DX Flat Repair

15:05 Take over telescope

16:30 Loop closed ready to test
  • Had to stop/start WFS SW because slope were not arriving at AdSec

16:35 Setup
  • Hex -2.1, -3.7, 0.97, 123, -50, 0
  • bayside -78.87, -48.73, 55.1

16:40 Original Performance
  • 300 mode
    • Rec 20150128_012542
    • autogain: 1.23, 0.5, 0.7
    • 164242
    • 164242_modalplot.jpg: looks good, but terrible modal plot
    • 164242_mirror_300.jpg
    • 165119: looks good, but terrible modal plot
      • increase high 1,2 gain but no improvement
      • is this the bump at 5:00?
    • flat test_164242_300.sav
    • OL default flat loaded
      • 135 modes applied
      • flat_default.jpg
      • wavy
    • OL test_164242_300 applied
      • 84 modes applied
      • flat_164242_300.jpg
      • wavy

17:14
  • 400 mode
    • Rec 20130919_014135
    • could not use auto gain, goes into safeskip
    • gain 1.25, 0.35, 0.2
      • command in bump at 5:00 gets big for HO1 gains. Not so much for HO2
    • 171443
    • 171442_mirror_400.jpg

17:49 Change act list
  • Add 465, 468, 471, 478, 484 back into list (bump @ 5:00)
  • flat_default_add_5_at_5_pos.jpg
  • 105 modes applied

17:58 CL
  • 300 mode
  • gains 1.15, 0.5, 0.35
  • 180208
  • 180208_modalplot.jpg
  • 180208_mirror_300
  • flat test_180208_300
    • load, 230 modes

18:12 CL
  • 300 modes
  • autogain 1.7, .28, .2
    • 181505
    • 181505_mirror_300.jpg
    • 181505_modalplot.jpg
  • manual gain 1.7, 0.5, 0.4
    • 181735
    • modal plot no better
  • Flat test_181505_300
    • 233 modes applied

18:29 CL
  • 400 modes
  • manual gain 1.2, 0.5, 0.3
  • 182906
  • 182906_mirror_400.jpg
  • flat test_182906
    • load with 238 modes * auto gain 1.2, 0.2, 0.4
    • 183933

18:58 CL with disturbance
  • /local/aomeas/adsec_calib/CMD/disturb/old_KL_v16/dist_flao_ts4_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • auto gain 1.23, 0.69, 0.68
  • 190252
  • 190252_mirror_400.jpg
  • flat test_190252_400
    • 207 modes applied

19:37 CL w/o disturbance
  • 400 modes
  • 193707
  • 193707_mirror_400.jpg
  • flat test_193797_400
    • 176 modes applied in flat

19:43 CL
  • 300 modes
  • gain 1.2, 0.4, 0.3
  • 194311
  • flat test_194311_300
    • 219 modes applied
    • 243 modes applied

19:50
  • add 338, 343, 351, 352, 453, 455, 460, 461
  • problem with skip frames. Add 461 back into list

20:17 CL
  • 300 modes
  • gains 1.2, 0.4, 0.2
  • 201744
  • flat test_201744_300
    • 190 modes

20:29 CL
  • 300 modes
  • gains 1.2, 0.4, 0.3
  • 202945
  • small bump in delta command at 5:00
  • flat test_202645_300

20:38 CL
  • 400 modes
  • gains 1.0, 0.3, 0.2
  • 203917
  • flat test_203917_400
    • 188

20:20

IDL> big = intarr(n_elements(force_vec[0,*])) 
IDL> for i = 1, n_elements(force_vec[0,*])-1 do big[i] = where(abs(reform(force_vec[cl_act,i])) EQ max(abs(reform(force_vec[cl_act,i]))))
IDL> value = fltarr(n_elements(force_vec[0,*]))
IDL> for i = 1, n_elements(force_vec[0,*])-1 do value[i] = max(abs(reform(force_vec[cl_act,i])))
IDL> print, cl_act[big[188]]
     455
IDL> mostra, cl_act[big[(where(value GE 0.3))[0]]]                                                              
%IDL> mostra, 455            
IDL> display, reform(force_vec[cl_act,(where(value GE 0.3))[0]]), cl_act, /sh, /no_n, max_v=-0.2, no_smooth=0
IDL> display, reform(force_vec[cl_act,(where(value GE 0.3))[0]]), cl_act, /sh,  max_v=-0.2, no_smooth=0
IDL> plot, big, value, yrange=[0,0.5], psym=4
IDL> plot, value, big          

23:03 CL
  • 300 modes
  • 230237
  • test_230237_300
    • 224 modes

23:05 CL
  • 300 modes
  • flat test_230700_300
    • 223 modes
  • CL
  • flat test_231200_300
    • 224 modes
  • CL
  • flat test_231600_300
    • 230 modes
  • CL
  • flat test_231900_400
    • 224

23:24 CL
  • 400 modes
  • gain 0.85, 0.4, 0.3
  • 232427
    • test_232427_400
  • 232550
    • test_232550_400
  • 232633
    • test_232633_400
  • 232746
    • test_232746
  • 232833
    • test_232746

00:08 Original
  • default flat
    • 138 modes
  • test_232833_400.sav
    • 172, 182, 174

-- %USERSIG{DougMiller - 2017-04-07}%
Topic revision: r4 - 08 Apr 2017, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.