SX AO Observations

07:27
  • Getting all GUIs started and such.
07:40
  • Had spurious TS8 error that resolved itself
  • CCD 47 error in which it said "Ready", but never went live.
  • Sun Mar 26 07:59:30.223 2017 aos.AcquireRef{FLAO}.failed - AcquireRef {FLAO}[204] - [AOException] centerPupils: WARNING - RETRY: Camera lens temperature too low. T= 2.42, minimum allowed is 8.00 (0) NO_ERROR
  • Actuator 259 may be causing issues.
0819
  • In bin 4, 36 modes, 100 Hz, CCD 39 says Mag 15.9, should be 14.7....
  • Optical Loop Data: 20170326_082036
  • Removed Actuator 259 from position and icc, It was causing a large bump in the adsec. Also changed it in confcalib. The change will take effect when we restart the processes.
08:51
  • Slewed to new target.
  • All processes disconnected and I had to restart everything on the WFS. I used my checout routine because it can kill disconnected processes.
  • Sun Mar 26 09:07:48.862 2017 aos.Resume{FLAO}.failed - Resume{FLAO}[198] - [AOException] RETRY: Magnitude calculation cannot take place with negative counts (-5002) VALUE_OUT_OF_RANGE_ERROR
  • Started safe-skipping.....
  • WTH
09:14
  • Preset working fine, bin 2, 153 modes, 400 Hz.
  • Massive Vibrations! The image was shaking all over the place. We just passed zenith.