • The AdSec was also unresponsive. I restarted it at 07:26
07:39
  • Changed thmin to 56 microns
  • /home/aoeng/int4d/idl/adsec_lib/adsec_utilities/short_dust_test.pro
08:02
  • Forgot to mention: Everything is now properly restarted.
  • We had an offset failure due to the loop opening for an unknown reason.
  • I think that the realtime pupil routine is getting confused in bin4. I've manually centered the lens and turned it off.
  • I've made some screenshots to capture the oddities that the realtime pupil routine is doing.
  • Andrew said something about problems that he was having with the same issue.
9:42
  • Still going.
10:01
  • New Preset.
  • Bin 1. 7.5 mag
  • Datasets: 101717, 101732, 101754, 101809
11:00
  • Still chugging along.
12:50
  • Finishing.