SX LBTB AO Science

08:00 Send binocular ACE preset to gravitational lens system

08:10 Fail on both sides during Optimize Gain
  • IDL variable not defined gainv

08:15 Send bino preset in Intervention mode
  • Optimize Gain works now
  • Loop Closed
  • Gain 3.4, 1.3, .8
  • bin 4
  • 36 modes
  • 100 Hz

08:26 Opt Loop Data
  • 082611

08:30 DX process disconnects from msg
  • We were collecting Optical Loop Data in closed loop
  • counter froze at 187 samples of 2000
[AOeng@flao-dxwfs UAO]$ w_check
Configuration directory: /home/aoeng/aoroot/conf/wfs/current/processConf

msgdrtdb             running
mirrorctrl           running but NOT connected to Msgd
pinger               running but NOT connected to Msgd
masterdiagnostic     running but NOT connected to Msgd
slopecompctrl        running but NOT connected to Msgd
techviewctrl         running but NOT connected to Msgd
powerboard           running but NOT connected to Msgd
picctrl              running but NOT connected to Msgd
ccd39                running but NOT connected to Msgd
ccd47                running but NOT connected to Msgd
ttctrl               running but NOT connected to Msgd
filterwheel1         running but NOT connected to Msgd
filterwheel2         running but NOT connected to Msgd
adc1                 running but NOT connected to Msgd
adc2                 running but NOT connected to Msgd
rerotator            running but NOT connected to Msgd
mercury              running but NOT connected to Msgd
cubestage            running but NOT connected to Msgd
baysidex             running but NOT connected to Msgd
baysidey             running but NOT connected to Msgd
baysidez             running but NOT connected to Msgd
cameralensctrl       running but NOT connected to Msgd
housekeeperWFS       NOT running
optloopdiag          NOT running
od_gopt              NOT running
truesense            NOT running
wfsarb               running but NOT connected to Msgd
dbfiller             running but NOT connected to Msgd
gopt                 NOT running
[AOeng@flao-dxwfs UAO]$ 
[

08:35 DX AdSec in failed mode
  • adsc_stop/start

08:45 bino preset
  • center pupil failed
    • center star with silver mirror then when move to dichroic the star shifts CCD47 and on CCD39, off pyramid
    • no signal so loop diverge
  • manually center star with dichroic in
  • click Center Pupils and continue
  • no problems
  • LUCI Script running

09:05 Bad actuator on SX FLAO GUI
  • 150
  • Clicked Remove

09
ER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_185|INF|       856|2017-03-30 09:40:25.343874|             MAIN > Received cmd: PauseAo()
AdSecArbitrator_185|INF|       857|2017-03-30 09:40:25.343937|   COMMANDHANDLER > FSM (status: AORunning) has received command 2027 (Pause Ao)
AdSecArbitrator_185|INF|       858|2017-03-30 09:40:25.359440|   COMMANDHANDLER > Command Pause Ao (code 2027) successfully completed
AdSecArbitrator_185|INF|       859|2017-03-30 09:40:40.274371|             MAIN > Received cmd: ResumeAo()
AdSecArbitrator_185|INF|       860|2017-03-30 09:40:40.274485|   COMMANDHANDLER > FSM (status: AOPause) has received command 2028 (ResumeAo Ao)
AdSecArbitrator_185|INF|       861|2017-03-30 09:40:40.277385|   COMMANDHANDLER > Command ResumeAo Ao (code 2028) successfully completed
AdSecArbitrator_185|INF|       862|2017-03-30 09:40:40.584001|             MAIN > Received cmd: Offloading(1)
AdSecArbitrator_185|INF|       863|2017-03-30 09:40:40.584115|   COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_185|INF|       864|2017-03-30 09:40:40.587978|   COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
20173003 09:41:01.8279 Alert: HighOrderOffload: HighOrderOffload vector ready
20173003 09:41:21.9705 Alert: HighOrderOffload: HighOrderOffload vector ready
20173003 09:41:42.1394 Alert: HighOrderOffload: HighOrderOffload vector ready
AdSecArbitrator_185|INF|       865|2017-03-30 09:41:57.615492|             MAIN > Received cmd: Offloading(0)
AdSecArbitrator_185|INF|       866|2017-03-30 09:41:57.615544|   COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_185|INF|       867|2017-03-30 09:41:57.617279|   COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_185|INF|       868|2017-03-30 09:41:57.617553|             MAIN > Received cmd: PauseAo()

