Title

02:58 Measured center of rotation of AGw#2
  • gcsclient readGuideCam right -e 90000
  • /lbt/data/share/tcs/lbto.20160911/right_guider000002.fits
  • 127.2, 115.4
  • 254.4, 230.8 + 18-1, 16-1 = 271.4, 245.8,
  • want 275, 243
  • -> with accuracy that we can position the center
  • No Change

03:05 Pupil Wobble
  • iif_register, LBFG (or RFBG)
  • pupil_collect, retval, EXP=10.0

03:50 Started both AO WFSs
  • No problems

04:18 Preset to BS9177 for pupil wobble
  • pupil_collect, retval, EXP=20.0
  • Move +200 tip on both axes. Both wrong direction
  • See plots

06:11 WUniit transformation on SX with LUCI1 script

06:15
  • 27.54, -14.04
  • -105, -63

06:17
  • -44, -63 No Star

06:19
  • -43, -39 Star Found

06:21 Preset On Axis
  • OC_M29, GS=23, RS=0
  • Sun Sep 11 06:31:59.804 2016 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[41] - Command timeout - -13107: Arbitrator interface generic error

06:34 SX
  • -77.3, -49.1 Preset
  • -76.7, -50.4

06:34 DX
  • -75.2, -47.95 Preset
  • -72.68, -48.53

06:40 RunAO
  • Success both sides
  • SX 065037
  • DX 065020
  • Looks like overdrive Tip/Tilt 1.7, 1.8. See two lopes on LUCI1 image
  • T/T to 1.5, 1.55
  • SX 065300
  • DX 065301
  • T/T to 1.2, 1.2
  • SX 065501
  • DX 065459
  • SX HO1/2 .59, .26 to .49, 21
  • SX 065718
  • DX HO1/2 .29, .31 to .29, .21
  • DX 065849

07:05 Optimize gain on both sides
  • SX
    • 1.05, .35, .26
    • 070703
  • DX
    • 1.41, 0.6, 0.49
    • 070756

07:08 DX Loop Open
  • adsec.R |INF| 10743|2016-09-11 07:08:05.136032| MAIN > Forces out of safe range ID: # 87
  • no jumping actuator

07:20 DX Preset with sethotspot 0,0. Failed
  • Sun Sep 11 07:18:47.498 2016 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[50] - Command timeout - -13107: Arbitrator interface generic error
  • WfsArbitator _26134|ERR| 528|2016-09-11 07:18:53.486138| MAIN > Timeout error on baysidex (-5001) TIMEOUT_ERROR

07:28 DX Preset with sethotspot 0,0. Failed

WfsArbitrator_26134|ERR|       607|2016-09-11 07:27:50.396497|             MAIN > Timeout error on baysidex (-5001) TIMEOUT_ERROR
WfsArbitrator_26134|ERR|       608|2016-09-11 07:27:50.396532|             MAIN > Traceback (most recent call last):
WfsArbitrator_26134|ERR|       609|2016-09-11 07:27:50.396552|             MAIN >   File "/home/flao/TEST/lib/python/AdOpt/WfsArbScripts.py", line 2993, in arbFunc
WfsArbitrator_26134|ERR|       610|2016-09-11 07:27:50.396571|             MAIN >     errstr = func(*args)
WfsArbitrator_26134|ERR|       611|2016-09-11 07:27:50.396589|             MAIN >   File "/home/flao/TEST/lib/python/AdOpt/WfsArbScripts.py", line 988, in doPrepareAcquireRef
WfsArbitrator_26134|ERR|       612|2016-09-11 07:27:50.396608|             MAIN >     app.stagex.waitTargetReached()
WfsArbitrator_26134|ERR|       613|2016-09-11 07:27:50.396643|             MAIN >   File "/home/aoacct/aoroot/lib/python/AdOpt/hwctrl/simplemotor.py", line 300, in waitTargetReached
WfsArbitrator_26134|ERR|       614|2016-09-11 07:27:50.396661|             MAIN > AdOptError: Timeout error on baysidex - code: -5001 TIMEOUT_ERRO

08:21 Ref Star 3 is 44" from Target
  • if_preset, OC_M29, MODE=ACE, GS=33, RS=3, TS=0, PA=45, /BOTH
    • Can See? Yes on both
  • if_preset, OC_M29, MODE=ACE, GS=20, RS=3, TS=0, PA=-45, /BOTH
    • Yes on SX
    • No on DX
  • if_preset, OC_M29, MODE=ACE, GS=12, RS=3, TS=0, PA=-135, /BOTH
    • Yes on SX
    • No on DX
  • if_preset, OC_M29, MODE=ACE, GS=27, RS=3, TS=0, PA=-135, /BOTH * Yes on SX * No on DX
  • if_preset, OC_M29, MODE=ACE, GS=13, RS=3, TS=0, PA=135, /BOTH * Yes on SX * Yes on DX

08:58 Set fw2_acq in SX table_LUCIFER_ACE_AO.txt

09:02 RunAO on a 17.5 Ref Star
  • Failed, safe skip
  • Just not enough light

10:52 RunAO failed
  • WfsArbitrator _18114|INF| 1951|2016-09-11 10:50:27.253586| MAIN > source_acq.acquire(): current error X=-0.041 mm, Y=1.608 mm
  • WfsArbitrator _18114|ERR| 1952|2016-09-11 10:50:43.589899| MAIN > Timeout waiting for baysidey.L.ENABLE.CUR (-5001) TIMEOUT_ERROR

11:06 Field Measurements
  • SX
                  lfbg
              From 45 Astigmatism Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip =  14.1", Delta IE = -14.1"
         Tilt = -11.0", Delta CA =  11.0"
  
  
                  lfbg
              From 0 Astigmatism Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip =  34.1", Delta IE = -34.1"
         Tilt = -18.1", Delta CA =  18.1"
  
                  lfbg
              From both Astigmatisms Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip =  24.7", Delta IE = -24.7"
         Tilt =  -0.9", Delta CA =   0.9"

*DX

                  rfbg
              From 45 Astigmatism Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip = -16.1", Delta IE =  16.1"
         Tilt = -35.7", Delta CA =  35.7"
  
  
                  rfbg
              From 0 Astigmatism Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip = -23.8", Delta IE =  23.8"
         Tilt = -27.4", Delta CA =  27.4"
  
                   rfbg
              From both Astigmatisms Measurement
                  On-Sky Arcseconds
         -------------------------------------------
          Tip =  -6.4", Delta IE =   6.4"
         Tilt = -35.6", Delta CA =  35.6"

11:12 SX deserialization error
  • AOARB.L |ERR| 90549|2016-09-11 11:12:56.916355| SERIALIZATION > Error in message deserialization from AOS.L (014000/2012)
  • RunAO take 5 minutes to timeout
  • Stop AO Arbitrator to force RunAO to fail.

11:24 SX RunAO failed

AOARB.L            |ERR|       127|2016-09-11 11:25:07.195611|    SERIALIZATION > Command stopped by external signal
AOARB.L            |INF|       128|2016-09-11 11:25:07.195670|   COMMANDHANDLER > Command execution failed: Fsm forced to Failure
AOARB.L            |INF|       129|2016-09-11 11:25:07.195686|   COMMANDHANDLER > forceFailure() opening loop...
AOARB.L            |INF|       130|2016-09-11 11:25:07.195722|            ADSEC > Requesting TTOffload to AdSecArbitrator
AOARB.L            |INF|       131|2016-09-11 11:25:09.415783|            ADSEC > Command returns success .
AOARB.L            |ERR|       132|2016-09-11 11:25:09.416064|            ADSEC > Command returns error : adsecarb:Illegal command for state AOSet (TTOffload)

11:27 SX RunAO failed

WfsArbitrator_18114|INF|      2438|2016-09-11 11:27:14.821922|             MAIN > source_acq.acquire(): current error X=-0.186 mm, Y=1.870 mm
WfsArbitrator_18114|ERR|      2439|2016-09-11 11:27:31.831998|             MAIN > Timeout waiting for baysidey.L.ENABLE.CUR (-5001) TIMEOUT_ERROR
  • Stopped and stated WFS Arbitrator
  • Nest RunAO get FLAOWFS> Command returns error but RunAO proceeds

11:43 RunAO fail waiting on baysidey
  • AOARB.L |ERR| 360|2016-09-11 11:42:17.505073| FLAOWFS > Command returns error .
  • AOARB.L |ERR| 361|2016-09-11 11:42:42.052249| FLAOWFS > Command returns error : Timeout waiting for baysidey.L.ENABLE.CUR (-5001) TIMEOUT_ERROR

11:45 bSX Opt Loop Data
  • Bin 4
  • 15.7 mag
  • 114939 (loop opened for second half of set)

11:52 SX RunAO failed
  • AOARB.L |ERR| 471|2016-09-11 11:51:49.599730| FLAOWFS > Command returns error : Timeout waiting for baysidey.L.ENABLE.CUR (-5001) TIMEOUT_ERROR
  • AOARB.L |WAR| 472|2016-09-11 11:51:49.599835| SERIALIZATION > AcquireRefAO on Wfs failed

11:56
  • bin 4
  • 16.3 mag
  • Opt Loop data 115900

12:01
  • bin 4
  • 15.8 mag
  • Opt Loop Data 120510

12:08
  • bin 4
  • 16.1 mag
  • Opt Loop Data 121136

12:15 LUCI images
  • luci1.20160911.0019.fits - luci1.20160911.0024.fits
  • luci1.20160911.0024.fits dark

12:40 Test remote LUCI acquisition
  • Have to type in password
  • /home/lucifer/lcsp/bin/commitReadout.sh not found
  • At luci.luci command line
    • % echo '/home/lucifer/lcsp/bin/commitReadout.sh wait pause 10.0000 1 0' | /bin/bash
    • Error while trying to reload DOM representation!

-- %USERSIG{DougMiller - 2016-09-11}%