03:15 ACE preset to both sides

03:31 Bayside X stage not home.
  • Homed on first manual try
Thu Nov 24 03:31:16.688 2016 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[975] - FLAO: unknown exception from arbitrator

WfsArbitrator_36452|INF|       194|2016-11-24 03:30:24.293583|             MAIN > calibTVDarkInternal(): done
WfsArbitrator_36452|ERR|       195|2016-11-24 03:31:54.316127|             MAIN > Timeout error on baysidex (-5001) TIMEOUT_ERROR
WfsArbitrator_36452|DEB|       196|2016-11-24 03:32:15.417766|             MAIN > arbFunc updateNcpaInternal

03:37 DX RunAO
  • failed
WfsArbitrator_75957|INF|       330|2016-11-24 03:45:11.918011|             MAIN > Starting gain optimization procedure
WfsArbitrator_75957|ERR|       331|2016-11-24 03:45:11.953775|             MAIN > [Errno 13] Permission denied: '/local/aomeas/adsec_data/20161124'
  • Changed permissions via flao account

03:45 SX RunAO failed because no slopes to AdSec
  • try cycling power to BCU
  • cycle slope process
  • No luck
  • Perform w_stop/start and power on. Success

05:15 Going to switch to binocular AO but RFBG rotator faulted twice.
  • no more RFBG tonight.

05:20 Start SX LUCI1/AO Commisioning

06:27 Loop open,
  • bin 3
  • Seeing spiked to 1.7 from 1.0 to 1.3.
  • Elevation 45
AdSecArbitrator_119|WAR|       888|2016-11-24 06:27:55.522331|             MAIN > Skip frame detected!
AdSecArbitrator_119|INF|       893|2016-11-24 06:27:59.663770|             MAIN > Received cmd: StopAo(1)
Last message repeated
AdSecArbitrator_119|INF|       894|2016-11-24 06:27:59.663838|   COMMANDHANDLER > FSM (status: AORunning) has received command 2015 (Stop Ao)
idl_119648         |INF|      1469|2016-11-24 06:27:59.664213|             MAIN > fsm_stop_ao start
idl_119648         |INF|      1470|2016-11-24 06:27:59.664282|             MAIN > fsm_skip_recovery FORCE set
  • Seeing calmed down to 1.2 so try again

07:21 Move to trapesium, bin 1

07:41 Opt Loop data
  • bin 1
  • 400 modes
  • NCPA on
  • 074115
  • 075056
  • 081344
  • 081520 HO2 gain down by 0.5 to 0.5
  • 081721 HO2 gain up by 0.5 to 0.9
  • 081935 NCPA off HO2 gain down by 0.5 to 0.5
  • 082129 GOPT off, rms=0.16
  • 082325 GOPT off, HO1 gain down .1 to 0.3
  • 082456 GOPT off, HO1 gain up .1 to 0.5
  • 082635 GOPT on, rms=1.1
  • 083036 GOPT off. Take another cluster offsets on Trapesium to compare to above data (081344), rms=0.17

08:36 After offset, loop fluctuate wildly, safe skip, loop open. Pupil position was OK

08:39 PresetAO failed
WfsArbitrator_55524|INF|      2246|2016-11-24 08:39:51.877343|             MAIN > Setting tiptilt to amp= 0.00, freq= 0.00, offx= 0.00, offy= 0.00
WfsArbitrator_55524|ERR|      2247|2016-11-24 08:39:51.928324|             MAIN > [Errno 4] Interrupted system call

08:47 Preset
  • GOPT off
  • rms = 0.15
  • 084906
  • rms = 0.15
  • 085422

08:59 Conclude that the analysis package is incorrect for GOPT on.
  • image quality from LUCI is a bit small and rounder with GOPT, NCPA on than without

09:02 Trapesium, bin 1, NCPA on
  • 090259 0.8
  • 091448 ?
  • 091630 slight striping 1.0

09:22 Preset with 300 modes, NCPA on, 0.7" seeing
  • 092839
  • 093131
  • 093429 HO2 up 0.5
  • 094715
  • 095016 0.96
  • 095156 1.2"
  • 095448 0.9"
  • 100520 1,0"

10:16 Preset bin 4
  • 102721 1.0"

12:31 ACE Preset to AO302
  • 6.8 mag catgalog
  • 9.0 mag measured (sirus)
  • bin 1
  • 300 modes
  • 69 elevation
  • NCPA on
  • BG 50
    • //dataA/readout1/DATA/20161124/luci1.20161124.0258.fits
  • Data 50
    • 125517
    • luci1.20161124.0259.fits
    • 125704
    • luci1.20161124.0260.fits (0.6" on guider) maybe .95"
    • 125834
    • luci1.20161124.0261.fits
  • GB 50
    • luci1.20161124.0262.fits

13:10 ACE Preset to AO302
  • 6.8 mag catgalog
  • 8.6 mag measured (sirus)
  • bin 1
  • 400 modes
  • 69 elevation
  • NCPA off
  • Data 50
    • 131352
    • luci1.20161124.0264.fits
    • 131445
    • luci1.20161124.0265.fits
  • Data 50 NCPA ON
    • 131539
    • luci1.20161124.0266.fits (may be ok, but loop opened)

-- %USERSIG{DougMiller - 2016-11-24}%