09:53 Close the loop manually
  • Since failed on an offset, IIF is still thinks the Loop is Paused
  • Close the loop manually
  • Pause the AO loop on the AdSec Control GUI and on the WFS Control GUI
  • Click Resume on AOS COmmand GUI.
  • Success and IIF realized the AO Loop is closed

10:00 Script Finished
  • Opt Loop Data 100013

10:02 New Target
  • bin 2
  • 990 Hz
  • 153
  • gain 1.88, 0.44, 0.44
  • 101148

10:35 Shell Rip
  • No elevation info
  • anemometer process die
  • AOSGUI Says FLAO disconnected
  • Stop start aoarb and adsecarb. No comm
  • Stop/Start AOSL
    • elevation available
IN > Received cmd: Offloading(0)
AdSecArbitrator_339|INF|       730|2017-03-30 10:35:52.844736|   COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_339|INF|       731|2017-03-30 10:35:52.862948|   COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_339|INF|       732|2017-03-30 10:35:52.863124|             MAIN > Received cmd: PauseAo()
AdSecArbitrator_339|INF|       733|2017-03-30 10:35:52.863197|   COMMANDHANDLER > FSM (status: AORunning) has received command 2027 (Pause Ao)
AdSecArbitrator_339|INF|       734|2017-03-30 10:35:52.867559|   COMMANDHANDLER > Command Pause Ao (code 2027) successfully completed
AdSecArbitrator_339|WAR|       735|2017-03-30 10:35:57.875434|             MAIN > Windspeed not available!
AdSecArbitrator_339|INF|       736|2017-03-30 10:35:57.875492|             MAIN > Enabling TSS
AdSecArbitrator_339|INF|       737|2017-03-30 10:35:57.875512|      ADAM-MODBUS > AdamModbus: enabling tss...
AdSecArbitrator_339|ERR|       738|2017-03-30 10:35:57.875615|      ADAM-MODBUS > [TcpReceiveException] Error receiving data (errno=0 ), socket closed. (code -15003) TcpConnection receive error File TcpConnection.cpp line 131
AdSecArbitrator_339|ERR|       739|2017-03-30 10:35:57.875660|      ADAM-MODBUS > SetSingleCoil: failed to RecvForceSingleCoil
AdSecArbitrator_339|ERR|       740|2017-03-30 10:35:57.902580|             MAIN > Elevation not available! Resting the shell.
AdSecArbitrator_339|INF|       741|2017-03-30 10:35:58.828797|   COMMANDHANDLER > forceRest(): disabling coils
AdSecArbitrator_339|INF|       742|2017-03-30 10:35:58.828869|      ADAM-MODBUS > AdamModbus: disabling coils...
AdSecArbitrator_339|INF|       743|2017-03-30 10:35:58.843879|   COMMANDHANDLER > FSM (status: AOPause) has received command 2010 (Rest)
AdSecArbitrator_339|INF|       744|2017-03-30 10:35:59.003204|      ADAM-MODBUS > AdamModbus: disabling coils...
idl_14378          |INF|      2093|2017-03-30 10:35:59.018110|             MAIN > Set  WFS port #0 on switch BCU
idl_14378          |INF|      2094|2017-03-30 10:35:59.019058|             MAIN > WFS port #0 on switch BCU successfully set.
idl_14378          |INF|      2095|2017-03-30 10:35:59.020003|             MAIN > Settings:
idl_14378          |INF|      2096|2017-03-30 10:35:59.020078|             MAIN > Skip threshold: MAX DELTA MODE   0.000260000
idl_14378          |INF|      2097|2017-03-30 10:35:59.020129|             MAIN > Skip threshold: MAX ABS MODE     0.000130000
idl_14378          |INF|      2098|2017-03-30 10:35:59.020173|             MAIN > Skip threshold: DELTA COMMAND    1.00000e-05
idl_14378          |INF|      2099|2017-03-30 10:35:59.020210|             MAIN > Skip threshold: MAX COMMAND      0.000120000
idl_14378          |INF|      2100|2017-03-30 10:35:59.020246|             MAIN > Skip threshold: MIN COMMAND          0.00000
idl_14378          |INF|      2101|2017-03-30 10:35:59.020299|             MAIN > Skip threshold: MAX FORCE           0.770000
idl_14378          |INF|      2102|2017-03-30 10:35:59.020335|             MAIN > Skip threshold: MIN FORCE          -0.770000
idl_14378          |INF|      2103|2017-03-30 10:35:59.029880|             MAIN > Safe skip thresholds successfully set.
idl_14378          |INF|      2104|2017-03-30 10:35:59.029917|             MAIN > Installing error handling code
idl_14378          |INF|      2105|2017-03-30 10:35:59.029943|             MAIN > Reset optical loop gain to zero
idl_14378          |INF|      2106|2017-03-30 10:35:59.033437|             MAIN > Disable actuators coils with ADAM
idl_14378          |ERR|      2107|2017-03-30 10:35:59.033538|      ADAM-MODBUS > [TcpReceiveException] Error receiving data (errno=0 ), socket closed. (code -15003) TcpConnection receive error File TcpConnection.cpp line 131
idl_14378          |INF|      2108|2017-03-30 10:35:59.138365|             MAIN > Coils disabled. Cleaning currents
idl_14378          |INF|      2109|2017-03-30 10:36:00.614704|             MAIN > Recontructor cleaning started...
idl_14378          |INF|      2110|2017-03-30 10:36:01.279724|             MAIN > Recontructor cleaning done.
