DX Commissioning on Sky

03:05
  • Opening after delay for wind.
  • Document restart procedures.
  • IDL controller off, Fastdiagn off, IDL on Fast on, restart arbitrator.
03:36
  • First preset failed, typo in script.
03:38
  • New preset.
  • Add how to adjust gain manually.
  • High forces on 460 and 462, probably due to a mismatch in the reconstructor and the FF matrix.
  • I'm going to temporarily take out #460 and will see what #462 does.
04:17
  • Redoing preset.
  • A lot of safeskips
  • Optical Loop Data:
  • 050311, 050338, 050407, 050428
  • Still getting high currents on #462.
  • I'm not sure if that is because the actuator is bad or the FF matrix is not well matched to the reconstructor.
  • If it's safe skipping because of a single actuator, why can't it just ignore that one.
  • #351 is also having issues.
  • If using the AOS command GUI manually, remember to hit Start AO.
  • DSP Write Failure.
05:36
  • Doing a preset into bin 1 to test the SX reconstructor.
  • Dataset: 054810
06:17
  • Fri Sep 22 06:17:49.071 2017 aos.AcquireRef{FLAO}.failed - AcquireRef {FLAO}[541] - [AOException] RETRY: Error in acquire ref: Star too bright: ccd39 is saturated, AO loop cannot work. - code: 0 NO_ERROR (-1000) GENERIC_SYS_ERROR
  • Closed due to wind.

-- %USERSIG{GregoryTaylor - 2017-09-22}%

Comments

No Comment
Topic revision: r2 - 22 Sep 2017, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.