DX TT LUT with CalUnit, Arcetri Tests

13:40 Arrived in Remote Room. Telescope, AO and CalUnit ready

13:45 Power on WUnit

14:00 Network to the mountain really slow

14:11 error when setting flat

idl_59930          |INF|      3989|2017-10-23 14:11:31.660628|             MAIN > No misfunctioning capacitive sensors found.
idl_59930          |INF|      3990|2017-10-23 14:11:31.660912|             MAIN > No gap contamination found.
idl_59930          |INF|      3991|2017-10-23 14:11:31.661149|             MAIN > Dust test successfully completed
idl_59930          |INF|      3992|2017-10-23 14:11:31.661513|             MAIN > Update progress successfully completed (2)
idl_59930          |INF|      3993|2017-10-23 14:11:31.661652|             MAIN > FSM Rip done.
AdSecArbitrator_39082|ERR|       233|2017-10-23 14:11:31.664430|              IDL > Command returns error : IDL Ramp gain error: something is not going well (mirror stuck, gain too high...)
Last message repeated
20172310 14:11:31.6651 Alert: Warning: setFlatAo failed or syntax error occured
AdSecArbitrator_39082|INF|       234|2017-10-23 14:11:31.664677|   COMMANDHANDLER > Command execution failed: Fsm forced to Failure
AdSecArbitrator_39082|ERR|       235|2017-10-23 14:11:31.664802|             MAIN > SetFlatAo: FATAL - IDL Ramp gain error: something is not going well (mirror stuck, gain too high...)

14:15 Set flat again worked.

14:20 Fastlink connection works

14:35 Discussion
  • Accuracy of PSF measurement
    • Repeatability
    • 10 Sets of LUCI data
    • 5 frames each set
    • Before used OH1060/Z filters to get 3.5 mag of extinction
    • Try FeII with light source at "82" so get Source and corner alignment spots in one image

14:40 Mountain tip over to fill LBC and then tip up to fill LBTI

15:00 Back Zenith. LBTI fills

  • FeII /clear
  • 2.5 sec int
  • TN 164602
  • LUCI 280
  • LUCI Dark 281 164944

17:00 Find we need -0.8 rotation on re-rotator offset
  • Edit ./conf/wfs/current/processConf/wfsarb.conf
  • stop/start wfsarb
  • correct offset

17:01 New flat
  • bin1, 990 Hz
  • 20171023_400_1.sav
  • 603 modes applied

18:37
  • bin1, 990 Hz
  • rotator 210
  • K/Clear
    • Source ~1500, corners
  • 10 sec integration, cube of 2
  • 183724 (302), 183802 (303), 184011 (304), 184039 (305)
  • Dark 184505 (306)

18:51 Start collection 180 rotation
  • 185108 (308), 185213 (309), 185240 (310)
    • Brakes Set
    • Open loop rotation -180 (210->30->210)
  • 185714 (311), 185756 (312), 185823 (313)
    • Brakes Set
    • Open loop rotation +180 (210->390->210)
  • 190403 (001), 190508 (002), 190533 (003)
    • Brakes Set
    • Open loop rotation -180 (210->30->210)
  • 191544 (004), 191608 (005), 191632 (006)
    • Brakes Set
    • Open loop rotation +180 (210->390->210)
  • 192113 (007), 192135 (008), 192157(009)

19:45
  • bin 4, 200 Hz, 12 modes
  • HKspec

19:58 Start collection 180 rotation
  • 195748 (019), 195834 (020), 195948 (021), 200018 (022), 200048 (023)
    • Brakes Set
    • Open loop rotation -180 (210->30->210)
  • 202049 (028), 202145 (029), 202208 (030)
    • Brakes Set
    • Open loop rotation +180 (210->390->210)
  • 202821 (032), 202855 (033), 202922 (034)
    • Brakes Set
    • Open loop rotation -180 (210->30->210)
  • 203615 (036), 203648 (037), 203720 (038)
    • Brakes Set
    • Open loop rotation +180 (210->390->210)
  • 204200 (039), 204634 (040), 204714 (041)

20:04 Dark
  • 2.5 Sec
  • 200458 (025)

20:48 Stages in closed loop
  • 204858 (042), 204938 (043), 205006 (044)
    • Stages closed loop
    • Open loop rotation -180 (210->30->210)
  • 205534 (045), 205618 (046), 205639 (047)
    • Stages closed loop
    • Open loop rotation +180 (210->390->210)
  • 210145 (048), 210209 (049), 210243 (050)
    • Stages closed loop
    • Open loop rotation -180 (210->30->210)
  • 210707 (051), 210729 (052), 210808 (053)
    • Stages closed loop
    • Open loop rotation +180 (210->390->210)
  • 211148 (054), 211210 (055), 211236 (056)

21:14 Hand over for OH106 + K
  • 2.5s Dark 20171023_212522 (0057)
  • 20s Dark 20171023_213008 (0058)
Angle OH1060 +z 2.5s K 20s Comment
210 214706 (0059) 214918 (0060) Starting At 210
210 215718 (0063) 215841 (0064) To 30 and Back to 210
210 220731 (0066) 220411 (0065) To 390 and Back to 210
210 221339 (0067) 221527 (0068) 30 and back
210 222238 (0070) 222044 (0069) 390 and back