idl_14378          |INF|      2111|2017-03-30 10:36:01.279869|             MAIN > Start sending diagnostic frames... 
idl_14378          |INF|      2112|2017-03-30 10:36:01.280552|             MAIN > Settings:
idl_14378          |INF|      2113|2017-03-30 10:36:01.280619|             MAIN > Skip threshold: MAX DELTA MODE   0.000260000
idl_14378          |INF|      2114|2017-03-30 10:36:01.280651|             MAIN > Skip threshold: MAX ABS MODE     0.000130000
idl_14378          |INF|      2115|2017-03-30 10:36:01.280693|             MAIN > Skip threshold: DELTA COMMAND    1.00000e-05
idl_14378          |INF|      2116|2017-03-30 10:36:01.280723|             MAIN > Skip threshold: MAX COMMAND      0.000120000
idl_14378          |INF|      2117|2017-03-30 10:36:01.280750|             MAIN > Skip threshold: MIN COMMAND          0.00000
idl_14378          |INF|      2118|2017-03-30 10:36:01.280775|             MAIN > Skip threshold: MAX FORCE           0.770000
idl_14378          |INF|      2119|2017-03-30 10:36:01.280800|             MAIN > Skip threshold: MIN FORCE          -0.770000
idl_14378          |INF|      2120|2017-03-30 10:36:01.307377|             MAIN > Done
idl_14378          |INF|      2121|2017-03-30 10:36:01.310547|             MAIN > Trigger disabled
idl_14378          |INF|      2122|2017-03-30 10:36:01.311017|             MAIN > Offset disabled
idl_14378          |INF|      2123|2017-03-30 10:36:01.311070|             MAIN > IF Trigger successfully set
idl_14378          |INF|      2124|2017-03-30 10:36:01.313650|             MAIN > Accelerometers successfully set
idl_14378          |INF|      2125|2017-03-30 10:36:01.313765|             MAIN > Update progress successfully completed (2)
idl_14378          |INF|      2126|2017-03-30 10:36:01.313813|             MAIN > Change RMS thresholds to avoid spurious diagnostic dumps when the shell is REST.
idl_14378          |INF|      2127|2017-03-30 10:36:01.328384|             MAIN > RMS thresholds successfully changed.
idl_14378          |INF|      2128|2017-03-30 10:36:01.333944|             MAIN > Oversampling frames disabled.
idl_14378          |INF|      2129|2017-03-30 10:36:01.334004|             MAIN > Disabling DSP WATCHDOG...
idl_14378          |INF|      2130|2017-03-30 10:36:01.336972|             MAIN > ... done.
idl_14378          |INF|      2131|2017-03-30 10:36:01.337378|             MAIN > Uploading reference wave paramenters...
idl_14378          |INF|      2132|2017-03-30 10:36:01.468550|             MAIN > ... done.
idl_14378          |INF|      2133|2017-03-30 10:36:01.468597|             MAIN > Enabling DSP WATCHDOG...
idl_14378          |INF|      2134|2017-03-30 10:36:01.471602|             MAIN > ... done.
idl_14378          |INF|      2135|2017-03-30 10:36:01.634638|             MAIN > Oversampling frames disabled.
idl_14378          |INF|      2136|2017-03-30 10:36:01.634742|             MAIN > Reference wave decreased
idl_14378          |INF|      2137|2017-03-30 10:36:01.748720|             MAIN > No misfunctioning capacitive sensors found.
idl_14378          |INF|      2138|2017-03-30 10:36:01.748954|             MAIN > No gap contamination found.
idl_14378          |INF|      2139|2017-03-30 10:36:01.749134|             MAIN > Dust test successfully completed
idl_14378          |INF|      2140|2017-03-30 10:36:01.749253|             MAIN > Update progress successfully completed (2)
idl_14378          |INF|      2141|2017-03-30 10:36:01.749383|             MAIN > FSM Rip done.
AdSecArbitrator_339|INF|       745|2017-03-30 10:36:02.741619|   COMMANDHANDLER > Command Rest (code 2010) successfully completed
20173003 10:36:03.5101 Alert: Error: COILS DISABLED
AdSecArbitrator_339|WAR|       746|2017-03-30 10:36:04.513038|             MAIN > Windspeed not available!
AdSecArbitrator_339|INF|       749|2017-03-30 10:36:06.689576|             MAIN > Received cmd: ResumeAo()
Last message repeated
AdSecArbitrator_339|INF|       750|2017-03-30 10:36:06.689647|   COMMANDHANDLER > FSM (status: Ready) has received command 2028 (ResumeAo Ao)
AdSecArbitrator_339|ERR|       751|2017-03-30 10:36:06.689688|   COMMANDHANDLER > adsecarb: Fsm is discarding command 2028 (ResumeAo Ao)...
AdSecArbitrator_339|WAR|       752|2017-03-30 10:36:06.689741|             MAIN > ResumeAo: WARNING - adsecarb:Illegal command for state Ready (ResumeAo Ao)
AdSecArbitrator_339|WAR|       753|2017-03-30 10:36:07.531583|             MAIN > Windspeed not available!
20173003 10:36:26.5625 Alert: HighOrderOffload: HighOrderOffload vector ready
20173003 10:36:46.7617 Alert: HighOrderOffload: HighOrderOffload vector ready
20173003 10:37:06.9635 Alert: HighOrderOffload: HighOrderOffload vector ready

11:21 DX Loop open with safe skip
  • Not sure why
  • IIF thinks loop is Closed because pauseAO failed
  • Close loop manually
  • DX side is not closed and IIF thinks/knows it is closed
  • SX side is paused
    • have to move bayside stages to get star on pyramid
    • Resume from AOS Command GUI
    • SX is now closed and IIF thinks/knows it is closed
  • Resume LUCI script: next offset fine.
 received command 2029 (TTOffload)
AdSecArbitrator_153|INF|      1121|2017-03-30 11:21:35.849661|   COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
20173003 11:21:55.1952 Alert: HighOrderOffload: HighOrderOffload vector ready
AdSecArbitrator_153|WAR|      1122|2017-03-30 11:21:56.330413|             MAIN > Skip frame detected!
AdSecArbitrator_153|INF|      1135|2017-03-30 11:22:08.979571|             MAIN > Received cmd: StopAo(1)
Last message repeated
idl_15235          |INF|       821|2017-03-30 11:22:08.980089|             MAIN > fsm_stop_ao start
idl_15235          |INF|       822|2017-03-30 11:22:08.980139|             MAIN > fsm_skip_recovery FORCE set
AdSecArbitrator_153|INF|      1136|2017-03-30 11:22:08.979655|   COMMANDHANDLER > FSM (status: AORunning) has received command 2015 (Stop Ao)
AdSecArbitrator_153|INF|      1137|2017-03-30 11:22:08.979686|    SERIALIZATION > StopAoImpl(): restoreShape = 1
idl_15235          |INF|       823|2017-03-30 11:22:09.195164|             MAIN > Null Gain application and start sending null slope...
idl_15235          |INF|       824|2017-03-30 11:22:09.321687|             MAIN > Cleaned last slopes command.

12:15 Script done

12:20 take images with LUCI2 with NCPA off and on for last object

12:27 Bright star, 8.8
  • 124109
  • 124600
  • OH1060/Z filters
  • 10 x2.5 sec
  • 1 113, 2 121 Off
  • 2 122 ON
  • offset
  • 1 114, 115 2 123 124

12:46 FeII filter
  • NCPA ON
  • 1 116, 2 125
  • NCPA OFF
  • 1 117, 2 126
  • 124820

-- %USERSIG{DougMiller - 2017-03-30}%
Topic revision: r2 - 30 Mar 2017, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.