LUCI1/AO Commissioning

02:00 Start up AO System
  • Had to manually reset CCD47
  • After that, all worked properly

02:25 Preset to 14.3 mag star
  • bin 4
  • 200 Hz
  • 55 elevation

02:28 RunAO: successful
  • Offsets every minute

02:47 Bin 4, 200 Hz * 024758

02:53 Bayside stage timeout
  • Retry success
WfsArbitrator_45489|INF|       499|2016-11-23 02:53:38.744162|             MAIN > offsetXYZ(): Focus correction (absolute): 0.000 mm
WfsArbitrator_45489|INF|       500|2016-11-23 02:54:18.773172|             MAIN > Stage movement timeout after 40 seconds. Retry 1 of 3
WfsArbitrator_45489|INF|       501|2016-11-23 02:54:24.473182|             MAIN > offsetXY(): done

02:54 025444

03:17 031737

04:24 After offset could see oscillation on the Pyramid
  • Bin 1
  • NCPA on
  • 990 Hz
20162311 04:24:06.2880 Alert: HighOrderOffload: HighOrderOffload vector ready
AdSecArbitrator_110|WAR|       998|2016-11-23 04:24:15.479466|             MAIN > Skip frame detected!
AdSecArbitrator_110|INF|      1004|2016-11-23 04:24:21.891772|             MAIN > Received cmd: StopAo(1)
Last message repeated
AdSecArbitrator_110|INF|      1005|2016-11-23 04:24:21.891864|   COMMANDHANDLER > FSM (status: AORunning) has received command 2015 (Stop Ao)
idl_110145         |INF|       827|2016-11-23 04:24:21.892196|             MAIN > fsm_stop_ao start
idl_110145         |INF|       828|2016-11-23 04:24:21.892248|             MAIN > fsm_skip_recovery FORCE set
AdSecArbitrator_110|INF|      1006|2016-11-23 04:24:21.891906|    SERIALIZATION > StopAoImpl(): restoreShape = 1
idl_110145         |INF|       829|2016-11-23 04:24:22.107816|             MAIN > Null Gain application and start sending null slope...
idl_110145         |INF|       830|2016-11-23 04:24:22.237330|             MAIN > Cleaned last slopes command.
idl_110145         |INF|       831|2016-11-23 04:24:22.237451|             MAIN > Modes out of safe range    ID: # -1
idl_110145         |INF|       832|2016-11-23 04:24:22.237521|             MAIN > Commands out of safe range ID: # -1
idl_110145         |INF|       833|2016-11-23 04:24:22.237575|             MAIN > Forces out of safe range   ID: # 26 # 54 # 82
idl_110145         |INF|       834|2016-11-23 04:24:22.237647|             MAIN > SafeSkipFrames caused by forces out of range detected.
idl_110145         |INF|       835|2016-11-23 04:24:22.237685|             MAIN > Recovery from any SafeSkipFrames started...
idl_110145         |INF|       836|2016-11-23 04:24:22.237725|             MAIN > Getting info on actual mirror position...
idl_110145         |INF|       837|2016-11-23 04:24:22.241244|             MAIN > Reconstructor filter cleaning...
idl_110145         |INF|       838|2016-11-23 04:24:22.247206|             MAIN > Control current integrator cleaning up...
idl_110145         |INF|       839|2016-11-23 04:24:22.266541|             MAIN > Null new delta command contribution
idl_110145         |INF|       841|2016-11-23 04:24:22.269269|             MAIN > Restore old oversampling...
idl_110145         |INF|       842|2016-11-23 04:24:22.276828|             MAIN > Done.
idl_110145         |INF|       843|2016-11-23 04:24:22.341238|             MAIN > Restore control current integrator action....
idl_110145         |INF|       844|2016-11-23 04:24:22.386844|             MAIN > Done.
idl_110145         |INF|       845|2016-11-23 04:24:22.386902|             MAIN > Reload last shape selected
idl_110145         |INF|       846|2016-11-23 04:24:23.452493|             MAIN > Loading shape: /local/aomeas/adsec_calib/flat/Applicable/LUCIFER.sav
idl_110145         |INF|       847|2016-11-23 04:24:23.454603|             MAIN > Selected flat: /local/aomeas/adsec_calib/flat/Applicable/LUCIFER.sav
idl_110145         |INF|       848|2016-11-23 04:24:24.570464|             MAIN > The threshold has been moved to 0.2 as the total forces are included in the computation of the modal filtering.
idl_110145         |INF|       849|2016-11-23 04:24:24.570540|             MAIN > The threshold has been additionally moved to 0.5 to try flattening better the external edge.
idl_110145         |INF|       850|2016-11-23 04:24:24.570628|             MAIN > Delta command modal filtering for reducing peak force to 0.30 [N]
idl_110145         |INF|       851|2016-11-23 04:24:24.604449|             MAIN > total current added in the force vector for the computation of the modal filtering
idl_110145         |INF|       852|2016-11-23 04:24:24.614203|             MAIN > Maximum mode index used to apply flat -> # 505
idl_110145         |INF|       853|2016-11-23 04:24:24.615126|             MAIN > Min/Max delta command to apply for flattening -> #  -4.01577e-06 /    5.54540e-06 [m]
idl_110145         |INF|       854|2016-11-23 04:24:24.760440|             MAIN > COMM  MAX [m] 6.88322e-05
idl_110145         |INF|       855|2016-11-23 04:24:24.760526|             MAIN > FORCE MAX [N] 0.298078

04:32 Opt Loop Data
  • bin 2
  • 990 Hz
  • NCPA on
  • 043253

04:33 Offset failed
   
20162311 04:33:22.7320 Alert: HighOrderOffload: HighOrderOffload vector ready
AdSecArbitrator_133|WAR|       138|2016-11-23 04:33:27.749832|             MAIN > Skip frame detected!
AdSecArbitrator_133|INF|       148|2016-11-23 04:33:39.487573|             MAIN > Received cmd: StopAo(1)
Last message repeated
AdSecArbitrator_133|INF|       149|2016-11-23 04:33:39.487651|   COMMANDHANDLER > FSM (status: AORunning) has received command 2015 (Stop Ao)
AdSecArbitrator_133|INF|       150|2016-11-23 04:33:39.487709|    SERIALIZATION > StopAoImpl(): restoreShape = 1
idl_110145         |INF|       943|2016-11-23 04:33:39.488085|             MAIN > fsm_stop_ao start
idl_110145         |INF|       944|2016-11-23 04:33:39.488162|             MAIN > fsm_skip_re

WfsArbitrator_45489|INF|      1444|2016-11-23 04:36:47.564862|             MAIN > Setting tiptilt to amp= 0.00, freq= 0.00, offx= 0.00, offy= 0.00
WfsArbitrator_45489|ERR|      1445|2016-11-23 04:36:47.588787|             MAIN > [Errno 4] Interrupted system call

04:40 Resend Preset
  • Success

04:49 Opt Loop Data
  • bin 2
  • 990 Hz
  • NCPA on

Wed Nov 23 04:52:20.361 2016 aos.info - WFS source set to: FLAOWFS
Wed Nov 23 04:54:30.382 2016 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[235] - FLAO: unknown exception from arbitrator

04:58 Resend Preset
Wed Nov 23 04:59:37.725 2016 aos.Stop{FLAO}.started - Stop{FLAO}[236] (msg:Starting Preset)
Wed Nov 23 05:00:27.752 2016 aos.Stop{FLAO}.failed - Stop{FLAO}[236] - FLAO: unknown exception from arbitrator
Wed Nov 23 05:00:32.177 2016 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[237] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:-4.48773[-4.48773],soy:4.48773,rox:-4.59735[-4.59735],roy:4.65413,mag:9.19,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Nov 23 05:00:32.188 2016 aos.info - WFS source set to: FLAOWFS
Wed Nov 23 05:02:42.209 2016 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[237] - FLAO: unknown exception from arbitrator

05:03 Stopped/Started the AO Arbitrator
  • Now PresetAO from AOS command GUI works properly

05:10 Preset
  • bin 2
  • No NCPA
  • No trouble for 4 offsets
  • Opt Loop 051617
  • During the previous failures

07:24 Trapesium
  • bin 1
  • 9.1 mag
  • 072357
  • Skip Frames on first offset (40")
AdSecArbitrator_133|INF|      1256|2016-11-23 07:25:17.411871|   COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_133|INF|      1257|2016-11-23 07:25:17.800714|   COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_133|WAR|      1258|2016-11-23 07:25:18.368575|             MAIN > Skip frame detected!
AdSecArbitrator_133|WAR|      1265|2016-11-23 07:25:27.368708|             MAIN > Cannot dispatch alert: TcpTransceiver.cpp:284: Ice::ConnectionLostException:
Last message repeated

07:36 Same skip frame on first offset (40")

07:50 Same skip frame on first offset * Turn on Opt Gain but on NCPA * 075534

07:58 Turn on NCPA
  • 075859
  • offset
  • 080042
  • offset
  • 080233

08:34 During normal operation (no offset) the Optical gain was 0.2 and Optical Gain ratio became large, 1.4, and gain set to small value.
  • Skip Frame and the open.
python_45522       |INF|      1962|2016-11-23 08:34:51.770745|             MAIN > Tracknum: 20161123_250000 demodDelta/demodCmd: 1e-08/4e-06 gopt: 0.1
python_45522       |INF|      1963|2016-11-23 08:34:58.102968|             MAIN > Tracknum: 20161123_250000 demodDelta/demodCmd: 1e-08/3e-06 gopt: 0.1
python_45522       |INF|      1964|2016-11-23 08:35:04.437125|             MAIN > Tracknum: 20161123_250000 demodDelta/demodCmd: 1e-08/2e-05 gopt: 0.1
python_45522       |INF|      1965|2016-11-23 08:35:10.827941|             MAIN > Tracknum: 20161123_250000 demodDelta/demodCmd: 1e-08/1e-05 gopt: 0.0

08:48 New Preset
  • bin 1

09:00 preset
  • NCPA on
  • GOPT gain 0.2
  • ratio 1.0
  • Offset 40"
  • After offset, pupil off by 0.5
  • get striping on pupils
  • safe skip and open loop
  • video 31 sec

09:10 preset same as above
  • NCPA off
  • GOPT on
  • Offset 40"
  • After offset, pupil off by 0.5
  • get striping on pupils
  • safe skip and open loop
  • video 28 sec

09:16 same preset as above
  • No NCPA
  • No Gopt
  • Same result, pupil off and skip frame

09:41 Preset
  • Open on 40" offset
  • reclose manually
  • NCPA on
  • continue with small offsets
  • Rolled edge at 11:00
  • Loop open

10:04 Opt loop data
  • Bin1
  • 300 modes
  • Gopt off *NCPA off
  • 990 Hz
  • 100422
  • 100732

11:34 Opt Loop Data
  • bin 3
  • 200 Hz
  • 113401

13:02 ACE Preset to AO307
  • bin 1
  • 8.9 mag
  • 300 modes
  • LUCI images
    • HeI and J cross filters
    • images 0271-3

13:16 ACE Preset to AO307
  • bin 1
  • 8.9 mag
  • 400 modes
  • LUCI images
    • HeI and J cross filters
    • 274-6

WfsArbitrator_12473|INF|      2670|2016-11-23 13:16:33.167342|             MAIN > Setting tiptilt to amp= 0.00, freq= 0.00, offx= 0.00, offy= 0.00
WfsArbitrator_12473|ERR|      2671|2016-11-23 13:16:33.203295|             MAIN > [Errno 4] Interrupted system call

-- %USERSIG{DougMiller - 2016-11-23}%
Topic revision: r3 - 23 Nov 2016, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback