SX Performance Measurement with RR

Comments

Hopefully the Table of Contents for this Wiki page clearly shows the tests we performed today to see if the performance of the 400 mode reconstructor improves with either a slight change in re-rotator position or slight shift in the CCD39 pupil shift.

We found no change in performance when we either rotated the re-rotator nor when we shifted the pupils on CCD39. This can be seen in the modal plot for each data set.

At the end of the day (bottom of the TWiki) we collected loop telemetry for the 300 mode reconstructor. Comparing the modal plot for 300 modes to the modal plot from data collected on 20160229, the day the 400 mode Interaction Matrix and Reconstructor were collected and calculated (shown at the very top of the TWiki page), shows very similar performance for the first 300 modes. Today, when we use 400 modes, we see the decrease in performance at all modes. This is consistent with the results we saw on-sky during LUCI1/AO commissioning.

For all the data sets we have include a jitter plot showing the strong uncorrected vibration at 14 and 15 Hz. In these plot we include the PSD for the first tracking number and the last tracking number in the set to see if the vibration conditions change while we were collecting data. In general, conditions did not change all day.

For the 300 mode reconstructor we created a jitter plot out to 70 Hz. We then found a strong vibration at 45 Hz and 60 Hz. Are these the vibrations that are causing our poor performance of the 400 mode reconstructor?

We do not see a strong 60 Hz vibrations on the OVMS system (OVMS 60 Hz power lower by an order of magnitude). Is this vibrations new in the last months and internal to the AO system?

John Hill reported that tonight LBTI does not see any strong 60 Hz vibration on-sky with their AO Systems. I have asked Amali to collect a couple sets of telemetry data on both sides of LBTI to compare to our daytime FLAO telemetry.

If this 60 Hz vibration is the problem, and internal to the AO system, the LBTI result implies it is not in the SX AdSec. Thus, is this a problem with the SX WUnit? Is it a mechanic vibration (failing power supply) or electronic noise on CCD39 at 60 Hz?

I have not had a chance look at the jitter plot of the telemetry data taken on 20160229, when the IM and RM were collected/created. I think these data are on the old CentOS 5 machine, adsecsx. I will search for these data tomorrow.

In the afternoon we tested that the operation of all the routines to collect a new interaction matrix and calculate a new reconstructor matrix work properly. They do. We collect a new 400 mode IM and calculated a new 400 mode RM. The performance for this new RM is about the same as with the original 400 mode RM. See http://wiki.lbto.org/bin/view/AdaptiveOptics/20171511_SX_Recon

Note: all of today's test and the measurement of a new IM an RM were performed with an SX AdSec actuator list that is the same as the reconstructor used. Specifically, the SX AdSec configuration had 29 actuator in the do-not-use list (total of curr and pos) which is the same as the SX AdSec configuration when the 400 mode reconstructor was measure on 20160229.

Doug

Good Performance of 400 Mode Reconstructor on 20160229

This is the measured performance on 20170229 of the currently used 400 mode reconstructor that we are using on-sky and are testing today (see results below). The optical setup today is the same as on 20160229.
  • Bin 1
  • RetroReflector
  • 400 Modes
  • 7.5 mag
  • dist_flao2_KL_v18_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • Gain 0.75, 0.5, 0.45
with the slight difference in the applied disturbance. The seeing value today and on 20160229 are the same, but today we are using the KL_v20 version of the disturbance.
  • dist_flao2_KL_v20_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd3892.fits

185606_new_bin1_meas_rec.jpg

400 Mode Reconstructor

Re-Rotator at 54.836 and Camera lens Centered

17:30 Source found, closed loop, ready to collect data
  • See 14 Hz and 15 Hz vibrations, like yesterday. 15 Hz is stronger

17:45 Initial data
  • Closed Loop
  • 0.8" seeing
  • Opt Gain 0.67, 0.49, 0.39
  • Data
    • 174534 - 180258

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_174534 ??? 7.8 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 436 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180033 ??? 7.8 324000 6 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 435 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180107 ??? 7.8 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 442 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180157 ??? 7.8 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 439 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180217 ??? 7.8 324000 5 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 438 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180229 ??? 7.8 324000 5 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 440 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180240 ??? 7.8 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 440 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_180258 ??? 7.8 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 440 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_174534_modal.jpg

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_174534_jitter.jpg

Re-Rotator test

18:21 Curent position 54.836
  • +0.4 degrees see pedal in the pupil
  • -0.3 degrees see about the same in the pupil

Re-Rotator position to 54.936

18:23 Set position to 54.936
  • Data
    • 182415-182416

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_182315 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 424 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_182341 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 426 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_182405 ??? 7.9 324000 5 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 423 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_182416 ??? 7.9 324000 5 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 424 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.936:
    20170511_182315_modal.jpg

