DX Adsec Actuators

18:35
  • Starting again after crash.
  • GOPT gain 0.35
  • Gains 0.44 0.23 0.23
  • print, sinusmode_generate(4, 10e-9, 30, 990, "KL_v16", /even, add_file="/local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.6_L040.0_v15.0_ovfreq1000.00_sd1983.fits")
  • CmdDisturb _20170509_173751.fits
  • Dataset: 174246
  • Gains 0.44, 0.23, 0.13
  • Dataset 174641
  • Gains 0.44 0.23 0.33
  • Dataset 174916

18:46
  • GOPT: 0.32
  • gains 0.40 0.28 0.27
  • Dataset: 184715

18:51
  • Putting actuators that are missing from the reconstructor into "no curr" list.
  • 441, 552, 262, 263, 260, 58, 169, 279, 404, 501, 573, 663
  • Gains: 0.94 0.69 0.70
  • GOPT gain 0.40
  • Dataset: 190448

19:06
  • Remove the above from current.
  • ...
  • Try with NCPA, what the heck.
  • Gopt 0.40
  • Gains 0.42 0.27 0.27
  • Dataset: 191642

20:50
  • Saving 50 Darks for LUCI
  • 128*128 subwindow
  • FeII and Clear
  • 0.157 sec exposure
  • GOPT off
  • Gains 1.0 0.65 0.65
  • Dataset w/ 50 LUCI frames: 205232

Tracknum 20170509_205232
Closed Loop Yes
Is OK? No - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 104.617256
Magnitude 9.408589
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.651000 1.000000
Telescope elevation 324000.062808
Wind speed 0.000000
seeing from OL modes 0.768140
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.571865
LUCI dark 20170509_205008.fits_cube.fits
  • GOPT gain: 0.39
  • Gains 0.39 0.26 0.26
  • Dataset w/ 50 LUCI frames: 205531

Tracknum 20170509_205531
Closed Loop Yes
Is OK? No - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 104.686211
Magnitude 9.407873
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.255116 0.391884
Telescope elevation 324000.062934
Wind speed 0.000000
seeing from OL modes 0.805314
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.569203
LUCI dark 20170509_205008.fits_cube.fits

  • NCPA On
  • Dataset w/ 50 LUCI frames: 205709
  • NCPA off 212347
  • NCPA on 212640, vibrations

Tracknum 20170509_205709
Closed Loop Yes
Is OK? No - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 104.407364
Magnitude 9.410769
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.255116 0.391884
Telescope elevation 324000.062577
Wind speed 0.000000
seeing from OL modes 0.809638
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.677619
LUCI dark 20170509_205008.fits_cube.fits

Tracknum 20170509_212347
Closed Loop Yes
Is OK? No - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 104.238388
Magnitude 9.412528
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.261364 0.401480
Telescope elevation 324000.064503
Wind speed 0.000000
seeing from OL modes 0.804359
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.573441
LUCI dark 20170509_210626.fits_cube.fits

Tracknum 20170509_212640
Closed Loop Yes
Is OK? No - FLTimeout error - PendingCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 104.104156
Magnitude 9.413927
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.261364 0.401480
Telescope elevation 324000.064088
Wind speed 0.000000
seeing from OL modes 0.813920
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.534905
LUCI dark 20170509_210626.fits_cube.fits
21:44
  • Now with the old supervisor.
  • Sets with 135 modes
  • Gains: 1.16 0.60 0.69
  • Dataset: 225046

Tracknum 20170509_225046
Closed Loop Yes
Is OK? No - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error - long stop of 478 frames at frame 3231
AntiDrift ON
nphotons/sub/fr 108.308464
Magnitude 9.370941
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.599000 1.159400
Telescope elevation 324000.430952
Wind speed 0.000000
seeing from OL modes 0.752806
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.618745
LUCI dark 20170509_224636.fits_cube.fits

  • GOPT on, Gain 0.30
  • Dataset: 225154

Tracknum 20170509_225154
Closed Loop Yes
Is OK? No - CRC error - FLTimeout error - PendingCounter error - SkipCounter error - TimeStamp error - WFSGlobalTimeout error
AntiDrift ON
nphotons/sub/fr 107.534859
Magnitude 9.378724
# Modes 400
Modal rec Rec_20130919_014135.fits
Binning 1
Frequency [Hz] 989
Modulation 2.996182
FW1 Dichroic 600-1000 nm
FW2 FB850-10
Gain minmax 0.191820 0.371280
Telescope elevation 324000.431415
Wind speed 0.000000
seeing from OL modes 0.874960
lambda [um] 1.646000
exptime [s] 0.156748
framerate [Hz] 6.379667
no. frames 50
SR SE 0.614160
LUCI dark 20170509_224636.fits_cube.fits

  • NCPA on:
  • Dataset: 225246