DX Actuator Check

Hex -3.440 -1.30 0.985 -108.797 -192.130

18:44
  • Reverting to most complete actuator map, from Thursday at about 20:18
  • Dataset 185507
  • Flat: test_185507_400.sav
19:00
  • Removing all but the inner ring of actuators from the "no icc" list
  • Gains: 1.00 0.30 0.30
  • Dataset: 190731
  • Flat: test_190731_400.sav
  • Test with Disturb: /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.6_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • Gains: 1.23 0.70 0.90
  • Dataset: 191344
  • Odd bump in modal plot b/c gains too high.
  • Gains 1.23 0.70 0.70
  • Dataset: 191617
19:17
  • Some actuators causing high forces and bumps, will try them in no icc.
  • 279, 404, 501, 573, 58, 169
  • Dataset: 192657
  • Flat: test_192657_400.sav
  • Disturb: /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.6_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • Autogains: ho2 still too hight, 1.23 0.74 1.45
  • gains: 1.23, 0.74, 0.75
  • Dataset: 193156
  • gains: 1.23 0.74, 0.65
  • Dataset: 193340
19:45
  • 455 and 469 possibly causing a bump in "Delta Command", removing from icc (adding to no icc list)
  • No change, I took them off of the "no icc" list.
19:55
  • Disturb: /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.6_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • Gains: 1.23 0.70 0.70
  • Dataset: 200055
  • I do not understand why it levels off between 100 and 300 modes, but there is improvement between 300 and 400 modes.

20:54
  • Leaving the loop closed to make sure that it is stable.
22:27
  • Removed actuators that were not in the Reconstructor.
  • /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao1_ts4_atm_s1.6_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • Dataset: 222745, 223508, 224847
  • Zero gain: 223702
-- %USERSIG{GregoryTaylor - 2017-05-08}%
Topic revision: r3 - 08 May 2017, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.