DX ARGOS TT LUT with CalUnit and Dichroic

08:15 MST Telescope, ARGOS and AO all set up and ready to find light source

08:16 DX CCD47 4x4 live display has a lot of pattern noise
  • Dark 20171016_151647 (Bayside stages powered off)

09:22 From April from ARGOS log (from Gustavo)
  • hexapod -2.04, -4.58, 1.25, 255, -18
  • bayside -73.3, -47.7, 55.7

10:07 Aligned source on LUCI and then on CCD47. Remove as much coma as I could
  • hexapod -3.4, -2.3, 1.15, -243, -360
  • bayside values from above
  • ready to close the loop

10:23 No slopes arriving at DX AdSec.
  • stopped and started BCU interface and Slope Computer processes. No slopes
  • Should have powered off BCU

10:26 Clicked operate button
  • Error could not connect to BCU47 timeout
WfsArbitrator_52910|ERR|        33|2017-10-16 17:24:49.323040|              WFS > Command returns error : Timeout waiting for pinger.R.bcu47 (-5001) TIMEOUT_ERROR
WfsArbitrator_52910|INF|        34|2017-10-16 17:24:49.323106|   COMMANDHANDLER > Command execution failed: Fsm forced to Failure
WfsArbitrator_52910|ERR|        35|2017-10-16 17:24:49.323150|             MAIN > Operate: FATAL - Timeout waiting for pinger.R.bcu47 (-5001) TIMEOUT_ERROR
  • Power was on and all processes were on
  • Manually powered off and w_stop/start

10:33 Now get can not connect to TS8!
WfsArbitrator_65568|ERR|        32|2017-10-16 17:31:50.423912|              WFS > Command returns error : Missing network connection with ts8 dx. Check network cable and 110V line (-5001) TIMEOUT_ERROR
WfsArbitrator_65568|INF|        33|2017-10-16 17:31:50.423967|   COMMANDHANDLER > Command execution failed: Fsm forced to Failure
WfsArbitrator_65568|ERR|        34|2017-10-16 17:31:50.423994|             MAIN > Operate: FATAL - Missing network connection with ts8 dx. Check network cable and 110V line (-5001) TIMEOUT_ERROR
  • Turned Box47 Power off
  • Tried Operation again. Same error
  • Turned Box47 Power off
  • w_stop/start
  • Operate

10:40 Elliott will unplug and plug 110 V power
  • Before hand, Chris powered off AGw2 off-axis. He will power on and restart OAC server when done.

10:56 Power cycled by Elliott
  • Now powers on correctly.

11:29 Closed loop with 400 modes * New Flat 20171016_ARGOS_Dichroic * Hexapod -3.44, -2.34, 1.07, -247, -382 * bayside -75, -53, 55.7 * On N30 source at 1126, 868

12:30 Start collecting data N30
  • 2.5 sec dark 0029
Angle Off On Z Comment
225 0002 0003 0004  
180 0007 0006 0005  
135 0008 0009 0010  
90 0014 0015 0016  
45 0019 0018 0017  
0 0020 0021 0022  
-45 0025 0024 0023  
-90 0026 0027 0028  

01:30 Start collecting data N3.37
  • bin 2, 400 Hz
  • 10.5 mag (120 counts)
  • 2.5 sec dark 0029
-90 0044 0043 0045  
-45 0048 0047 0046  
0 0049 0050 0051  
45 0055 0059 0058  
90 0065 0064 0063  
135 0066 0067 0068  
180 0071 0070 0069  
225 0077 0076 0075  
270 0078 0079 0080  

03:05 Start collecting TT calibration data
  • bin 2
  • 400 Hz

X Y LUCI
0.0 0.0 109
0.0 -1.5 108
0.0 -3.0 107
0.0 -1.5 106
0.0 0.0 105
0.0 1.5 104
0.0 3.0 103
0.0 1.5 102
0.0 0.0 101
-1.5 0.0 100
-3.0 0.0 99
-1.5 0.0 98
0.0 0.0 97
1.5 0.0 96
3.0 0.0 95

-- %USERSIG{DougMiller - 2017-10-16}%