  • Re-Rotator at 54.936:
    20170511_182315_jitter.jpg

Re-Rotator position to 55.036

18:36 Set position to 55.036
  • Data
    • 183657-183830

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_183657 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 424 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_183713 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 422 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_183736 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 421 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - long stop of 53 frames at frame 2198
20170511_183830 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 423 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 55.036:
    20170511_183657_modal.jpg

  • Re-Rotator at 55.036:
    20170511_183657_jitter.jpg

Re-Rotator position to 54.736

18:44 Set position to 54.736
  • Data
    • 184422-184457

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_184422 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 424 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184435 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184446 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 426 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184457 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.736:
    20170511_184422_modal.jpg

  • Re-Rotator at 54.736:
    20170511_184422_jitter.jpg

Re-Rotator position to 54.636

18:48 Set position to 54.636
  • Data
    • 184855-184932

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_184855 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 428 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184908 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184920 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_184932 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.636:
    20170511_184855_modal.jpg

  • Re-Rotator at 54.636:
    20170511_184855_jitter.jpg

Re-Rotator at 54.836 and Camera lens Centered

18:48 Set position to 54.836
  • Data
    • 185844-185951

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_185844 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 427 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_185856 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 427 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_185912 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 428 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_185924 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 425 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_185939 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 426 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_185951 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 427 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_185844_modal.jpg

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_185844_jitter.jpg

Camera Lens Position Test

19:05 Camer Lens at Y=-23 um
  • -15 um stripes
  • +9 um strips

Camera lens dY=-12 um

19:11 Camera lens dY=-12 um
  • Data
    • 191113-191229

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_191113 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_191131 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 432 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_191148 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 427 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_191229 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target

  • Camera Lens dY=-12 um:
    20170511_191113_modal.jpg

  • Camera Lens dY=-12 um:
    20170511_191113_jitter.jpg

Camera lens dY=+6 um

19:18 Camera lens dY=+6 um
  • Data
    • 191809-191928

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_191809 ??? 7.8 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 433 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_191821 ??? 7.8 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 433 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_191917 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 430 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_191928 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 430 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Camera Lens dY=+6 um:
    20170511_191809_modal.jpg

  • Camera Lens dY=+6 um:
    20170511_191809_jitter.jpg

Camera lens dX=+3 um

19:18 Camera lens dX=+3 um
  • Data
    • 192827-192905

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_192827 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_192841 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_192852 ??? 7.9 324000 4 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_192905 ??? 7.9 324000 3 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 428 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Camera Lens dX=+3 um:
    20170511_192827_modal.jpg

  • Camera Lens dX=+3 um:
    20170511_192827_jitter.jpg

Camera lens dX=-8 um

19:32 Camera lens dX=-8 um
  • Data
    • 193249-193347

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_193249 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 430 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_193322 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 426 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_193335 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 424 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target
20170511_193347 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 431 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - Pupils not on target

  • Camera Lens dX=-8 um:
    20170511_193249_modal.jpg

  • Camera Lens dX=-8 um:
    20170511_193249_jitter.jpg

Re-Rotator at 54.836 and Camera lens Centered

19:36 Re-Rotator at 54.836 and Camera lens Centered
  • Data
    • 193829-193905

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_193829 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_193842 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 429 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_193853 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 430 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_193905 ??? 7.9 324000 2 NaN -1.00 184300 1 400 990 0.7 0.5 0.4 3 431 ON -1.0 ? -1 -1 ON NO - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_193829_modal.jpg

  • Re-Rotator at 54.836 and Camera lens Centered:
    20170511_193829_jitter.jpg

* 400 Mode Recon:
20170511_193829_70_jitter.jpg

300 Mode Reconstructor Test

Re-Rotator at 54.836 and Camera lens Centered

22:45 Closed Loop
  • 300 Modes
  • 29 unused actuator (same as Reconstructor)
  • 0.8"
  • Gain 0.67, 0.51, 0.48
  • Data
    • 224401 - 224627

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20170511_224513 ??? 7.9 324000 3 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 432 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_224529 ??? 7.9 324000 3 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 429 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_224543 ??? 7.9 324000 1 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 431 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_224555 ??? 7.9 324000 2 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 430 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_224616 ??? 7.9 324000 2 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 431 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
20170511_224627 ??? 7.8 324000 5 NaN -1.00 062006 1 300 990 0.7 0.5 0.5 3 434 ON -1.0 ? -1 -1 ON NO - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error

  • 300 Mode Recon:
    20170511_224513_modal.jpg

  • 300 Mode Recon:
    20170511_224513_jitter.jpg

-- %USERSIG{DougMiller - 2017-05-11}%
I Attachment Action Size Date Who Comment
20170511_174534_jitter.jpgjpg 20170511_174534_jitter.jpg manage 119 K 11 May 2017 - 18:32 DougMiller Re-Rotator at 54.836 and Camera lens Centered
20170511_174534_modal.jpgjpg 20170511_174534_modal.jpg manage 112 K 11 May 2017 - 18:31 DougMiller Re-Rotator at 54.836 and Camera lens Centered
20170511_182315_jitter.jpgjpg 20170511_182315_jitter.jpg manage 101 K 11 May 2017 - 18:40 DougMiller Re-Rotator at 54.936
20170511_182315_modal.jpgjpg 20170511_182315_modal.jpg manage 112 K 11 May 2017 - 18:40 DougMiller Re-Rotator at 54.936
20170511_183657_jitter.jpgjpg 20170511_183657_jitter.jpg manage 112 K 11 May 2017 - 18:47 DougMiller Re-Rotator at 55.036
20170511_183657_modal.jpgjpg 20170511_183657_modal.jpg manage 101 K 11 May 2017 - 18:46 DougMiller Re-Rotator at 55.038
20170511_184422_jitter.jpgjpg 20170511_184422_jitter.jpg manage 116 K 11 May 2017 - 18:58 DougMiller Re-Rotator at 54.736
20170511_184422_modal.jpgjpg 20170511_184422_modal.jpg manage 104 K 11 May 2017 - 18:57 DougMiller Re-Rotator at 54.736
20170511_184855_jitter.jpgjpg 20170511_184855_jitter.jpg manage 101 K 11 May 2017 - 19:04 DougMiller Re-Rotator at 54.636
20170511_184855_modal.jpgjpg 20170511_184855_modal.jpg manage 109 K 11 May 2017 - 19:03 DougMiller Re-Rotator at 54.636
20170511_185844_jitter.jpgjpg 20170511_185844_jitter.jpg manage 108 K 11 May 2017 - 19:22 DougMiller Re-Rotator at 54.836 and Camera lens Centered
20170511_185844_modal.jpgjpg 20170511_185844_modal.jpg manage 101 K 11 May 2017 - 19:22 DougMiller Re-Rotator at 54.836 and Camera lens Centered
20170511_191113_jitter.jpgjpg 20170511_191113_jitter.jpg manage 98 K 11 May 2017 - 19:28 DougMiller Camera Lens dY=-12 um
20170511_191113_modal.jpgjpg 20170511_191113_modal.jpg manage 103 K 11 May 2017 - 19:28 DougMiller Camera Lens dY=-12 um
20170511_191809_jitter.jpgjpg 20170511_191809_jitter.jpg manage 96 K 11 May 2017 - 19:36 DougMiller Camera Lens dY=+6 um
20170511_191809_modal.jpgjpg 20170511_191809_modal.jpg manage 97 K 11 May 2017 - 19:36 DougMiller Camera Lens dY=+6 um
20170511_192827_jitter.jpgjpg 20170511_192827_jitter.jpg manage 104 K 11 May 2017 - 19:44 DougMiller Camera Lens dX=+3 um
20170511_192827_modal.jpgjpg 20170511_192827_modal.jpg manage 101 K 11 May 2017 - 19:44 DougMiller Camera Lens dX=+3 um
20170511_193249_jitter.jpgjpg 20170511_193249_jitter.jpg manage 97 K 11 May 2017 - 20:24 DougMiller Camera Lens dX=-8 um
20170511_193249_modal.jpgjpg 20170511_193249_modal.jpg manage 98 K 11 May 2017 - 20:24 DougMiller Camera Lens dX=-8 um
20170511_193829_70_jitter.jpgjpg 20170511_193829_70_jitter.jpg manage 98 K 12 May 2017 - 07:34 DougMiller 400 Mode Recon
20170511_193829_jitter.jpgjpg 20170511_193829_jitter.jpg manage 102 K 11 May 2017 - 19:54 DougMiller Camera Lens dX=-8 um
20170511_193829_modal.jpgjpg 20170511_193829_modal.jpg manage 101 K 11 May 2017 - 19:54 DougMiller Camera Lens dX=-8 um
20170511_224513_jitter.jpgjpg 20170511_224513_jitter.jpg manage 122 K 12 May 2017 - 07:28 DougMiller 300 Mode Recon
20170511_224513_modal.jpgjpg 20170511_224513_modal.jpg manage 111 K 12 May 2017 - 07:28 DougMiller 300 Mode Recon
Topic revision: r3 - 12 May 2017, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback