DX Reconstructor Measurements

17:00
  • Searching for Source
18:14
  • Making flats
  • Vgood 20170612_300_flat_2.sav
  • 20170612_300_flat_3.sav
18:18
  • Starting Reconstructors
  • 10 Modes, open loop
  • Disturb 182112,
  • Amplitude file: optimized_amano_6l.fits, 48 push-pull cycles
  • Intmat: 182353
  • Recon: 183208
  • 10 Modes in closed loop with 183208 as recon
  • same disturb, gain 0.05
  • Intmat: 183435
  • Recon: 183638
  • 36 mode disturbance, same Amplitude file, 13 push-pull cycles
  • Disturb: 184129
  • Closed loop with previous recon
  • Intmat: 184226
  • Recon: 184406
  • 66 mode disturb, same Amplitude file, 4 frames per movement, 7 push-pull cycles
  • Disturb 184753
  • Intmat: 184918
  • Recon: 185050
  • 153 mode disturb 3 push-pull cycles
  • Disturb: 185620
  • Intmat: 185724
  • Recon: 185909
  • New Flat: 20170612_300_flat_4.sav
19:05
  • 400 Mode Disturb, 2 frames/movement, 2 push-pull, same amplitude file
  • Disturb: 190705
  • Intmat: 190820
  • Recon: 190946
  • New flat, using/local/aomeas/adsec_calib/M2C/KL_v16/RECs/Rec_20150125_161348.fits, an old recon
  • 20170612_300_flat_5.sav, much better!
19:17
  • Iterating,
  • moved cube stage to -5.4 because of dust
  • Same disturb, intmat in closed loop with 190946
  • Lost the bottom edge, I'm going to try and use 20150125_161348, to take a new intmat
  • This keeps the bottom under control.
  • Intmat: 192401
  • Recon: 192626
19:38
  • Moved cube to -2.5, vastly improving dust situation
  • New cube angle: 203.636
  • New flat wit latest recon: 20170612_400_flat
19:44
  • Iterating
  • Intmat: 194556
  • Recon: 194747
  • Flat 20170612_400_flat_2
19:55
  • Cube to -2.2
  • rotator 203.727
  • Intmat: 200014
  • Recon: 200205
  • Flat: 20170612_400_disturb
20:11
  • Intmat: 201249
  • Recon: 201442
  • Flat: 20170612_400_flat_3
20:21
  • 48 iteration intmat
  • Intmat: 202238
  • Recon: 203644
20:43
  • Testing with disturb: /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.2_L040.0_v15.0_ovfreq1000.00_sd3892.fits
  • Dataset: 204535, 204625, 204641, 204655
20:47
  • Disturb: /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits
  • Datasets: 205035, 205115, 205130, 205146
22:05
  • Datasets with 20 luci images and 1/2NCPA magnitude with disturb:/local/aomeas/adsec_calib/M2C/KL_v16/disturb/CmdDisturb_20170612_214846.fits, which is the same disturb as before, but with a sinus mode added.
  • Datasets: 221513, 221532, 221604, 221624
22:24
  • Dataset: /local/aomeas/adsec_calib/M2C/KL_v16/disturb/CmdDisturb_20170612_222150.fits, a 0.8, arcsec disturb
  • No NCPA: 222518, 222541, 222609, 222626,
  • NCPA: 223215, 223231, 223247, 223307
22:50

  • New disturb with: 600_optimized_6l.fits, 400 modes, 2 frames, 2 push-pull.
  • Disturb: 225127
  • 4 iterations
  • Intmat: 225941
  • Recon: 230117

Comments:

No Comment

-- %USERSIG{GregoryTaylor - 2017-06-12}%

Comments:

No Comment
Topic revision: r5 - 13 Jun 2017, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback